Title: FLUKE 5700A/5720A ARTIFACT CALIBRATION


1
FLUKE 5700A/5720AARTIFACT CALIBRATION
  • MAC Study Group
  • R.Kletke, D.Deaver, B.Sorrells, M.Duffin, S.
    Dizor

2
PURPOSE OF STUDY
  • A2LA presently accepts Flukes claim for
    specifications based on the calibration cycle of
    its artifact calibration, provided the instrument
    is given a classical external calibration at
    least bi-annually. The task groups intent is to
    determine what should be done, if anything, to
    help verify those claims and to determine any
    contingencies that may be involved.

3
Artifact Calibration
  • The Fluke 5700A/5720A makes use of internal check
    standards and measurement systems so it can be
    completely calibrated in place to full
    specifications using a small number of convenient
    portable standards
  • (Operations Manual, Paragraph 1-15, p 1-8)

4
5700A/5720A with ARTIFACT CAL.
10 V DC Standard
6.5 V DC Reference.
26 Bit DAC
Calibrator Circuitry
13 V DC Reference
Switching Matrix
1 Ohm Standard
1 Ohm Reference
Multi-Range TVC Standard 1kHz -1MHz
10 kOhm Standard
1.9 Ohm Reference
5
(No Transcript)
6
(No Transcript)
7
IMPORTANT OBSERVATIONS
  • Works well for DCV, DCI and R with unbroken
    traceability chain
  • No traceability for AC parameters with Artifact
    Cal
  • Does adjust calibrator within the 24 hour
    specification for all parameters provided TVC
    standard remains stable
  • Cal Shift Report accurately reflects adjustment
    changes

8
PROBLEMS
  • IF AC traceability is established by initial
    overall calibration, then artifact calibration
    during the cal period breaks traceability
  • Measurement values and uncertainties are not
    given on the Cal Shift Report

9
SOLUTIONS
  • AC traceability and measurement values can be
    established from the artifact calibration Cal
    Shift report.
  • Measurement uncertainties are bounded by the
    calibrator 24 hour specification

10
RECOMMENDATIONS BY EU NATIONAL LAB STUDY
  • The Artifact Cal report, complete with shift data
    for all parameters, should be examined and stored
    by the user
  • The user of an instrument should be able to
    justify the calibration interval he has chosen
  • Full calibrations should be made at least every
    two years

11
CALIBRATION RECIPE GUIDELINE
EA 1 year CAC CAC CAC CAC CAC CAC
Fluke 1 year AC A AC A AC A
Study 1 year AC AC A AC A AC
Study 90 Day AC AC A AC A A A AC A A A A A A A AC A A A A A
0 3m 6m 9m 1yr 3m 6m 9m 2yr 3m 6m 9m 3yr 3m 6m 9m 4yr 3m 6m 9m 5yr
12
MAC WORKGROUP RECOMMENDATIONS
  • Accept EU National Lab study with recommendations
  • Schedule Complete Cal and Artifact Cal so that
    performance at the desired cal interval is
    confirmed by the As Found data collected during
    Complete Cal. This should be repeated at every
    Complete Cal.

13
WORKGROUP CAL GUIDELINE
CAL PERIOD
12 Mo AC AC A AC
6 Mo AC A AC A A A AC
3 Mo AC A A A AC A A A A A A A AC
Mo 0 3 6 9 12 15 18 21 24 27 30 33 36
Yr 0 1 2 36
14
ARTIFACT CAL SHIFT REPORTSample Page
15
Concluding Remarks
  • Collect, review and store Cal Shift report after
    every Art Cal. Verify cal shifts do not exceed
    claims
  • Take care during Art Cal to set 5700A/5720A to
    proper specifications
  • If User claims Fluke 99 CL specs, cal shift
    limit on report is 100
  • If User claims Fluke 95 CL specs, cal shift
    limit on report is 77
  • Take care that As Found data collected during
    full calibration is checked against the
    performance claimed by Art Cal period.