Title: Measurement of the Calorimetric Energy Scale in MINOS
1Measurement of the Calorimetric Energy Scale in
MINOS
- Jeff Hartnell
- Rutherford Appleton Laboratory
- (University of Oxford)
- Thesis Talk
- Week in the Woods 2005
2Talk Outline
- Motivation
- Track window technique
- Stopping muon selection
- Results
- Bethe-Bloch Measurements
- Sources of systematic error
- Performing the relative calibration in MC
- Impact on MINOS CC Measurements
- Summary
3Motivation
- Perform relative calibration
- Enable accurate spectral comparisons of beam
- Relative energy scale effects measurement of both
dm2 and s2t. - Tune the MC so that muons give correct response
at all detectors - MINOS relative calibration target 2
4What to use?
- Unfortunately no source of, say, exactly 1 GeV
electrons at all detectors (had them at CalDet!). - Average dE/dx of through-going cosmic muons
differs by 50 between 3 detectors. - Use stopping muons
- Well understood same at all detectors.
- Future cross-checks
- Use neutrino beam stopping muons.
- Use magnetic field to select, say, 20 GeV muons.
5Track Window Technique
- Sum energy deposition in a window on the
relativistic rise. - Robust measurement not reliant on perfect track
finding. - Away from sharp change in dE/dx as muon stops.
6Defining a MEU
- A Muon Energy Unit (MEU) is the same in all 3
detectors - As a rule of thumb
- 1 MEU equals the detector response to a
perpendicular 1 GeV muon traversing 1 plane of
scintillator - More technically
- Median of detector response in track window per
plane - Note only calibrate scintillator not steel!
- The first 190 planes in FD are 1 thicker than
the rest -- reconstruction must account for this - CalDet steel is 1.6 thinner than FD
7Stopping muon selection
- At CalDet use test-beam muons (1.4-2.0 GeV/c)
- Range cut to remove 98 of pions
- Fiducial volume to ensure containment
- Timing cuts to reject overlapping events
- At FarDet use stopping cosmics (80k data, 40k MC)
- Tight fiducial volume (Rlt3 m)
- 0.5 m away from coil hole
- Track quality checks (views agree, untrkd hits)
- Track projection to detector edge (traceZ)
- Signalbackground is 10001 in MC
- At NearDet (to be started soon)
8CalDet Results
- Measure response in 3 run periods
- T11 2002
- T7 2002
- T7 2003 (Near/Far)
9Cross-check with test-beam
Compare calibrated electron response per GeV
between run periods Detector response measured
with muons and electrons is consistent to better
than 2 Proves that the calibration at the CalDet
works!
10Far Detector Results
- Use same livetime as atmospheric analysis Aug 03
to Oct 04 - 80000 stopping muons
- Just one number!
- MEU 505 SigMaps
11Stopping muons as a Standard Candle
Asymmetric (longer clear fibres at top)
Uniform response (almost!)
Symmetric about centre of strips
12Spatial variations (zoom)
2 stripends x 2 plane views 4
combinations Purple lines show the edge of the
detector read out Clear correlation of
yellow/red patches with stripend read out 15
differences! Looks like problems with both
attenuation and strip-to-strip calibration MC is
perfectly flat!
13Overall Average Response
MC
Data
Y-dependence of detector response
Uniform response
5 fall in response
14Temporal fluctuations
4
Clear correlation of detector response with
temperature
No drift points used (not all available)
Drift 1.0 0.1 per degC
At CalDet gain drift 0.8 0.2 (LI)
15Effect of Spatial and Temporal Variations in
Detector Response
- If only a single neutrino event is considered
then errors are crudely - Temporal 2 error
- Spatial 2.5 error
- BUT, any reasonable analysis will use 100s or
1000s of events errors average out assuming a
uniform distribution - Temporal sqrt(1002)/100 0.2 error
- Spatial sqrt(1002.5)/100 0.2 error
- However, stopping muon sample is not uniformly
distributed (concentrated at top). - 0.6 correction and assume a 100 error on
correction
16Why should problems be fixed if small error?
- We dont understand our detector fully could be
a deeper problem? - MC simulation is inaccurate calibration
constants are used as the true light output of
each strip - Potentially susceptible to biases have to check
that events are uniform - Worse energy resolution
17Angular dependence
Only use events with cosThZ gt 0.3
Have a 1 dependence Average is 0.5 different
from horizontal correct Assign a 100 error to
size of correction (0.5)
18Measuring Bethe-Bloch Curve
- Measure muon energy deposition as a function of
distance from the end of the track - Convert
- Distance travelled -gt momentum
- Detector response -gt MeV cm2/g
- Provides valuable cross-check that stopping muon
reconstruction and calibration is correct
19FarDet Bethe-Bloch Curve
Good agreement with Bethe-Bloch and MC
Disagreement in data at high momenta
Data and MC slowly start to diverge
Muons all come from above!
Why?
20FarDet Bethe-Bloch Curve (2)
Make a correction for y-dependence
Now good agreement
21CalDet Bethe-Bloch Curve
MC agrees well with Bethe-Bloch calculation up to
here
1.8 GeV/c test-beam muons
Data AND MC disagree with Bethe-Bloch calculation
Energy loss ! energy deposited
Delta-rays can punch through more than 1 plane
Data disagrees with BOTH here
Really interesting dependence of BB curve on muon
range at CalDet.
(see next talk)
22Errors Statistical and Systematic
- Calibration detector
- Statistical lt 0.3
- No atten. correction 0.2
- Strip-to-strip calibration 1.2
- Temporal fluctuations 0.3
- Pion background 0.25
- Off-momentum muons 0.25
- Far detector
- Statistical 0.1
- Temporal drifts 0.2
- Spatial variations 0.6
- Angular dependence 0.5
23Total Errors
- Error on relative calorimetric energy scale
measurement - CalDet total 1.3
- FarDet total 1.1
- So, error on relative calibration of scintillator
between CalDet and FarDet is - sqrt(1.32 1.12) 1.6
- This is better than the 2 target
24Performing the Relative Calibration in MC
- Important cross-check of entire relative
calibration procedure - Shows self-consistency of MC calibration/de-calibr
ation process - Ensures that the light-level of each detector
in MC is an accurate simulation of reality
25Procedure
- PhotonTransport uses CALMIPCALIBRATION table to
set each detectors light level - Tune GeVPerMip until MC stopping muons give the
same response as data at ONE detector only - Now expect others detectors to give their correct
response do they?
26Results in MC
- FD and CD 02 all have the same deviation
(0.45) - Very good result
- Stopping muon reconstruction and calibration good
in both detectors - MC is self-consistent
- FD CD 02 differ from CD 03 by 1.6
- Due to sparsification of individual ND buckets
and cross-talk to unconnected pixels (electronics
pmt differences!) - Need to figure out a way to compensate for this
27Impact of Relative Calibration on MINOS CC
Measurements
- Fast MC loglikelihood analysis (by AW)
- Use MC truth smear muons and showers
- Use CC/NC selection efficiencies (by KG)
- Generate ideal spectrum using 32x1020 POT
sample - Introduce energy shift to shower in each event
for ideal spectrum - 11 points between -15 and 15 energy shift
- Use 500 data samples with only 8x1020 POT
statistics - Determine average fractional bias on best fit
value of dmsq and s2t
28Fractional bias on dmsq
Bias (fit true) / true Bias changes with the
size of the relative energy shift Size of bias
depends on dmsq
29Why is effect dependent on dmsq?
If you shift a flat energy spectrum it just
hardly matters!
30Results of Study
- A shift of 2 in the relative energy scale (for
SK best fit) means - 0.6 error on both dmsq and s2t
- For high values of dmsq (0.005) the effect of a
relative miscalibration is much smaller but at
the same time we make a 3x more precise
measurement of dmsq anyway.
31Summary
- Robust track window technique developed
- Precise measurements of the calorimetric energy
scale have been made - Relative calibration (of scintillator) between FD
and CD performed - to 1.6 in data
- to better than 2 in MC
- Calibration is below 2 target
- Important cross-checks shown to work
- Bethe-Bloch curve measured
- At CalDet, detector response consistent with
test-beam E - MC now tuned to give accurate realistic
calorimetric response and shown to be
self-consistent - Study shows a 2 relative miscalibration gives a
0.6 error on dmsq and s2t
32Now to pass my viva Thanks to everyone who
provided advice/software/MC/data/etc.