Title: Performance of ventilators at simulated altitude: study of FiO2


1
Performance of ventilators at simulated
altitude study of FiO2
Jean-Pierre Tourtier , Thomas Leclerc, Marc
Borne Military hospital Val-de-Grâce, Paris,
France
2
Introduction (1)
  • Aero medical evacuation ventilatory support
  • Ventilators turbine technology
  • Hypobaric
  • FiO2 preset / assessed / delivered

3
Introduction (2)
  • 2 ventilators
  • T-bird VS02
    LTV-1000

4
Methods (1)
  • ?
  • Cabin altitude simulation

5
Methods (2)
  • ?
  • Altitude simulated
  • 1500 m (4000 feet)
  • 2500 m (6700 feet)
  • 3000 m (3000 feet)

6
Methods (3)
  • ?
  • Lung simulation VT1 Bioteck
  • Normal lung
  • (resistance  5.6 cm H2O/L/min  compliance  50
    ml/cm H2O)
  • Acute Respiratory Distress Syndrom
  • (resistance  3.9 cm H2O/L/min  compliance  20
    ml/cm H2O)

7
Methods (4)
  • ?
  • Ventilation parameters

Lung model Normal lung ARDS
Vt set (ml) 700, 400 400, 250
Respiratory rate 12 20
I/E 1/2 1/1
Fi02 () 21,50,90 21, 50, 90
8
Methods (5)
  • ?
  • Measurements
  • Actual FiO2
  • paramagnetic analysis

9
Methods (6)paramagnetic analysis
  • focused magnetic field
  • Oxygen is attracted
  • electric potential
  •  gold standard 

10
Results (1) Normal lung, ARDS T-bird VS02
Cabin Altitude (m) 90 1500 2500 3000
FiO2 set 50 50 50 50
Fi02 assessed 48 43 55 50
FiO2 delivered 49 53 57 61
11
Results (2) Normal lung, ARDS T-bird VS02
Cabin Altitude (m) 90 1500 2500 3000
FiO2 set 90 90 90 90
Fi02 assessed 87 76 94 84
FiO2 delivered 88 96 99 99
12
Results (3) Normal lung, ARDS LTV-1000
Cabin Altitude (m) 90 1500 2500 3000 3000 ARDS Vt 450 ml
FiO2 set 50 50 50 50 50
Fi02 assessed 49 49 50 50 53
FiO2 delivered 50 51 53 54 50
13
Results (4) Normal lung, ARDS LTV-1000
Cabin Altitude (m) 90 1500 2500 3000 3000 ARDS Vt 450 ml
FiO2 set 90 90 90 90 90
Fi02 assessed 87 76 88 92 92
FiO2 delivered 88 96 94 95 90
14
Results (5) exampleNormal lung, FiO2 set50
15
Conclusion (1) T-bird VS02
  • ?
  • Variations between FiO2 set and delivered high
  • Variations between FiO2 delivered and assessed
    high
  • Inefficacity of ventilators Fi02 analysis

16
Conclusion (2) LTV-1000
  • ?
  • Variations between FiO2 set and delivered
    moderate
  • Variations between FiO2 delivered and assessed
    high
  • Inefficacity of ventilators Fi02 analysis

17
Conclusion (3) OXYMAN
  • ?