Title: A1258568572xhvJr
 1 A USAXS STUDY OF THE MORPHOLOGY OF 
POLYETHYLENE USED IN ORTHOPAEDIC IMPLANTS Mary 
Turell, Anuj Bellare Department of Orthopaedic 
Surgery, Brigham and Womens Hospital, Harvard 
Medical School, Boston, MA 
RESULTS AND DISCUSSION USAXS scattering curves 
were obtained by plotting the scattered intensity 
I versus q where q  (4p/?)sin? (1) such
 that ? equals one half of the scattering angle, 
and ? is the wavelength of x-rays (see Fig. 2). 
USAXS curves revealed a linear Porod region at 
ultra-low q values suggesting the presence of 
large micrometer size scatterers such as voids, 
regardless of thermal history. The scattering 
functions were converted to paired distance 
distribution functions (PDDF) or p(r) using the 
program ITP 2. PDDF is related to the 
scattering function I(q) by the following 
equation (2) where p(r) is the paired 
distance distribution function, A is the area of 
the lamella, I(q) is the experimental scattering 
function, q is the scattering vector, and r is 
the radial distance perpendicular to lamellar 
surfaces within a stack of lamellae. The USAXS 
long period for all samples was measured from the 
first maximum of p(r) (Figure 3). Together, the 
long period and DSC crystallinity were used to 
calculate the lamellar thickness (see Table 
1) These results show that the chosen thermal 
histories strongly affect the morphological 
parameters such as inter-lamellar spacing and 
lamellar thickness. Preliminary wear tests have 
shown that these parameters strongly affect wear 
rates under conditions that simulate wear in the 
joint replacement prostheses. These results will 
be used to optimize wear performance of UHMWPE in 
these orthopedic implants. 2 Glatter, Acta 
Phys.Austriaca, 47, 83-102 (1977) ACKNOWLEDGEMENT
 We gratefully acknowledge Dr. P. Jemian and Dr. 
J. Ilavsky for their assistance with USAXS 
measurements. The UNI-CAT facility at the 
Advanced Photon Source (APS) is supported by the 
University of Illinois at Urbana Champaign, 
Materials Research Laboratory (U.S. Department of 
Energy, the State of Illinois-IBHE-HECA, and the 
National Science Foundation), the Oak Ridge 
National Laboratory (U.S. DOE), the National 
Institute of Standards and Technology (U.S. 
Department of Commerce), and UOP LLC. The APS is 
supported by the U.S. Department of Energy, 
Office of Science, Office of Basic Energy 
Sciences, under Contract No. W-31-109-ENG-38.
- INTRODUCTION 
- Ultra-high molecular weight polyethylene (UHMWPE) 
 has attained worldwide acceptance as a bearing
 material used in total knee replacement
 prostheses. Each year, over 300,000 total knee
 replacement prostheses are implanted in the U.S.
 alone (Figure 1).
- The lifetime of these implants is governed by 
 UHMWPE tribological and fracture properties,
 specifically those that are related to fatigue
 failure.
- The crystalline morphology of polymers have a 
 strong influence on their mechanical and wear
 properties 1.
-  The objective of this research is to investigate 
 crystallization conditions that would optimize
 the relevant wear and mechanical properties of
 UHMWPE for application in total knee replacement
 prostheses.
- In this study, UHMWPE was subjected to four 
 different thermal histories with the objective of
 developing structure-property relationships that
 are relevant to their application as orthopedic
 bearing materials. The crystalline morphology
 resulting from these treatments was characterized
 using ultra small-angle x-ray scattering (USAXS)
 at the UNI-CAT beamline of the Advanced Photon
 Source (APS).
- 1 R.W. Truss et al., Polym Eng Sci, 20, 747-755 
 (1980).
- MATERIALS  METHODS 
-  A ram extruded rod stock (PolyHi Solidur Inc) of 
 GUR 1050 UHMWPE resin (Ticona, Bayport, TX) was
 purchased and used as a control.
-  UHMWPE rod stock were melted and subjected to  
 (1) slow cooling to room temperature, (2)
 isothermal crystallization at 125C for a period
 of 48 hours, (3) quenching in liquid nitrogen,
 (4) quenching in liquid nitrogen followed by
 isothermal crystallization at 80C for a period of
 48hrs.
-  Samples were characterized using a combination 
 of differential scanning calorimetry (DSC) and
 ultra-small angle x-ray scattering. DSC was
 performed on a Perkin Elmer Pyris 1 to determine
 the degree of crystallinity in each sample.
 Percent crystallinity was calculated by
 normalizing the heat of fusion of each sample to
 the heat of fusion of polyethylene crystal (293
 J/g).
FIGURE 1
FIGURE 2
FIGURE 3
TABLE 1