Title: Chlorine Residual


1
Ana-Lab Corp.
P.O. Box 9000 2600 Dudley Road Kilgore, TX 75663
Employee Training
2
Presents
3
Field pH
  • By
  • Roy White

4
Field pH
  • Read and follow SOP reference materials
  • Follow safety plan
  • Collect representative aliquot
  • Protect integrity of sample
  • Record results and conditions
  • Complete Chain of Custody

5
Each Participant shall be able to
  • Turn ph meter on
  • Calibrate meter
  • Prepare sample
  • Measure pH
  • Read temperature
  • Record data

6
VWR SP20 pH Meter
7
VWR SP20 pH Meter
1. Switch unit on using power switch
8
VWR SP20 pH Meter
  • Meter Calibration
  • 1. Determine preliminary pH of sample to decide
    which buffers to use for two point calibration.
  • If pH lt 7.00 select 7.00 4.00
  • or
  • If pH gt 7.00 select 7.00 10.00

9
VWR SP20 pH Meter
  • Calibrate meter in a contamination free
    environment. Avoid direct sunlight, rain, and
    blowing dust.
  • Remove electrode from storage solution
  • Rinse with distilled water and blot
  • Place electrode in the 7.00 buffer

10
VWR SP20 pH Meter
  • When ready appears in display, Press OK
  • P2 will appear in display

OK
11
VWR SP20 pH Meter
  • Remove electrode from buffer 7.00
  • Rinse with distilled water and blot
  • Place electrode in second buffer (determined in
    step 1)
  • When ready appears in display, Press OK

OK
12
VWR SP20 pH Meter
  • P3 will be displayed in display
  • Press mode to complete two point calibration

ooo mode
13
VWR SP20 pH Meter
  • 16.Select buffer 6.00 or 8.00 to use as a
  • Check Standard (PreCS - should be between
    the calibration buffers)
  • 17. Read and record (within /- 0.05)
  • 18.Read and record buffer 7.00 (within
  • /- 0.05)

14
VWR SP20 pH Meter
  • 19. Read and record second buffer (within 0.05)
  • 20. When all three buffers read to within 0.05 of
    their value, calibration is complete
  • 21. Meter can be used to measure pH
  • 22. Check Record mid-point after each measure
    (Post CS)

15
VWR SP20 pH Meter
  • Record Calibration
  • pHlt7.00
  • pH /7.00 /4.00 PreCS
    /6.00 PostCS /6.00
  • pHgt7.00
  • pH /7.00 /10.00 PreCS
    /8.00 PostCS /8.00


16
VWR SP20 pH Meter
  • Record Calibration
  • pHlt7.00
  • pH 7.01 /7.00 4.02 /4.00 PreCS
    5.95 /6.00 PostCS 5.97 /6.00
  • pHgt7.00
  • pH 7.01 /7.00 10.02 /10.00 PreCS 8.03
    /8.00 PostCS 8.01 /8.00


17
VWR SP20 pH Meter
  • Measuring pH

18
VWR SP20 pH Meter
  • Collect sample
  • Analyze sample within 15 minutes of collection
  • Rinse the electrode with distilled water and blot
  • Pour sample in container to a depth that covers
    sensing elements of the electrode
  • Place the electrode in sample and stir gently

19
VWR SP20 pH Meter
  • Allow meter to stabilize ( ready )
  • Read pH
  • Record pH to nearest tenth (0.1)
  • Record temperature to nearest degree
  • Rinse electrode and return to storage bottle

20
VWR SP20 pH Meter
  • Record pH to nearest 0.1 SU
  • Record temperature to nearest degree
  • Record Name, Date Times
  • pH (On Site) Collected by Roy
    White Date 3/3/03 Time 0912
  • Results____8.3____Temp __28__C Analyzed by Roy
    White Date 3/3/03 Time 0917
  • Duplicate ______ Temp _______C

21
VWR SP20 pH Meter
22
VWR SP20 pH Meter
Questions?
Questions?
Questions?
Questions?
23
VWR SP20 pH Meter
Let's Roll