Introduction Heinz-Dieter Nuhn, SLAC / LCLS November 14, 2005 - PowerPoint PPT Presentation

About This Presentation
Title:

Introduction Heinz-Dieter Nuhn, SLAC / LCLS November 14, 2005

Description:

... for 200-pC ('chargito') SASE performance with AC Wake, Jim Welch ... Measure FEL u, f(x,y), spectral imaging of SP and FEL harmonics, attenuator calibration ... – PowerPoint PPT presentation

Number of Views:31
Avg rating:3.0/5.0
Slides: 33
Provided by: AMUE5
Category:

less

Transcript and Presenter's Notes

Title: Introduction Heinz-Dieter Nuhn, SLAC / LCLS November 14, 2005


1
IntroductionHeinz-Dieter Nuhn, SLAC /
LCLSNovember 14, 2005
  • Need for Beam Based Undulator K Measurements
  • Review of Beam Based K Measurement Discussions
  • LCLS Undulator Diagnostics Baseline Components
  • LCLS FEL Commissioning Milestones
  • Workshop Objective and Agenda
  • Charge to the Workshop

2
Linac Coherent Light Source
3
(No Transcript)
4
Summary of Nominal Undulator Parameters
  • Undulator Type planar hybrid
  • Magnet Material NdFeB
  • Wiggle Plane horizontal
  • Gap 6.8 mm
  • Period Length 30.0 0.05 mm
  • Effective On-Axis Field 1.249 T
  • Standard Effective K 3.500 0.015
  • Range of Effective Undulator Parameter K 3.500 -
    3.493 (3.480)
  • Accumulated Segment Phase Error Tolerance
    10 degrees(at any point along segment)
  • Module Length 3.40 m
  • Number of Modules 33
  • Undulator Magnet Length 112.2 m
  • Standard Break Lengths 48.2 - 48.2 - 94.9 cm
  • Nominal Total Device Length 130.954 m
  • Quadrupole Magnet Technology EMQ
  • Nominal Quadrupole Magnet Length 7 cm

5
Undulator Segment Prototype
6
Undulator Pole Canting
  • Canting comes from wedged spacers
  • 4.5 mrad cant angle
  • Gap can be adjusted by lateral displacement of
    wedges
  • 1 mm shift means 4.5 microns in gap, or 8.2 Gauss
  • Beff adjusted to desired value

7
Using Undulator Roll-Away and K Adjustment
Function
  • Horizontal position of undulator segment can be
    remotely controlled correct Keff on beam axis
  • This adjustment range goes from fraction of a
    percent to a complete field turn-off.

Neutral K3.5000 Dx0.0 mm
PowerTp K3.4804 Dx8.5 mm
Beam Axis
SpontTp K3.4929 Dx3.0 mm
RollAway K0.0000 Dx100 mm
8
Measurement of Spontaneous Radiation Using Rollout
Undulator Segments can be removed by remote
control from the end of the undulator. They will
not effect radiation produced by earlier segments.
9
Insufficient Knowledge of Actual K Seen by
Electrons
  • Effects Influencing Keff
  • Undulator Segment Tuning
  • Undulator Temperature
  • Transverse Segment Position
  • Segment Fiducialization and Alignment
  • Electron Beam Trajectory
  • Environmental Field in Undulator Hall
  • Radiation Damage

See Tolerance Budget on next Slide
Need for Beam Based Undulator K Measurements
10
Segment Detuning Sub-Budget
Parameter pi Typical Value rms dev. dpi Note
KMMF 3.5 0.0003 0.015 uniform
aK -0.0019 C-1 0.0001 C-1 Thermal Coefficient
DT 0 C 0.32 C 0.56 C uniform without compensation
bK 0.0023 mm-1 0.00004 mm-1 Canting Coefficient
Dx 1.5 mm 0.05 mm Horizontal Positioning
11
Discussions of Beam Based K Measurements Based on
Spontaneous Undulator Radiation
  • January 2004 Z. Huang
  • Sven Reiche
  • September 2004 LCLS Diagnostics and
    Commissioning Workshop
  • High-Resolution Effective K Measurements Using
    Spontaneous Undulator Radiation, Bingxin Yang
    http//www-ssrl.slac.stanford.edu/lcls/workshops/
    2004-09-22_diag_comm/bxyang_CommWorkshop200409.ppt
  • October 2004 LCLS Week
  • Undulator / FEL Diagnostics, Bingxin Yang
    https//www-ssrl.slac.stanford.edu/lcls/fac/talks
    _oct2004/Yang_FAC200410.ppt
  • January 2005 LCLS FEL Physics Meeting
  • Simulation Results for 200-pC ("chargito") SASE
    performance with AC Wake, Jim Welch
    http//www-ssrl.slac.stanford.edu/lcls/internals/
    felphysics/2005-01-18/k_meas_talk.ppt
  • April 2005 ICFA Commissioning Workshop at Zeuthen
    (Work Package 6)
  • Measurement of Undulator Segment K_effective
    using Spontaneous Radiation in the Near Hall of
    the LCLS, Jim Welch http//adweb.desy.de/mpy/ICFA2
    005_Commissioning/Talks(PDF)/April202120(Thursda
    y)/WP6_1/Welch_Undulator20Commissioning.pdf
  • High resolution undulator measurements using
    angle-integrated spontaneous spectra, Bingxin
    Yang http//adweb.desy.de/mpy/ICFA2005_Commission
    ing/Talks(PDF)/April202120(Thursday)/WP6_2/Yang_
    High20Resolution20Undulator20measurements.pdf
  • July 2005 LCLS Week
  • K-Measurement Strategies discussion presented by
    Jim Welch and Bingxin Yang
  • October 2005 FAC Meeting
  • X-Ray Diagnostic, Richard Bionta
    http//ssrl.slac.stanford.edu/lcls/fac/talks_oct_
    2005/bionta_xtod_diagnostics_fac.ppt

12
Review of Existing LCLS Baseline Diagnostics
  • Diagnostics presently being developed to
    Characterize Electron Beam and X-Ray Properties
    include
  • Electron Beam Diagnostics in the
    Linac-To-Undulator (LTU) Beamline
  • Electron Beam and X-Ray Diagnostics in the
    Undulator
  • Electron Beam Diagnostics after the Undulator
    (Dump Line)
  • X-Ray Diagnostics in the Front End Enclosure (FEE)

13
LTU Electron Beam Diagnostics
  • Control of Electron Beam Properties before
    Entrance into the Undulator

14
LTU
abort dump
1 OTR ? slice-(e, b, g)
OTR ? slice E-spread (0.02)
4 wires ? e, b, g ( collimators)
2 BPMs ? energy jitter
x1
x2
relative energy centroid resolution 0.003 (5-mm
BPMs)
Courtesy of Paul Emma
15
X-Ray Beam Diagnostics
  • Control of Electron Beam Trajectory inside the
    Undulator

16
Short Break Section Components
Beam Finder Wire
RF Cavity BPM
Quadrupole
Undulator Segment
Cherenkov Detector
Undulator Segment
Courtesy of Dean Walters
17
Long Break Section Components
Beam Finder Wire
Diagnostics Tank
RF Cavity BPM
Quadrupole
Undulator Segment
Cherenkov Detector
Undulator Segment
Courtesy of Dean Walters
18
After Undulator Electron Beam Diagnostics
  • Electron Beam Diagnostics after Undulator

19
Dump-Line
1 OTR ? energy-spread (0.001)
1 BPM ? energy-jitter (0.003)
Courtesy of Paul Emma
20
Diagnostics in the Front End Enclosure
  • Measurement of X-Ray Beam Properties in FEE

21
FEE Layout
Slit
Be Mirrors 2 3
Diagnostics Package
Solid Attenuator
Collimator 1
Fast close valve
Gas Attenuator
Ion Chamber
SiC Mirror 1
Ion Chamber
Diagnostics Package
SiC Mirror 2
Courtesy of Richard Bionta
22
Desired Spontaneous Measurements
f(x,y,l1) Spatial distribution around l1
l1 1st harmonic Photon wavelength
Dl/l1 1st harmonic wavelength spread
Beam direction
u Total energy / pulse
su,sl1 Temporal variation in beam parameters
Courtesy of Richard Bionta
23
FEE Cartoon
Start of Experimental Hutches
5 mm diameter collimators
Windowless Ion Chamber
Diagnostic Package
Spectrometer / Indirect Imager mirror
Solid Attenuator
High-Energy Slit
Total Energy Calorimeter
e-
WFOV Direct Imager
Gas Attenuator
FEL Offset mirror system
Spectrometer camera
Windowless Ion Chamber
Muon Shield
Courtesy of Richard Bionta
24
Redundant Commissioning Instrumentation
Instrument Purpose Adjustment Calibration and Physics risks
Direct Imager SP f(x,y), look for FEL, measure FEL u, f(x,y), x,y ND filter, Attenuators Scintillator linearity, Attenuator linearity and background
Indirect Imager Measure FEL u, f(x,y), spectral imaging of SP and FEL harmonics, attenuator calibration Mirror Angle Mirror reflectivity, damage
Total Energy FEL u Attenuators Energy to Heat, damage
Ion Chamber FEL u, x,y,x',y' Pressure Signal strength
Spectrometers FEL, SP spectra Attenuators Resolution, damage
Courtesy of Richard Bionta
25
LCLS FEL Commissioning Milestones
  • MS3BO_040 Front End Beneficial Occupancy
    (9/5/2007)
  • MS3BO_030 Undulator Facility Beneficial
    Occupancy (12/3/2007)
  • MS3_XT040 Solid Attenuator Installation Complete
    (12/14/2007)
  • MS3_XT045 Gas Attenuator Installation Complete
    (12/14/2007)
  • MS3_XT080 Start Front End Enclosure
    Commissioning (3/4/2008)
  • MS3_LN015 Start Linac-to-Undulator (LTU)
    Commissioning (5/12/2008)
  • MS3_XT066 Start Near Experimental Hall Checkout
    (6/12/2008)
  • MS3_UN020 Undulator System Installation Complete
    (7/18/2008)
  • MS3_UN025 Start Undulator Commissioning (1st
    Light) (7/24/2008)

Diagnostics needed around July 2008
26
Workshop Objective
  • Define a strategy for using spontaneous undulator
    radiation to measure the K value of every
    individual LCLS Undulator Segment after
    installation in the Undulator Hall.
  • To reach the objective, the physics and
    technologies necessary need to be identified.
    Workshop discussions will include
  • Usable spectral features of spontaneous radiation
  • Strategies for beam-based K measurements
  • Specifications for suitable instruments
  • Scheduling issues
  • Three Work Packages have been defined and
    assigned to three different groups. Work
    described by these Work Packages has been carried
    out in preparation of the workshop and will be
    presented and discussed at the workshop.

27
Workshop Agenda
28
Work Package 1 Angle Integrated Measurement
  • Group B. Yang, R. Dejus
  • Task Examine robustness of angle-integrated
    measurements of undulator spectrum. Consider
    effects of errors in beam alignment, undulator
    magnet structure, straightness of vacuum pipe,
    alignment of spectrometer, etc. Consider effects
    of location of undulator segment being tested.
    Determine what are realistic values for the
    precision with which the value of K can be
    determined for an undulator segment at the
    beginning, middle, and end of the undulator.
    This task explores the use of the high-energy
    edge of the fundamental spectral peak (the third
    harmonic may also be considered) of a single
    undulator to measure its K parameter. The
    measuring spectrometer will be located in the
    LCLS FEE, roughly 100 m downstream from the final
    undulator segment. Realistic values for the
    angular acceptance of the measurement (limited by
    beam-pipe apertures, or apertures at the
    measuring point) should be considered.

29
Work Package 2 Pinhole Measurement
  • Group J. Welch, R. Bionta, S. Reiche
  • Task Examine robustness of pinhole measurements
    of undulator spectrum. Consider effects of errors
    in beam alignment, undulator magnet structure,
    straightness of vacuum pipe, alignment of pinhole
    and spectrometer, etc. Consider effects of
    location of undulator segment being tested.
    Determine what are realistic values for the
    precision with which the value of K can be
    determined for an undulator segment at the
    beginning, middle, and end of the undulator.
    This task explores the use of the fundamental
    spectral peak (the third harmonic may also be
    considered) of a single undulator, as seen
    through a small angular aperture, to measure its
    K parameter. The measuring spectrometer will be
    located in the LCLS FEE, roughly 100 m downstream
    from the final undulator segment. Realistic
    values for the angular acceptance of the
    measurement should be determined, and the effects
    of misalignment of the aperture or undulator axis
    should be carefully considered.

30
Work Package 3 Single-Shot Spectral Measurement
  • Group J. Hastings, et al.
  • Task Assume that a single shot spectral
    measurement is needed for an LCLS spontaneous
    undulator pulse. What are the best options for
    doing the measurement? What spectral resolution
    can be obtained using these methods? What are the
    effects of beam jitter, spectrometer
    misalignment, etc? This task explores the
    design and performance of x-ray spectrometers
    capable of providing centroid or edge position
    with high resolution, on a single-shot of
    radiation from a single LCLS undulator. The
    spectrometer will most likely be located in the
    LCLS FEE, about 100 m downstream from the final
    undulator segment.

31
Workshop Charge
  • Characterize the spectral features of spontaneous
    synchrotron radiation that are usable for
    beam-based K-measurements.
  • Identify the most appropriate strategy for
    beam-based K-measurements.
  • Specify suitable instruments for the identified
    beam-based K-measurement strategy.
  • List expected performance parameters such as
    resolution of K measurement as function of beam
    charge, and segment location as well as expected
    tolerances to trajectory and energy jitter.
  • List any open questions regarding the feasibility
    of the most appropriate strategy.
  • List the RD activities, if any, needed before
    the design of a measurement system can be
    completed and manufacturing/procurement can start.

32
End of Presentation
Write a Comment
User Comments (0)
About PowerShow.com