Title: Evaluation of various diagnostic x-ray measuring devices
1Evaluation of various diagnostic x-ray measuring
devices
- MDCH Radiation Safety Section
- Don Parry, CHP
2Evaluation of X-ray Measuring Devices
- Evaluation Criteria (Don)
- Accuracy
- Reliability
- Range of Application
- Durability
- Ease of Use
- Cost
3Evaluation of X-ray Measuring Devices
- Evaluation Criteria (Dons Bosses)
- Cost
- Durability
- Reliability
- Range of Application
- Accuracy
- Ease of Use
4Evaluation of X-ray Measuring Devices
- Evaluation Criteria (Dons Staff)
- Ease of Use
- Reliability
- Range of Application
- Durability
- Accuracy
- Cost
5Evaluation of X-ray Measuring Devices
- Types of Devices
- Ion Chambers
- Solid State Detectors
6Evaluation of X-ray Measuring Devices
- Ion chambers
- Advantages
- Flat Energy Response
- History of Reliability
- Disadvantages
- Weight
- Chambers can be fragile
7Evaluation of X-ray Measuring Devices
- Solid State Detectors
- Advantages
- Small and lightweight
- Detectors are more robust
- Disadvantages
- More energy dependant
- Generally not as simple to use
8Evaluation of X-ray Measuring Devices
- Application of X-ray Measuring Devices
- Dental X-ray ( 50 90 kVp)
- Medical X-ray (45 150 kVp)
- Mammography (24 35 kVp)
9Evaluation of X-ray Measuring Devices - Accuracy
- Ion Chambers Medical and Dental
- Tested at four x-ray beam codes
- M50 (50 kVp 1.0 mm Al HVL)
- L80 (80 kVp 1.8 mm Al HVL)
- L-100 (100 kVp 2.8 mm Al HVL)
- M-100 (100 kVp 5.0 mm Al HVL)
- MoMo 25 (25 kVp, Moly target 0.30 mm Al HVL)
- MoMo 30 (30 kVp, Moly target 0.35 mm Al HVL)
10(No Transcript)
11(No Transcript)
12(No Transcript)
13Evaluation of X-ray Measuring Devices
- Solid State Detectors Medical and Dental
- Tested at three x-ray beam codes
- L80 (80 kVp 1.8 mm Al HVL)
- L-100 (100 kVp 2.8 mm Al HVL)
- M-100 (100 kVp 5.0 mm Al HVL)
14(No Transcript)
15(No Transcript)
16(No Transcript)
17(No Transcript)
18Evaluation of X-ray Measuring Devices
- Comparison
- Ion Chambers vs. Solid State Detectors
- Do not use any detector outside its specified
energy range
19(No Transcript)
20All the solid state detectors showed under
response at low energies
21Evaluation of X-ray Measuring Devices
- Ion Chambers and Solid State Detectors In
Mammography - Tested at two beam codes
- MoMo 25 (25 kVp, Moly target 0.30 mm Al HVL)
- MoMo 30 (30 kVp, Moly target 0.35 mm Al HVL)
22(No Transcript)
23(No Transcript)
24(No Transcript)
25(No Transcript)
26(No Transcript)
27Evaluation of X-ray Measuring Devices
- Some Mammography Machines now use beams with
higher energies - NIST currently does not provide calibration
standards for these beams - Since the energy response is typically flat for
ionization chambers, no additional corrections
should be needed. - Solid state instruments, however, may require
appropriate corrections obtained from the
instrument manufacturer.
28Evaluation of X-ray Measuring Devices
- Ion Chambers Reliability
- Ion chamber calibrations have historically been
very stable. - Ion chambers can be fragile and require
appropriate handling by field staff
29(No Transcript)
30Evaluation of X-ray Measuring Devices
- Solid State Detectors Reliability
- While reliability has been good, radiation
response has drifted or failed on some sensors - Detectors appear more robust and are less often
damaged by misuse
31(No Transcript)
32Evaluation of X-ray Measuring Devices - Range of
Applicability
- Field Data IC 1 versus SS 3
Type IC 1 mR SS 3 mR diff
Dental 190.9 197.9 3.7
Podiatrist 7.2 7.3 1.4
Medical 49.8 47.7 -4.2
Mammo 287.2 296.1 3.1
33Evaluation of X-ray Measuring Devices Ease of
Use
- IC1- IC3
- Care needed in connecting probes
- SS1
- Set up difficult when used on some machines
- SS2
- Detector cable fixed to base
- SS3
- Some felt it was not as simple to use
34Evaluation of X-ray Measuring Devices Our
Conclusions
- Ion chambers exhibit good energy response over a
wide range of beam qualities new chambers are
lighter and smaller - SSDs can be designed to have acceptable energy
responses in the diagnostic x-ray range, but know
your instruments limitations - Field staff like the small size and light weight
of SSD detectors
35Evaluation of X-ray Measuring Devices Our
Conclusions
- Send equipment to an accredited calibration lab
annually to ensure radiation response remains in
specification - We use SSD for equipment with known limits on the
energy range - We use ion chambers for mammography beams and on
x-ray beams in which beam qualities can vary
significantly from machine to machine.