Title: Inside the black box
1Inside the black box
2fluidics
electronics
light
optics
3Flow Cytometry Problems with early systems.
Detector
4Hydrodynamic focusing
V. Kachel, H. Fellner-Feldegg E. Menke -
MLM
Chapt. 3
5Fluidics Particle Orientation and Deformation
V. Kachel, et al. -
MLM
Chapt. 3
6The Flow Cell
The introduction of a large volume into a small
volume in such a way that it becomes focused
along an axis is called Hydrodynamic Focusing.
Original from Purdue University Cytometry
Laboratories, Modified by R. Duggan
7Sample differential pressure
10 psi
10 psi
10 psi
10.2 psi
10.4 psi
10.8 psi
- If differential is too large, cells will no
longer line up single file - Results in wider CVs and increase in multiple
cells passing through the laser at once.
No more single cell analysis!
8(No Transcript)
9What can we measure
- Scattered light
- Fluorescence
- Thats it
10Forward Angle Light Scatter
Laser
FALS Sensor
Intensity is said to be proportional to cell size
1190 Degree Light Scatter
Laser
FALS Sensor
Intensity is said to be proportional to
Granularity or internal complexity of the cell
90LS Sensor
12Lymphozyt
Basophiler
Monozyt
Neutrophiler
Eosinophiler
FACSAcademy
13Why Look at FSC v. SSC
- Since FSC size and SSC internal structure, a
correlated measurement between them can allow for
differentiation of cell types in a heterogenous
cell population
14How to visualize a specific antibody
fluorescein
15Fluorescence
Stokes Shift is 25 nm
Fluorescein molecule
520 nm
495 nm
Fluorescnece Intensity
Wavelength
low energy
high energy
wavelength (nm)
16Spectra of Common Fluorochromes
Laser Lines (nm)
PE-Texas Red
Texas Red
PI
Ethidium
PE
FITC
cis-Paranaric Acid
300
400
500
600
700
Original from Purdue University Cytometry
Laboratories, Modified by R. Duggan
17Fluorescence Detectors
Laser
FALS Sensor
Fluorescence detector (PMT3, PMT4 etc.)
18The photoreceptors of a flow cytometer
longpass filter
bandpass filter
19Flow Cytometry Optics dissection of fluorescence
PMT
4
Dichroic Longpass filters
PMT
3
Flow cell
PMT
2
Bandpass
Filters
PMT
1
Laser
20(No Transcript)
21Signal Detection - PMTsphoto multiplier tube
Secondary emission
Cathode
Anode
Amplified Signal Out
Photons in
End Window
Dynodes
Voltage
22The color blindness of a flow cytometer
Laser Lines (nm)
PE-Texas Red
Texas Red
PI
Ethidium
PE
FITC
cis-Paranaric Acid
300
400
500
600
700
Original from Purdue University Cytometry
Laboratories, Modified by R. Duggan
23(No Transcript)
24(No Transcript)
25(No Transcript)
26Measurements of the Pulse
Voltage Intensity
Time
27A mixed cell population
28(No Transcript)
29(No Transcript)
30One parameter display (Histogram)
31(No Transcript)
32(No Transcript)
33Two parameter display (Dot Plot)
FSC SSC FL1 FL2 FL3 FL4 75 42 320 10 10 9
FSC SSC FL1 FL2 FL3 FL4 20 41 7 480 8 8
FSC SSC FL1 FL2 FL3 FL4 20 41 320 480 8 8
34Two parameter display (Dot Plot)
CD45 PE
CD14 FITC
35Two parameters displayS
dot plot
3D plot
contour plot
density plot
36Data Analysis Concepts
Regions A certain characteristic (blond, little,
man, woman) Gates Logical combination of
characteristics (blond women, little men)
37Characterization of subpopulations
Gate
Auswertung
CD45 PE
CD14 FITC
3811-color, 13-parameter flow cytometry
Identification of human naive T cells by
phenotype, function, and T-cell receptor
diversity S C De Rosa, L A Herzenberg, L A
Herzenberg M Roederer Nature Medicine 7, 245 -
248 (2001).
39(No Transcript)
40http//www.cyto.purdue.edu/ Jean Paul Robinson