Title: Stability of the electron energy
1Stability of the electron energy
- Recycler department meeting
- April 5th, 2006
- A. Shemyakin, M. Sutherland
2Motivation
- In calibrations of the electron energy by cooling
of pbars with RF off, we saw the energy shifting
by more than 1 keV at the same settings - Sometimes adjusting the Pelletron voltage seemed
to improve the cooling rate - Part of the problem is combination of a drift of
the charging current and a finite amplification
of the HV regulation loop
3Regulation circuitry
Generation voltmeter (GVM)
Pelletron HV regulation circuitry
The terminal voltage (GVMVLT), grid voltage
(GRIDV, and the needle current (TPSPRB) as
functions of the chain current. 4-Apr-2006. The
voltage changes as 0.15 kV/?A.
Needle current as a function of the grid voltage
4Dispersion based energy deviation measurement
- The idea is to use the high dispersion in U-bend
to trace the energy changes, while Autotunes keep
the beam positions constant in BPMS A05 and A07
U-bend
5BPM drifts
BXD08S
0.02 mm
- RMS of the BPM noise corresponds to 20 eV
BYD08S
36 hours
BPM readings of a calibration signal. 32.05 kHz
modulation, 1 Hz recording. Standard deviations
in 2 hours are 1.8 ?m for X and 3.4 ?m for Y.
6Calibration
BPM D08 reading as a function of the Pelletron
voltage. 14-Mar-06, Mary The Y derivative is
0.219 mm/kV and is consistent with the bending
radius. The X derivative is -0.001 mm/kV.
7Data from 3-hour run
- GVM readings and BYD08S, BXD08S (multiplied by
the calibration coefficient). 1 Hz data.
29-Mar-2006, Mary.
8Data with 10 sec averaging
The drift of the GVM value is 0.9 kV and is
consistent with the drop of the chain current by
6 ?A. The drift measured by BYD08S is 2.7 keV.
9Comparison of energy deviations estimated in 3
BPMs
1.5 kV
BXD07
BYD07
6 min
BYD08
Changes of the beam positions in three BPMs
multiplied by corresponding dispersion
coefficients and adjusted to begin from zero.
10Possible reasons for the discrepancy
- Drift of the GVM electronics
- Kermit is preparing a test of its stability
- Drift of the GVM rotation frequency
- Kermit may try to measure the frequency
- Changes in the distance between GVM and the
terminal - GVM measures the electric field, not the
potential - If the distance change, the measured voltage
deviates as 5.2 kV/mm - Axially symmetrical thermal effect is 7?10-6/K
- The distance stays the same within 0.05 mm from
opening to opening of the Pelletron tank - Non-axially symmetric motion is still not
excluded
11Conclusion
- The long-term stability of the Pelletron voltage
is poor, at least after turning the Pelletron on.
Kermit plans to determine stability of
electronics. - The procedure of measuring the energy deviation
by the beam position in a high- dispersion region
works to some extend. A beam motion upstream the
bend seems to be an issue. - One can consider implementing similar procedure
in the full line - First, make an absolute energy calibration with
pbars - Save e-beam settings as energy calibration file
- In normal operation, compare readings in BYR01S
with the calibration
12180 deg bend as spectrometer
Fix the beam positions in BPMs C90, Q01 with an
Autotune
Measure shifts in BYR01S