Title: COAGULATION TESTS


1
COAGULATION TESTS
2
  • PT
  • PTT
  • Fibrinogen

3
Equipment
  • Water bath
  • Centrifuge
  • Reagents Buffers
  • Plastic glass tube
  • Pipettes
  • Stopwatches
  • Automated coagulation analyzers

4
PreAnalytical variables
  • Collection of venous blood
  • Exessive stress, Tourniquet, Needle
  • Anticoagulant
    Trisodium
    citrate(3.2 0.109 mol/L ) nine volume
    of blood added to one volume of anticoagulants
  • Under over filling of tubes
  • Severe anemia, polycytemia

5
PreAnalytical variables(cont.)
  • Centrifugation
    2000 g 15 min 4oC
  • Storage of sample
  • Time of sample collection
  • Icteric , Lipemic plasma-Hemolysis-Clot formation

6
Analytical variables
  • Sample , Delay in sample processing
  • Equipment
  • Reagents )thromboplastin , Ca )
  • Contamination of the reagents
  • Test method

7
Quality Control
  • Control plasma(normal abnormal)
  • Pooled plasma
  • EQAS

8
Prothrombin Time (PT)
9
Prothrombin Time (PT)
  • Samples should be stored in room temprature
  • Reagent is Thromboplastine
  • Expression of results
  • Normal Reference Range

10
Thromboplastine
  • Tissue factor Apoprotein,Phospholipid
  • Source of Tissue Factor
  • rabbit ,human ,monkey brain
  • rabbit lung, human placenta
  • Recombinant

11
  • International Normalized Ratio (INR) system was
    introduced by the WHO in 1983
  • International Sensitivity Index (ISI)
  • patient PT
  • INR( )
  • mean of normal PT

ISI
12
Therapeutic Range
13
ISI
  • The lower the ISI value the more sensitive the
    reagent.(0.9-1.7)
  • Use of reagents with lower ISI,s improve the
    precision of INR.

14

(PT increasing from 23 secs to 25 secs will
increase the INR 23
1 25 1
ISI 1 ( )2.1
( ) 2.4
10.5 10.5 ISI 2
23 2 25
2 ( )4.8
( )5.7
10.5
10.5
15
  • a variation in the time of clot detection will
    result in amplification in INR calculation
    by higher ISI value.
  • Loss of accuracy and precision with insensitive
    Thromboplastin (ISIgt1.5)
  • Insensitive Thromboplastin caused low PT ratio

16
  • Thromboplastine reagents with low ISI will result
    in a longer PT.

17
Choice of Thromboplastine
  • The lower the ISI value the more sensitive the
    reagent.
  • Use of reagents with lower ISI,s improve the
    precision and accuracy of INR.
  • Use of Thromboplastine with ISI
  • 0.9-1.7 is highly recommended

18
Activated Partial ThrombineTime (PTT)
  • Sample Storage
  • Reagent
    Contact factor,Ca
  • Method
  • Ca 0.025 mol/l
  • Normal Reference Range
  • Expression of results

19
Fibrinogen Assay
  • Reagents
    Fibrinogen Standard
  • Thrombin solution
  • Owren Buffer
  • Calibration curve

20
Coagulation Analyzer
  • Electromechanical
  • Photooptical
  • Electrochemical

21
  • ?? ????