Title: Process Validation
1Process Validation
- Case Study Heat Sealing Operation for Sterile
Product An Approach to Process Validation
2Regulatory Requirements (21CFR 820.75)
- Validate It
- Control It
- Revalidate When Appropriate
- Document Throughout
3Benefits of Validation
- Requires thorough understanding of a process
- Provides basis for Process Control
- Improves Product Quality and Reliability
- Reduces risk of patient hazard
- Improves Productivity
- Reduces adverse Regulatory Inspection Findings
- Reduces Field Actions / Complaints
4What Needs Validation?
- Where in-process tests have insufficient
sensitivity to verify the desired safety and
efficacy of finished devices - Where clinical or destructive tests are required
to verify the process produces the desired result - Where in-process tests do not reveal all
variations in safety and efficacy of finished
devices - Where process capability is unknown, or is
suspected barely capable - From FDA 1996 Compliance Guide
5Examples of Processes That Should Be Validated
- Sterilization Methods
- Clean Room Ambient Condition
- Aseptic Filling Procedures
- Sterile Packaging Sealing Processes
- Lyophilization Process
- Heat Treating Processes
- Plating Processes
- Plastic Injection Molding Process
- Welding Process
6Process Validation Decision Tree
7CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Pre-Validation
- From Design Requirements What aspects of the
product play a role in the process, e.g. - Sterilization Method
- Packaging
- Equipment Required to Seal
- Shipping Container
- Economics
- Identified Through the Manufacturing Plan Which
Is Developed in Early Design Stages
8CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Prepare Validation Plan
- Installation Qualification
- Software Validation
- Operational Qualification
- Process Characterization/Screening Experiment
- Gage RR as needed
- Performance Qualification
- Ongoing Process Controls
9CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Develop and Approve Protocols
- Approvals should include the appropriate
functional areas and levels - Define Success Criteria
- What do you expect from the process
- Prepare and Approve Reports
- Executive Summary and/or Conclusion Statement
- Was Success Criteria Met
- When to Revalidate
- Monitor the Process
10CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Installation Qualification (IQ)
- Confirm proper equipment installation, utilities,
and configuration/setup - Establish Maintenance and Calibration Needs
- Identify Worker Training/Safety Requirements
- Perform Tooling Qualifications Mini-IQ (if
applicable) - Define the tool/jig/fixture typically through a
drawing - Confirm it matches the drawing
- Are there any maintenance requirements for it
- Are there any utilities associated with it
- Are there any safety concerns
- Are there any measuring gages associated
11CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Sealer IQ protocol
- Engineering Documentation/Specification
(Instruction Manual, Standards e.g. ASTM, peel
pouch specs, etc.) - Electrical Utility Verification (Voltage, Safety
Cutoff, Overload Protection, Circuit Breakers,
etc.) - Instrument List Within Sealer (Thermocouple and
Timer Calibrations) - Preventive Maintenance Established (Manufacturer
Recommendations)
12CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Sealer IQ protocol cont
- Equipment Operation Procedure Drafted
- Software Validation Is a separate validation
needed - If the software is embedded factory software, it
shall be considered validated as a result of the
overall validation as long as all fault
conditions, alarms, and functionality is tested
within the validation program - If the software is a separate package such as a
Windows-based application, then a separate
software validation protocol may be required
13CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Sealer IQ protocol cont
- Equipment Control System Installation (on/off
switch, indicator lights, stops, breakers,
guards) - Test Instrument Control System Installation
(multimeter) - Any Discrepancies Resolved
14CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Process Characterization/Screening
- Designed experiment to identify key input
variables - Run process at different settings for the inputs
(trials) and measure the outputs - Determine which inputs affect the outputs
- For this example, the process inputs are in the
form of manufacturers recommendations for
temperature, hot dwell, and cool dwell - Outputs will be burst pressure and visual
examination of the seal
15CASE STUDY Heat Sealer Validation for Sterile
Packaging
- One example of characterization techniques is the
L-8 Array
Center Point
16CASE STUDY Heat Sealer Validation for Sterile
Packaging
17CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Operational Qualification (OQ)
- From the screening experiment you can determine
the inputs that contribute to the maximum output
variation, i.e., your worst-case condition - The OQ further challenges the process by
subjecting it to conditions such as anticipated
breakdowns like cleaning and maintenance, as well
as unanticipated conditions like a power failure,
etc.
18CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Operational Qualification (OQ)
- Measuring equipment needs to be appropriate for
intended use. One technique to determine this is
a Gage RR (Reference ASTM F1469-99 standard for
Repeatability and Reproducibility) - Gage Repeatability Reproducibility (Gage RR)
- Measuring instruments are subject to variation,
therefore, a process validation study cannot be
meaningful unless the process measuring equipment
possesses accuracy and precision
19CASE STUDY Heat Sealer Validation for Sterile
Packaging
- A Gage RR measures
- Repeatability Amount of variation obtained by
measuring the same unit (precision) with the
measuring equipment and same operator - Reproducibility Amount of variation observed in
the average of the measurements made by different
operators using the same unit and same equipment
20CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Sealer OQ protocol
- Process Characterization complete showing worst
case parameters - Process Challenges by generating fault conditions
- Power Failure
- Overload
- Gage RR Completed
- Any Discrepancies Resolved
21CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Performance Qualification (PQ)
- Demonstrate the process will consistently produce
acceptable product under normal operating
conditions - The PQ then challenges the process by performing
replicate runs at the defined worst case
acceptable condition - Multiple runs/lots captures lot to lot
variability including raw materials - Multiple operators captures operator to operator
variability - Multiple runs also captures any tear down, setup,
cleaning, and maintenance activities routinely
required
22CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Performance Qualification (PQ) cont
- Running at this worst case condition assures that
the product will meet all defined requirements
through all anticipated conditions of
manufacturing - Provides data for calculation of process
capability - Can the process consistently
produce product that meets specification
23CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Sealer PQ
- Replicate Runs/Lots Performed
- No Visual Seal Failures
- All Seals Burst Greater Than Spec
- Process Capability Determined to Be ? Cpk of 1.33
- Multiple Operators
- Any Discrepancies Resolved
24CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Final Validation Report Almost Done
- Summarize and reference all protocols and results
- Conclude the validation status of the process
- State routine monitoring parameters and range
- Criteria for revalidation
25CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Revalidation Requirements the following are
examples of conditions in which revalidation may
be necessary or at least evaluated - Changes in the actual process that may affect the
quality of the product - Negative trends in your quality indicators
- Change in product design affecting process
- Transfer of process to another facility
26CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Revalidation Requirements cont.
- Changes in external requirements (regulations,
standards) - New piece of equipment or modification of
existing equipment - Revalidate entire process?
- Not necessarily
- Dependent upon the extent of change
- New equipment may only require an IQ
- Document Decision
27CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Quality System Mechanism for Revalidation
- Change Order Process
- Are questions built into the CO process to force
a decision on revalidation, e.g. - Change to Equipment
- Change to process
- Change to Design
- Change to Supplier
- Nonconformance System
- Calibration/Preventive Maintenance Systems
- Corrective/Preventive Action System
28CASE STUDY Heat Sealer Validation for Sterile
Packaging
- Continuous Monitoring and Process Control
- Ensure the process remains within the established
parameters - Control Charts if applicable
- Use the data
- Process shifting, evaluate and adjust while it is
in range and before there is a need to revalidate - Nonconformance reports
- Investigate to determine if there is a need for
corrective action or revalidation
29Conclusions
- Dont be like this
- Engage the validation activities early
30Regulatory Requirements (21CFR 820.75)
- Validate It
- Control It
- Revalidate When Appropriate
- Document Throughout