BTM 024 qualification
  • MRB Meeting dd month yyyy

2
Description and reference documentation
  • Description of nonconformance
  • Problem description
  • Instances
  • Where was it found
  • Suspected Root cause
  • description
  • Reference NCR List
  • NCR n

3
Vacuum test
  • Vacuum test
  • Performed in the vacuum chamber in INFN-Roma 2.
  • Tray connected to the readout electronics by
    means of a single flex cable in order to minimize
    the number of the lines to be brought out of the
    vacuum chamber (all the FE chips read by the left
    GTRC).
  • 20 hours _at_ P lt 0.1 mbar, room temperature (25
    C).
  • Noise and gain of each single channel
    continuously monitored throughout the test.

4
Vacuum profile
5
Functional tests
  • Functional tests
  • Test suite run continuously during the night.
  • Noise and gain measured at the level of the
    single strip every 15 minutes.
  • 80 tests performed in total, no failures
    detected.
  • One noisy channel since the beginning, no more
    channels were lost during the test.

6
Thermal cycles
  • Thermal cycles
  • Performed in the climatic chamber _at_ INFN-Pisa.
  • Tray connected to the TEM through both flex
    cables.
  • 4 complete cycles from -15 C to 45 C (standard
    tray thermal cycles).
  • Noise and gain and noise occupancy of each single
    channel continuously monitored throughout the
    test.

7
Functional tests
  • Functional tests
  • Test suite run continuously throughout the whole
    test.
  • Noise and gain/noise occupancy measured at the
    level of the single strip every 15 minutes.
  • 93 tests performed in total, no failures
    detected.
  • Final complete test suite after the thermal
    cycles successfully performed.

8
Conclusion
  • Description of nonconformance
  • Problem description