Title: EPI


1
EPI Echo Planar Imaging
  • Joakim Rydell
  • 2004-03-26

2
Echo Planar Imaging
  • Like ordinary MR, but faster
  • Ordinary MR approximately one minute
  • EPI less than one second
  • The Nobel Prize in Medicine 2003
  • Lauterbur (MR)
  • Mansfield (EPI)

3
Repetition, k-space
  • Fourier transform of image
  • Traversed using magnetic field gradients

4
Repetition, pulse sequence
  • Slice selection
  • Phase encoding
  • Frequency encoding

5
Introducing EPI
Ordinary MR
  • Acquire all rows in k-space with just one
    excitation

EPI
6
EPI The Movie
7
Going even faster Partial Fourier
  • Real image ? hermitian symmetric k-space
  • Acquire half of k-space, compute the rest
  • Not always as good as it seems...

8
Partial Fourier, drawbacks
  • Phase errors
  • Slowly varying
  • Approximately 65 of k-space required
  • SNR decrease
  • Full k-space acquisition gives sqrt(2) SNR
    improvement

9
Why use ordinary MR?
  • Signal decay
  • Low SNR
  • N/2 ghosting
  • Geometric distortion
  • dB/dt

10
Signal decay
  • Long readout period
  • Weak signal near end of acquisition
  • Blurring in phase-encoding direction

11
Low SNR
  • Fast acquisition
  • Less averaging
  • Noisy images

12
N/2 ghosting
  • Alternating acqusition directions
  • Shifted sampling positions
  • N/2 ghosting
  • Reference scans

13
Geometric distortion
  • Low bandwidth/pixel
  • Pixels shifted in phase encoding direction
  • Field maps

14
dB/dt
  • Fast gradient switching induces eddy currents,
    causing field inhomogenity
  • Rapid changes in the magnetic field can cause
    peripheral nerve stimulation

15
k-space trajectories
  • Cartesian
  • Blurring
  • N/2 ghosting
  • High dB/dt
  • Spiral
  • Non-uniform sampling
  • Cartesian spiral

16
Ramp sampling
  • Sampling on gradient ramps
  • Non-uniform sampling
  • In time
  • In k-space

17
Multishot
  • Both EPI and ordinary imaging
  • Collect several rows per excitation
  • Alignment difficulties

18
Applications
  • Cardiac imaging
  • fMRI
  • ...

19
EVI Echo Volumar Imaging
  • 3D EPI
  • Acquire a full volume with just one excitation
  • Not quite possible yet