1
Implanterbare høreapparater 10 februar
2005 Sten Harris
2
Sten Harris ? Ørekirurg Audiolog Otoaudiologisk
seksjon RH
3
Implanterbare høreapparater ? Semi-implantable
hearing aids Totally implantable hearing aids (
BAHA, CI)
4
Varför implantera hörapp? Kombination av
problem/missnöje med konventionell HA och nya
tekniska möjligheter
5
Problem konventionell HA
Begränsad frekvensgång Dålig ljudkvalitet Hörappar
atproppen ocklusion,infektion,allergi Tål ej
vatten Kosmetik/fåfänga
6
mikr
HA
hörtelefon
förstärkare
Implanterb HA mekanisk stimulering
7
Implantabla hörapparater?
  • God ljudkvalitet
  • Bred frekvensgång
  • Mindre återkoppling

8
Semi- implantabla hörapparater
9
Historik
  • Piezoelektriska system
  • Vernon 1972
  • Suzuki-Yanagihara

10
Piezoelektriskt System RION
11
Historik
  • Elektromagnetiska system
  • Fredrickson 1973

12
Elektromagnetiskt system
Richards
13
Elektromagnetiskt system Soundtec

14
Semi-implantabla HA idag
Vibrant--MED-EL MET--Otologics
15
symphonix
Vibrant Soundbridge
MED-EL
Audio Processor
Receiver (VORP)
Conductor lead
FMT Floating Mass Transducer
16
External Part Audio Processor
The external part contains the microphone,
battery, a Digital signal processor (Signia /
Siemens) and a RF modulator. 2.5 Cm in
diameter. Held in place over the implant with
a magnet.
3 Colors Brown, Beige, Grey
17
Audio Processor in Place
  • no occlusion (ear canal free)
  • easily hidden (covered by hair)
  • Held by magnetic attraction

18
Vibrating Ossicular Prosthesis (VORP)
Demodulator
Magnet
Receiver coil
Silicone
Floating Mass Transducer
19
FMT Floating Mass Transducer
The Floating Mass TM is attached to the long
process of the incus with a Titanium clip
20
Position of the FMT
Left Ear
The FMT should be in close contact and parallel
to the stapes
21
Vibrant Soundbridge Middle Ear Implant
April 2003
22
Geoff Ball
The inventor of the Vibrant Soundbridge is
himself implanted (bilaterally)
23
Vibrant Soundbridge Selection Criteria
  • Word recognition score of 50 or better - to
    assess the cochlear integrity
  • Normal middle ear anatomy and sensorineural
    hearing loss - so as not to impede the energy
    transfer or the ossicular movement
  • Realistic expectations of the benefits and
    limitations of the device

24
MET Ossicular Stimulator
OTOLOGICS
  • Electromechanical transducer
  • Direct coupling to incus by advancing transducer
    probe tip into laser made hole within body of
    incus
  • Fibrous union formed during healing period
  • Peak Output 118 dB, Frequency range lt200 to
    6250 Hz
  • No disarticulation of Ossicles
  • Nothing in ear canal

25
Otologics Semi-Implantable Technology

MET Ossicular Stimulator

Button Audio Processor
26
Incision
27
Atticotomy
28
Fastening the Mounting Bracket
29
Transducer Placement
30
Securing the Electronics
31
Patient Selection Criteria
  • Adults with sensorineural (cochlear) hearing loss
  • Post-lingual onset of hearing loss
  • No signs or symptoms of conductive, mixed,
    retrocochlear, central auditory, or functional
    hearing loss
  • Stable hearing loss
  • No medical contraindications to surgery or middle
    ear implantation
  • Realistic expectations of device benefit

32
Thresholds
Bone Conduction
Air Conduction
N112
33
  • RH-erfarenheter- Otologics
  • Kvinna f 57
  • Sensorineural hns 40 dB
  • Missnöjd HA ssk ocklusionseffekt
  • Mycket nöjd efter op-
  • men tekniskt breakdown efter ett år
  • ej bättre efter reop

34
RH-erfarenheter- Otologics 2. Man f 49 komb
hns- cochleär hudsjukdom. PTA 30/50 Klarar ej
HA el BAHA. Efter en hel del programmerings- beky
mmer nu en mycket belåten bärare Åter i yrkeslivet
35
Pas 3.
Kvinna f.1962 Progressiv hns Kan ej ha propp Op
aug 03 vä Önskar op hö
36
Helt implanterade hörapparater
37
Implex det første helt implanterte hørepparatet
38
OTOLOGICS
39
Implant
Microphone
Digital Signal Processor
Battery
IS-1 Connector
Lead
Receiver Coil
Magnet
Transducer
40
Upgrade Surgical Procedure
Semi-Implantable
Fully-Implantable
41
MET Fully-Implantable System
Implant
Remote
Charging System
42
Charging
  • Implant Charger
  • Recharges battery of implant
  • Daily recharge takes less than 1 hour
  • Patient can use implant during re-charge
  • Base Station
  • Charges implant charger
  • Implant Battery
  • Warranted for 5 years
  • Life gt 12 years

43
Remote
  • Volume control
  • On/Off

44
Programming System
  • NOAHlink System
  • Fitting System
  • Diagnostic System

45
FIMOS FDA Clinical Trial
  • Implants to start Q1 2005
  • US
  • IDE Approval to begin recruiting patients
  • 20 patients from up to 15 centers
  • EU
  • 25 Patients from 7-10 sites

46
Tack för uppmärksamheten