Title: Wavefront aberrations after Penetrating Keratoplasty
1Wavefront aberrations after Penetrating
Keratoplasty
- Adriana Forseto, MD Telma Pereira, MD Lucila
Pinto, MD Vera Mascaro, MD Walton Nosé, MD
Eye Clinic Day Hospital São Paulo - Brazil
2- The authors do not have any commercial or
proprietary interest in the subject matter of
this paper
3Purpose
- To analyze ocular order aberrations derived from
LADARWave (Alcon) and Zywave (Bausch Lomb)
devices in eyes after penetrating keratoplasty
(PKP) eyes and to compare the clinical refraction
with the lower order aberrations obtained from
these two aberrometers
4Methods
- Thirty post-PKP eyes from keratoconic patients
were evaluated using both the LADARWave and the
Zywave aberrometers, according to each respective
manufacturers guidelines
5Methods
- Ocular lower-order aberrations were evaluated for
3.5mm diameter pupil. The obtained values were
compared with both manifest and cycloplegic
refractions (clinical refraction) - To be eligible for this study all the eyes should
have no sutures
6Methods
- Total higher-order aberrations (HOA), 3rd HOA and
4th HOA were documented for a 6.00mm pupil size
and compared between the two aberrometers
7Methods
- For statistical analysis the manifest,
cycloplegic and aberrometers refractions were
also converted to power vectors coordinates where
M represents the spherical equivalent, and J0 and
J45 the cylinder components of the refraction
8Methods
- Statistical analysis
- Refractions Power vectors(1)
- M Spherical Equivalent
- J0 (-Cyl/2)cos(2axis)
- J45 (-Cyl/2)sin(2axis)
(1)Thibos LN, Wheeler W, Horner DG. Power
vectors an application of Fourier analysis to
the description and statistical analysis of
refractive error. Optom Vis Sci 199774367-375.
9Results
- The mean time between the keratoplasty and the
wavefront examinations was 5.924.29 years (range
1 to 15 years) - The Zywave aberrometer was unable to capture the
images in 3 eyes and the LADARWave in one of them
10Results
Cycloplegic refraction vs Aberrometer derived
refraction
Both aberrometers showed similar spherical
equivalent values when compared with the
cycloplegic refraction. Statistical significant
differences were noted with the Zywave sphere
measurements and with both aberrometers cylinder
measurements (plt0.05)
11Results
Manifest refraction vs Aberrometer derived
refraction
When compared with the manifest refraction just
LADARWave showed similar spherical equivalent
values. Statistical significant differences were
noted with both aberrometers sphere and cylinder
measurements (plt0.05)
12Results
- Both wavefront devices derived refractions were
highly correlated to clinical cycloplegic and
manifest M, J0 and J45 components as will be
shown in the next slides - However these correlations tended to be worst
regarding the cylinder power vector coordinates
(J0 and J45)
13Cycloplegic vs Wavefront (M)
- Pearson correlation
- LADARWave 0.990
- Zywave 0.989
14Cycloplegic vs Wavefront (J0)
- Pearson correlation
- LADARWave 0.961
- Zywave 0.934
15Cycloplegic vs Wavefront (J45)
- Pearson correlation
- LADARWave 0.904
- Zywave 0.935
16Manifest vs Wavefront (M)
- Pearson correlation
- LADARWave 0.963
- Zywave 0.959
17Manifest vs Wavefront (J0)
- Pearson correlation
- LADARWave 0.949
- Zywave 0.951
18Manifest vs Wavefront (J45)
- Pearson correlation
- LADARWave 0.917
- Zywave 0.915
19Results
Mean values of total HOA, 3rd and 4th HOA
(6.00mm pupil)
- No statistically significant discrepancies in HOA
measurements between the two Hartmann-Shack
devices were observed. - Trifoil aberrations were dominant when compared
to coma or spherical aberrations in post-PKP eyes.
20Conclusions
- This study demonstrated that both wavefront
devices derived refraction were highly correlated
to clinical measurements - Zywave showed more discrepancies when compared
with both cycloplegic and manifest refraction - These discrepancies should be taken into account
while performing customized ablation in post PKP
eyes
21Conclusions
- Technical problems preventing the centroids
acquisition were more frequent with the Zywave
system - Post PKP eyes had increased values of HOA,
specially trifoil - The mean values for all higher-order aberrations
components were similar between the two
aberrometers