Title: CTD CALIBRATION


1
CTD CALIBRATION
Paul Ridout
2
Why Calibrate?
OSIL,Ltd 2004
3
Why Calibrate ?
  • 2 2 ??

OSIL, Ltd 2004.
4
Accuracy and Precision
Precision Accuracy True
OSIL, Ltd 2004.
5
Basic Principle of Calibration
  • Primary Standard - national or international
  • Transfer Standard - recognised method
  • Instrument for Calibration - functional
  • (Ancillary equipment to maintain stable
    conditions)

OSIL, Ltd 2004.
6
Accuracy
  • WOCE requirements for CTD
  • Temperature 0.002 degrees C
  • Salinity 0.002 (PSS78)
  • Pressure 3dBar(0.05FS)

OSIL, Ltd 2004
7
CTD Calibration
Temperature
OSIL, Ltd 2004
8
Temperature Calibration
  • Primary Standard is triple point cell
  • Transfer standard is PRT with resistance
    bridge
  • Standard Resistor

OSIL,Ltd 2004
9
Measurements
  • Temperature is measured according to the
    International Temperature Scale1990 ( ITS90)
  • Oceanographic range Primary Standards
  • Mercury -38.8344
  • Water 0.010
  • Gallium 29.7646
  • No extrapolation

OSIL, 2004
10
Temperature Calibration
  • Fluid Bath
  • 2 X prt
  • Bridge
  • Std. Resistor
  • Temperature controlled laboratory

OSIL, 2004
11
Data Recording
Data Recording
  • PRT resistance ratio ( with Std. Resistor)
  • Nominal temperature of bath
  • Std. Resistor temperature
  • Laboratory temperature
  • Instrument raw data
  • Serial nos., date, analyst, comments, etc.

OSIL, 2004
12
Calculations
  • Temperature from resistance using
    calibrations from prt and std. Resistor with
    deviation functions for ITS90
  • Polynomial fit of instruments data to standard
    data

OSIL, 2004
13
CTD Calibration
Pressure
OSIL, 2004
14
Pressure Calibrations
  • Primary Standard comprises known masses and
    defined piston (NPL)
  • Transfer Standard DWT mass (or quartz sensor)
  • Dead weight tester (DWT)

OSIL, 2004
15
Measurements
  • Pressure using dead weight tester (and quartz
    standard)
  • Temperature
  • Atmospheric pressure
  • Relative heights pressure of standards and
    instruments

OSIL, 2004
16
Pressure Calibration
  • DWT used to increment pressure
  • Exercise sensor prior
  • Quartz sensor values
  • increasing/decreasing
  • Changes with temperature
  • Barometer
  • Height adjustments

OSIL, 2004
17
Data recording
  • Standard pressure
  • Atmospheric pressure
  • Height of instruments
  • Temperature
  • Instrument raw data
  • Serial nos., analyst, notes, etc

OSIL, 2004
18
Calculations
  • Adjustments for atmospheric pressure
  • Adjustments for height
  • Polynomial fit instrument data to standard data

OSIL, 2004
19
CTD Calibration
Conductivity
OSIL, 2004
20
Conductivity Calibration
  • Comparison of the measured value with a
    recognised, traceable standard
  • Salinity primary standard is KCl
  • Transfer standard is IAPSO Standard Seawater

www.seawatersolutions.com
OSIL, 2004
21
Measurements
  • Salinity - using salinometer to PSS78
  • Temperature - using prt according to ITS 90
  • Calculate IPTS68 temperature from ITS90
  • (1.00024 ITS90 IPTS68, Saunders 1990)
  • Calculate conductivity from the salinity
    (PSS78)
  • temperature (IPTS68)

OSIL, 2004
22
Conductivity by temperature change
  • Salinometer
  • Temperature controlled seawater bath(high to
    low)
  • Platinum thermometer
  • Salinometer pump
  • Lab Assistant software
  • IAPSO Std. Seawater

OSIL, 2004
23
Conductivity by salinity change
  • Seawater tanks of various salinities
  • Salinometer
  • Platinum thermometer
  • Salinometer pump
  • Lab Assistant software
  • IAPSO Std. seawater

OSIL, 2004
24
Data Recording
  • Salinometer salinity/conductivity ratio
  • PRT temperature of bath
  • Instrument raw data
  • Laboratory temperature
  • Atmospheric pressure
  • Serial numbers
  • Date, analyst, remarks

OSIL, 2004
25
Calculations
  • IPTS68 temperature from ITS90 of PRT
  • Conductivity (mS/cm) from salinity or
    conductivity ratio (Rt) and temperature
  • Use polynomial curve fitting program to compare
    standard readings with instruments data

OSIL, 2004
26
Post Calibration Checks
  • Temperature 2 points in range
  • Pressure 0, 10, 50, 100
  • Conductivity 2 points in range

OSIL, 2004
27
CTD Calibration
Paperwork
OSIL, 2004
28
Incoming Inventory
OSIL, 2004
29
Diagnostics
OSIL, 2004
30
Diagnostics (contd.)
OSIL, 2004
31
Service Report
OSIL, 2004
32
Datasheet
OSIL, 2004
33
Calibration Report
OSIL, 2004
34
Calibration Certificate
OSIL, 2004
35
Calibration Centre
OSIL, 2004