Title: Magnetic Resonance Imaging IV
1Magnetic Resonance Imaging IV
- Angiography
- Instrumentation
- Quality Assurance
2Magnetic Resonance Angiography
- Signal from blood dependent on relative
saturation of surrounding tissues and incoming
blood flow in the vasculature - In multislice volume, repeated excitation of
tissues and blood causes partial saturation of
the spins, depending on the T1 characteristics
and the TR of the pulse sequence - Blood outside of imaged volume does not interact
with RF excitations - Unsaturated spins enter imaged volume produce
large signal relative to blood within volume
(flow-related enhancement)
3(No Transcript)
4Flow presaturation
- In some cases, flow-related enhancement is
undesirable and may be eliminated using
presaturation pulses applied to volumes just
above and below the imaging volume - Same presaturation pulses helpful in reducing
motion artifacts caused by adjacent tissues
outside the imaging volume
5Flow-related signal loss
- Flow-related signal loss occurs when rapidly
flowing blood moves through the excited slab of
tissue, but does not experience the full
refocusing 180-degree pulse, resulting in a flow
void - Blood appears black in the image
6Time-of-flight angiography
- Technique relies on tagging of blood in one
region of the body and detecting it in another - Differentiates moving blood from surrounding
stationary tissues - Tagging accomplished by spin saturation,
inversion, or relaxation to change the
longitudinal magnetization of moving blood - Penetration of tagged blood into volume depends
on T1, velocity, and direction of blood
7Time-of-flight MRA (cont.)
- Detectable range limited by eventual saturation
of tagged blood - Long vessels difficult to visualize
simultaneously in a 3D volume - 2D stack of slices typically acquired, where even
slowly moving blood can penetrate region of RF
excitation in thin slices
8Time-of flight MRA (cont.)
- Each slice acquired separately
- Blood moving in one direction can be selected by
delivering presaturation pulse on adjacent slab
superior or inferior to slab of data acquisition - Thin slices helpful in preserving resolution of
flow pattern
9(No Transcript)
10MR angiography maximum intensity projection of
carotid arteries
11Phase contrast angiography
- Relies on phase change that occurs in moving
protons such as blood - One method uses a bipolar gradient
- One gradient with positive polarity followed by
second gradient with negative polarity, separated
by a delay time ?T - Second acquisition of same view of data (same
PEG) reverses polarity of bipolar gradients - Moving spins encoded with negative phase
- Stationary spins exhibit no phase change
12(No Transcript)
13Phase contrast MRA (cont.)
- Subtracting second excitation from first cancels
magnetization due to stationary spins but
enhances magnetization due to moving spins - Degree of phase shift directly related to
velocity encoding time, ?T, and the velocity of
the spins in the excited volume - Intensity variations are dependent on amount of
phase shift - Image is inherently quantitative and can provide
estimate of mean blood flow velocity and direction
14Magnitude and phase images provide contrast of
flowing blood. Magnitude images are sensitive to
flow, but not to direction phase images provide
direction and velocity information.
15Magnet
- Heart of the MR system
- Performance criteria include field strength,
temporal stability, and field homogeneity - Parameters affected by magnet design
- Air core magnets
- Wire-wrapped cylinders main field parallel to
long axis of cylinder - horizontal - Solid core magnets
- Permanent magnet or electromagnet main field
runs between poles of magnet - vertical
16Air core magnet (A) and solid core magnet (B)
17Resistive magnet
- Most of solid core design
- Use continuous electric power to produce magnetic
field - Field strengths from 0.1 T to about 0.3 T
- Turn off field in emergency
- High electricity costs relatively poor
uniformity/homogeneity of field - Open design enables claustrophobic patients to
tolerate examination - Fringe fields of main magnet better confined
18(No Transcript)
19Superconductive magnet
- Typically use air core electromagnet
configuration - High field strengths (0.3 T to 3.0 T clinical
systems 4.0 T to 7.0 T research systems) - High field uniformity
- High initial capital and siting costs, cryogen
costs, difficulty in turning off main magnet in
emergency, extensive fringe fields - Uncontrolled quenching can cause explosive
boiling of liquid He and severe damage to magnet
windings
20Cross-sectional view of a superconducting magnet
21(No Transcript)
22Ancillary equipment
- Shim coils (adjust main magnetic field improve
homogeneity in central volume) - Gradient coils
- RF coils (create B1 field detect transverse
magnetization) - Control interfaces, RF source, detector, and
amplifier, A-to-D converter, pulse programmer,
computer, gradient power supplies, image display
23Proximity head coil (A) and surface coil detector
(B)
24Head coil with built-in mirrors
25Surface coil in use for extremity imaging
26Block diagram of a typical MRI system
27Magnet siting
- Patients with pacemakers or ferromagnetic
aneurysm clips must avoid fringe fields above 0.5
mT image intensifiers, gamma cameras, and color
TVs severely impacted by fringe fields lt 0.3 mT - Administrative control for magnetic fringe fields
is 0.5 mT (5 G), requiring controlled access to
areas that exceed this level - Disruption of fringe fields can reduce
homogeneity of active imaging volume
28(No Transcript)
29(No Transcript)
30Quality control
- Components of MR system that must be periodically
checked - Magnetic field strength
- Magnetic field homogeneity
- System field shimming
- Gradient linearity
- System RF tuning
- Receiver coil optimization
- Environmental noise sources
- Power supplies
- Peripheral equipment
- Control systems
31A quality assurance phantom for MRI
32(No Transcript)
33Another quality assurance phantom for MRI
34(No Transcript)