Title: Introduction Heinz-Dieter Nuhn, SLAC / LCLS November 14, 2005
1IntroductionHeinz-Dieter Nuhn, SLAC /
LCLSNovember 14, 2005
- Need for Beam Based Undulator K Measurements
- Review of Beam Based K Measurement Discussions
- LCLS Undulator Diagnostics Baseline Components
- LCLS FEL Commissioning Milestones
- Workshop Objective and Agenda
- Charge to the Workshop
2Linac Coherent Light Source
3(No Transcript)
4Summary of Nominal Undulator Parameters
- Undulator Type planar hybrid
- Magnet Material NdFeB
- Wiggle Plane horizontal
- Gap 6.8 mm
- Period Length 30.0 0.05 mm
- Effective On-Axis Field 1.249 T
- Standard Effective K 3.500 0.015
- Range of Effective Undulator Parameter K 3.500 -
3.493 (3.480) - Accumulated Segment Phase Error Tolerance
10 degrees(at any point along segment) - Module Length 3.40 m
- Number of Modules 33
- Undulator Magnet Length 112.2 m
- Standard Break Lengths 48.2 - 48.2 - 94.9 cm
- Nominal Total Device Length 130.954 m
- Quadrupole Magnet Technology EMQ
- Nominal Quadrupole Magnet Length 7 cm
5Undulator Segment Prototype
6Undulator Pole Canting
- Canting comes from wedged spacers
- 4.5 mrad cant angle
- Gap can be adjusted by lateral displacement of
wedges - 1 mm shift means 4.5 microns in gap, or 8.2 Gauss
- Beff adjusted to desired value
7Using Undulator Roll-Away and K Adjustment
Function
- Horizontal position of undulator segment can be
remotely controlled correct Keff on beam axis - This adjustment range goes from fraction of a
percent to a complete field turn-off.
Neutral K3.5000 Dx0.0 mm
PowerTp K3.4804 Dx8.5 mm
Beam Axis
SpontTp K3.4929 Dx3.0 mm
RollAway K0.0000 Dx100 mm
8Measurement of Spontaneous Radiation Using Rollout
Undulator Segments can be removed by remote
control from the end of the undulator. They will
not effect radiation produced by earlier segments.
9Insufficient Knowledge of Actual K Seen by
Electrons
- Effects Influencing Keff
- Undulator Segment Tuning
- Undulator Temperature
- Transverse Segment Position
- Segment Fiducialization and Alignment
- Electron Beam Trajectory
- Environmental Field in Undulator Hall
- Radiation Damage
See Tolerance Budget on next Slide
Need for Beam Based Undulator K Measurements
10Segment Detuning Sub-Budget
Parameter pi Typical Value rms dev. dpi Note
KMMF 3.5 0.0003 0.015 uniform
aK -0.0019 C-1 0.0001 C-1 Thermal Coefficient
DT 0 C 0.32 C 0.56 C uniform without compensation
bK 0.0023 mm-1 0.00004 mm-1 Canting Coefficient
Dx 1.5 mm 0.05 mm Horizontal Positioning
11Discussions of Beam Based K Measurements Based on
Spontaneous Undulator Radiation
- January 2004 Z. Huang
- Sven Reiche
- September 2004 LCLS Diagnostics and
Commissioning Workshop - High-Resolution Effective K Measurements Using
Spontaneous Undulator Radiation, Bingxin Yang
http//www-ssrl.slac.stanford.edu/lcls/workshops/
2004-09-22_diag_comm/bxyang_CommWorkshop200409.ppt
- October 2004 LCLS Week
- Undulator / FEL Diagnostics, Bingxin Yang
https//www-ssrl.slac.stanford.edu/lcls/fac/talks
_oct2004/Yang_FAC200410.ppt - January 2005 LCLS FEL Physics Meeting
- Simulation Results for 200-pC ("chargito") SASE
performance with AC Wake, Jim Welch
http//www-ssrl.slac.stanford.edu/lcls/internals/
felphysics/2005-01-18/k_meas_talk.ppt - April 2005 ICFA Commissioning Workshop at Zeuthen
(Work Package 6) - Measurement of Undulator Segment K_effective
using Spontaneous Radiation in the Near Hall of
the LCLS, Jim Welch http//adweb.desy.de/mpy/ICFA2
005_Commissioning/Talks(PDF)/April202120(Thursda
y)/WP6_1/Welch_Undulator20Commissioning.pdf - High resolution undulator measurements using
angle-integrated spontaneous spectra, Bingxin
Yang http//adweb.desy.de/mpy/ICFA2005_Commission
ing/Talks(PDF)/April202120(Thursday)/WP6_2/Yang_
High20Resolution20Undulator20measurements.pdf - July 2005 LCLS Week
- K-Measurement Strategies discussion presented by
Jim Welch and Bingxin Yang - October 2005 FAC Meeting
- X-Ray Diagnostic, Richard Bionta
http//ssrl.slac.stanford.edu/lcls/fac/talks_oct_
2005/bionta_xtod_diagnostics_fac.ppt
12Review of Existing LCLS Baseline Diagnostics
- Diagnostics presently being developed to
Characterize Electron Beam and X-Ray Properties
include - Electron Beam Diagnostics in the
Linac-To-Undulator (LTU) Beamline - Electron Beam and X-Ray Diagnostics in the
Undulator - Electron Beam Diagnostics after the Undulator
(Dump Line) - X-Ray Diagnostics in the Front End Enclosure (FEE)
13LTU Electron Beam Diagnostics
- Control of Electron Beam Properties before
Entrance into the Undulator
14LTU
abort dump
1 OTR ? slice-(e, b, g)
OTR ? slice E-spread (0.02)
4 wires ? e, b, g ( collimators)
2 BPMs ? energy jitter
x1
x2
relative energy centroid resolution 0.003 (5-mm
BPMs)
Courtesy of Paul Emma
15X-Ray Beam Diagnostics
- Control of Electron Beam Trajectory inside the
Undulator
16Short Break Section Components
Beam Finder Wire
RF Cavity BPM
Quadrupole
Undulator Segment
Cherenkov Detector
Undulator Segment
Courtesy of Dean Walters
17Long Break Section Components
Beam Finder Wire
Diagnostics Tank
RF Cavity BPM
Quadrupole
Undulator Segment
Cherenkov Detector
Undulator Segment
Courtesy of Dean Walters
18After Undulator Electron Beam Diagnostics
- Electron Beam Diagnostics after Undulator
19Dump-Line
1 OTR ? energy-spread (0.001)
1 BPM ? energy-jitter (0.003)
Courtesy of Paul Emma
20Diagnostics in the Front End Enclosure
- Measurement of X-Ray Beam Properties in FEE
21FEE Layout
Slit
Be Mirrors 2 3
Diagnostics Package
Solid Attenuator
Collimator 1
Fast close valve
Gas Attenuator
Ion Chamber
SiC Mirror 1
Ion Chamber
Diagnostics Package
SiC Mirror 2
Courtesy of Richard Bionta
22Desired Spontaneous Measurements
f(x,y,l1) Spatial distribution around l1
l1 1st harmonic Photon wavelength
Dl/l1 1st harmonic wavelength spread
Beam direction
u Total energy / pulse
su,sl1 Temporal variation in beam parameters
Courtesy of Richard Bionta
23FEE Cartoon
Start of Experimental Hutches
5 mm diameter collimators
Windowless Ion Chamber
Diagnostic Package
Spectrometer / Indirect Imager mirror
Solid Attenuator
High-Energy Slit
Total Energy Calorimeter
e-
WFOV Direct Imager
Gas Attenuator
FEL Offset mirror system
Spectrometer camera
Windowless Ion Chamber
Muon Shield
Courtesy of Richard Bionta
24Redundant Commissioning Instrumentation
Instrument Purpose Adjustment Calibration and Physics risks
Direct Imager SP f(x,y), look for FEL, measure FEL u, f(x,y), x,y ND filter, Attenuators Scintillator linearity, Attenuator linearity and background
Indirect Imager Measure FEL u, f(x,y), spectral imaging of SP and FEL harmonics, attenuator calibration Mirror Angle Mirror reflectivity, damage
Total Energy FEL u Attenuators Energy to Heat, damage
Ion Chamber FEL u, x,y,x',y' Pressure Signal strength
Spectrometers FEL, SP spectra Attenuators Resolution, damage
Courtesy of Richard Bionta
25LCLS FEL Commissioning Milestones
- MS3BO_040 Front End Beneficial Occupancy
(9/5/2007) - MS3BO_030 Undulator Facility Beneficial
Occupancy (12/3/2007) - MS3_XT040 Solid Attenuator Installation Complete
(12/14/2007) - MS3_XT045 Gas Attenuator Installation Complete
(12/14/2007) - MS3_XT080 Start Front End Enclosure
Commissioning (3/4/2008) - MS3_LN015 Start Linac-to-Undulator (LTU)
Commissioning (5/12/2008) - MS3_XT066 Start Near Experimental Hall Checkout
(6/12/2008) - MS3_UN020 Undulator System Installation Complete
(7/18/2008) - MS3_UN025 Start Undulator Commissioning (1st
Light) (7/24/2008)
Diagnostics needed around July 2008
26Workshop Objective
- Define a strategy for using spontaneous undulator
radiation to measure the K value of every
individual LCLS Undulator Segment after
installation in the Undulator Hall. - To reach the objective, the physics and
technologies necessary need to be identified.
Workshop discussions will include - Usable spectral features of spontaneous radiation
- Strategies for beam-based K measurements
- Specifications for suitable instruments
- Scheduling issues
- Three Work Packages have been defined and
assigned to three different groups. Work
described by these Work Packages has been carried
out in preparation of the workshop and will be
presented and discussed at the workshop.
27Workshop Agenda
28Work Package 1 Angle Integrated Measurement
- Group B. Yang, R. Dejus
- Task Examine robustness of angle-integrated
measurements of undulator spectrum. Consider
effects of errors in beam alignment, undulator
magnet structure, straightness of vacuum pipe,
alignment of spectrometer, etc. Consider effects
of location of undulator segment being tested.
Determine what are realistic values for the
precision with which the value of K can be
determined for an undulator segment at the
beginning, middle, and end of the undulator.
This task explores the use of the high-energy
edge of the fundamental spectral peak (the third
harmonic may also be considered) of a single
undulator to measure its K parameter. The
measuring spectrometer will be located in the
LCLS FEE, roughly 100 m downstream from the final
undulator segment. Realistic values for the
angular acceptance of the measurement (limited by
beam-pipe apertures, or apertures at the
measuring point) should be considered.
29Work Package 2 Pinhole Measurement
- Group J. Welch, R. Bionta, S. Reiche
- Task Examine robustness of pinhole measurements
of undulator spectrum. Consider effects of errors
in beam alignment, undulator magnet structure,
straightness of vacuum pipe, alignment of pinhole
and spectrometer, etc. Consider effects of
location of undulator segment being tested.
Determine what are realistic values for the
precision with which the value of K can be
determined for an undulator segment at the
beginning, middle, and end of the undulator.
This task explores the use of the fundamental
spectral peak (the third harmonic may also be
considered) of a single undulator, as seen
through a small angular aperture, to measure its
K parameter. The measuring spectrometer will be
located in the LCLS FEE, roughly 100 m downstream
from the final undulator segment. Realistic
values for the angular acceptance of the
measurement should be determined, and the effects
of misalignment of the aperture or undulator axis
should be carefully considered.
30Work Package 3 Single-Shot Spectral Measurement
- Group J. Hastings, et al.
- Task Assume that a single shot spectral
measurement is needed for an LCLS spontaneous
undulator pulse. What are the best options for
doing the measurement? What spectral resolution
can be obtained using these methods? What are the
effects of beam jitter, spectrometer
misalignment, etc? This task explores the
design and performance of x-ray spectrometers
capable of providing centroid or edge position
with high resolution, on a single-shot of
radiation from a single LCLS undulator. The
spectrometer will most likely be located in the
LCLS FEE, about 100 m downstream from the final
undulator segment.
31Workshop Charge
- Characterize the spectral features of spontaneous
synchrotron radiation that are usable for
beam-based K-measurements. - Identify the most appropriate strategy for
beam-based K-measurements. - Specify suitable instruments for the identified
beam-based K-measurement strategy. - List expected performance parameters such as
resolution of K measurement as function of beam
charge, and segment location as well as expected
tolerances to trajectory and energy jitter. - List any open questions regarding the feasibility
of the most appropriate strategy. - List the RD activities, if any, needed before
the design of a measurement system can be
completed and manufacturing/procurement can start.
32End of Presentation