Dosimetry by PulseMode Detectors II - PowerPoint PPT Presentation

1 / 44
About This Presentation
Title:

Dosimetry by PulseMode Detectors II

Description:

... as detectors of ionizing radiation, especially in nuclear physics ... for an ionization chamber in a -ray survey meter for health-physics applications ... – PowerPoint PPT presentation

Number of Views:90
Avg rating:3.0/5.0
Slides: 45
Provided by: michae354
Category:

less

Transcript and Presenter's Notes

Title: Dosimetry by PulseMode Detectors II


1
Dosimetry by Pulse-Mode Detectors II
  • Scintillation Dosimetry
  • Semiconductor Detectors for Dosimetry

2
Scintillation Dosimetry Introduction
  • Many transparent substances, including certain
    solids, liquids, and gases, scintillate (i.e.,
    emit flashes of visible light) as a result of the
    action of ionizing radiation
  • By using a sensitive light detector such as a
    photomultiplier (PM) tube, the light emitted can
    be converted into an electrical signal

3
Introduction (cont.)
  • A light photon incident on the photocathode of a
    PM tube may release an electron, which is then
    numerically amplified as much as ? 107 in passing
    through the dynode chain in the tube
  • Either the output electrical pulses from numerous
    such events can be counted (with or without
    pulse-height analysis), or the output current can
    be measured

4
General schematic design for a scintillation
detector for dosimetry applications. A light
pipe optically couples the scintillator to the
photomultiplier tube. The scintillator is
otherwise enclosed in an optically opaque and
internally reflective envelope that may also
filter out short-range radiation and may
additionally serve as a CPE buildup layer for
indirectly ionizing radiation.
5
Introduction (cont.)
  • Scintillators have been widely applied as
    detectors of ionizing radiation, especially in
    nuclear physics
  • Very fast decay times, down to 10-9 s, make
    organic liquid and plastic scintillators
    excellent choices for coincidence measurements
    with good time-resolution, and they can occupy
    whatever volume shape and size one wishes

6
Introduction (cont.)
  • Scintillators, especially NaI(Tl), also have been
    used extensively for x- and ?-ray energy
    spectrometry but have been largely replaced by
    Si(Li) and Ge(Li) semiconductor detectors for
    best energy resolution
  • However, scintillators continue to be widely used
    for those ?-ray spectrometry applications in
    which resolution is less critical, because of
    lower cost and the greater convenience of
    operating the detector at room temperature
    instead of inside a cryostat

7
Light Output Efficiency
  • Only a very small part of the energy imparted to
    a scintillator appears as light the rest is
    dissipated as heat
  • In typical situations 1 keV of energy is spent
    in the scintillator for the release of one
    electron from the PM tubes photocathode
  • However, the large gain available in the PM tube
    and external amplifiers still provides an
    adequate output signal

8
Efficiency (cont.)
  • The light generated in a scintillator by a given
    imparted energy depends on the LET of the charged
    particles delivering the energy
  • In typical organic scintillators, increasing the
    particle LET decreases the light output for a
    given energy imparted, as can be seen in the
    following diagram

9
Light output vs. particle energy for electrons
and protons stopped in the plastic scintillator
NE-102. The light output is proportional to
electron energy, but not to proton energy.
10
Efficiency (cont.)
  • The light response from electrons that spend
    their full track length in the scintillator is
    found to be proportional to their starting energy
    above about 125 keV
  • For protons, the light output is only about 15
    as great as for electrons at 1 MeV, rising to
    about 40 as great at 10 MeV
  • The technique of pulse-shape discrimination
    allows the separation of dose components on the
    basis of particle LET

11
Efficiency (cont.)
  • For dosimetry of ?-rays or electrons, either the
    PM-tube output should be measured as an electric
    current or the pulse-heights must be analyzed and
    calibrated in terms of dose, as discussed for
    proportional counters
  • Simple counting of pulses without regard to their
    size is not a measure of the dose in a
    scintillator

12
Scintillator Types
  • For most dosimetry applications where soft tissue
    is the dose-relevant material, organic plastic
    scintillators such as NE-102, organic liquids
    such as NE-213, and the organic crystals stilbene
    and anthracene are the most useful because they
    are made mostly of the low-atomic-number elements
    C and H
  • Thus they do not overrespond to photons through
    the photoelectric effect, and the hydrogen
    content makes the (n, p) elastic-scattering
    interaction the main process for fast-neutron
    dose deposition, as it is in tissue

13
Characteristics of some scintillators
14
Light Collection and Measurement Scintillator
Enclosure
  • A light reflector, optimally a thin layer of MgO
    powder, is useful to maximize light-collection
    efficiency from a scintillator
  • If the scintillator has polished surfaces, all
    the light incident from the inside is reflected
    if the angle of incidence is greater than the
    critical angle
  • The MgO reflector will recapture most of the
    light that escapes at smaller angles

15
Scintillator Enclosure (cont.)
  • For small or thin scintillators (plastic ones may
    be as thin as 20 ?m) one should keep in mind
    cavity-theory considerations
  • Simplest dosimetric interpretation for indirectly
    ionizing radiation calls for surrounding the
    scintillator by a nonscintillating layer of the
    same composition, thick enough to provide CPE

16
Scintillator Enclosure (cont.)
  • In the case of plastic scintillators a shell of
    Lucite will usually suffice, surrounding the thin
    reflector
  • Outside of this an opaque covering such as
    aluminum foil is required to exclude ambient
    light
  • NaI and CsI scintillators require hermetic seals,
    as they are hygroscopic

17
Light Collection and Measurement Light Pipe and
PM Tube
  • The exit surface of a scintillator is optically
    coupled to the PM-tube photocathode through a
    light pipe, usually consisting of a solid
    cylinder of polished Lucite
  • The interfaces are filled with an optical
    coupling agent such as high-viscosity silicone
    oil or transparent epoxy cement
  • Ideally all materials along the optical path
    should have nearly the same refractive index as
    the glass face of the PM tube, ? 1.5

18
Light Pipe and PM Tube (cont.)
  • The main purpose of the light pipe in dosimetry
    is to remove the PM tube from the radiation field
    that the scintillator is measuring
  • PM tubes are capable of responding to ionizing
    events occurring within their structure
  • The interactions occur in different media than
    the scintillator, at different locations, and
    with variable gain factors
  • Large doses can so damage a PM tube that its
    light sensitivity is permanently decreased

19
Comparison with an Ionization Chamber
  • Scintillators are often used as a more sensitive
    substitute for an ionization chamber in a ?-ray
    survey meter for health-physics applications
  • It is instructive to consider what factors are
    involved in estimating the difference in current
    output from a scintillator and an ion chamber of
    the same volume

20
Comparison with an Ionization Chamber (cont.)
  • The analogue in a scintillator of W in a gas is
    the average energy spent by an electron per light
    photon produced
  • For plastic scintillators this is around 60 eV
    (about twice that for gases)
  • For good optical coupling 1/3 of the photons
    reach the photocathode (typical efficiency about
    15 and tube gain about 106)
  • Thus for equal masses of chamber gas and plastic
    scintillator, the output current of the latter is
    3 ? 104 greater

21
Pulse-Shape Discrimination
  • In most scintillators, the promptly emitted light
    comprises nearly all of the observed
    scintillation
  • In some materials a sizable longer-time-constant
    component exists that is LET-dependent
  • Particles with denser tracks thus have a more
    pronounced component of longer decay time
    constant, as shown in the following diagram

22
Time dependence of scintillation pulses in
stilbene, normalized to equal heights at time
zero, when excited by radiations of different LET
23
Pulse-Shape Discrimination (cont.)
  • Suitable electronic discrimination can be
    provided to count pulses of different lengths
    separately, correlated with the LET of the
    particles that produced them
  • Thus it becomes possible to apply different dose
    calibrations to pulse heights for radiations
    having different LETs
  • This feature is especially useful for dosimetry
    in combined neutron-?-ray fields

24
Pulse-Shape Discrimination (cont.)
  • Combinations of two different scintillators
    coupled to the same PM tube are useful for some
    dosimetry situations
  • The scintillators chosen have different decay
    times so pulse-shape discrimination can be
    applied to separate the signals
  • One thin scintillator can be used to stop a
    relatively non-penetrating component of radiation
    while a thicker scintillator behind the first
    interacts more strongly with more penetrating
    ?-rays

25
Beta-Ray Dosimetry
  • A plastic scintillator covered by a thin opaque
    window and coupled to the PM tube face can be
    used to measure the planar energy-flux density
    due to incident ?-rays, assuming that the
    scintillator is thick enough to stop them, and
    that the light output is proportional to ?-ray
    energy
  • The distribution of dose vs. depth can be
    obtained from the reductions in light output
    observed when a series of tissue-equivalent
    plastic absorbing layers are placed over the
    front of the scintillator

26
Semiconductor Detectors Introduction
  • Semiconductor detectors have characteristics that
    make them very attractive as dosimeters, for
    measuring either dose or dose rate, as a
    substitute for an ion chamber
  • They can also serve as a solid-state analogue of
    a proportional counter, since the ionization
    produced by a charged particle in traversing the
    sensitive volume of the detector is proportional
    to the energy spent, irrespective of LET, for
    particles lighter than ?s

27
Introduction (cont.)
  • Some internal amplification is even possible in
    the avalanche detector mode of operation, but
    external amplification is usually preferred
  • The broad lack of LET-dependence is an advantage
    over scintillation detectors, allowing simpler
    interpretation of pulse heights in terms of
    energy imparted
  • Semiconductor detectors may be employed as
    neutron dosimeters by measuring the resulting
    radiation damage done by the neutrons

28
Basic Operation of Reverse-Biased Semiconductor
Junction Detectors
  • The following diagram illustrates the operation
    of a typical reverse-biased semiconductor, the
    silicon p-n junction
  • The bulk of the crystal consists of a p region
    having an excess of holes, while a thin layer
    at the surface is an n region having an excess
    of electrons
  • Electrical conduction in each region occurs
    through motion of these majority charge carriers

29
(No Transcript)
30
Reverse-Biased Detectors (cont.)
  • Then a positive potential (10 103 V) is
    applied to the n-terminal relative to the
    opposite evaporated-metal surface contact,
    electrons and holes are pulled out of an
    intermediate region called the depletion layer,
    and current cannot then flow across the junction
    except for some leakage current
  • If a charged particle passes through the
    depletion layer while the junction is in this
    reverse-biased condition, it forms electron-hole
    pairs by the usual collision processes

31
Reverse-Biased Detectors (cont.)
  • The mean energy spent per electron-hole pair in
    Si at 300 K is 3.62 eV for ?s and 3.68 for
    electrons, and in Ge at 77 K it is 2.97 eV for
    both
  • These figures are only about one-tenth of the
    analogous W-values for gas ion chambers hence
    10 times as much ionization is formed in
    semiconductor detectors as in ion chambers for
    the same energy expenditure
  • This also helps account for the good energy
    resolution of Si and Ge detectors

32
Reverse-Biased Detectors (cont.)
  • Electrons have mobilities of 1350 cm/s per V/cm
    in Si and 3900 in Ge, at 300 K
  • Hole mobilities are 480 cm/s per V/cm in Si and
    1900 in Ge, at 300 K
  • Thus typically they can reach the boundary of the
    depletion layer in 10-7 10-8 s, producing a
    comparable voltage-pulse rise time
  • A charge-sensitive linear preamplifier and linear
    voltage amplifier comparable to those used for
    proportional counters, but with suitably shorter
    time constants, are used to amplify the charge
    pulses for charge measurement or pulse-height
    analysis and counting

33
Silicon Diodes without Bias
  • Although the sensitivity is greater and the
    response time is less for Si diode detectors with
    reverse bias applied, for DC operation there is
    an advantage in operating without any external
    bias
  • As the bias voltage is reduced to zero, the DC
    leakage current decreases more rapidly than the
    radiation-induced current
  • The residual zero-bias radiation-induced current
    results from alteration of charge-carrier
    concentrations, and in turn gives rise to a
    potential difference between the electrodes

34
Design of unbiased silicon p-n junction
35
Operation of unbiased silicon p-n junction
36
Silicon Diodes without Bias (cont.)
  • The ranges of dose rate that are measured in
    radiotherapy applications (0.03 3 Gy/min)
    produce adequate output currents from an unbiased
    silicon diode detector with a typical sensitivity
    of 2 ? 10-11 A per R/min

37
Lithium-Drifted Si and Ge Detectors
  • These are prepared by diffusing Li ions into
    high-purity (but slightly p-type) Si or Ge
    crystals
  • The Li ions lodge at interstitial positions next
    to the electron-acceptor sites, then capture
    electrons to become electron donor sites, which
    thereby neutralize the acceptor sites
  • The crystal is then said to be compensated, by
    having the same number of electrons in the
    conduction band as it has holes in the valence
    band

38
Lithium-Drifted Detectors (cont.)
  • In this condition it acts like an intrinsic
    material, that is, one that is free of all donor
    and acceptor sites, being almost completely pure
  • Drifted regions up to almost 2 cm in thickness
    can be achieved in this way, and the entire
    intrinsic volume acts as the dosimeters
    sensitive volume
  • Changing the applied potential varies the
    electric field strength across this volume, but
    doesnt change its depth

39
Lithium-Drifted Detectors (cont.)
  • Si(Li) and Ge(Li) detectors can be made as thin
    as 10 ?m to serve as dE/dx measuring devices
    for charged particles passing through, by which
    is meant that they respond proportionally to the
    collision stopping power of the material
    (ignoring ?-ray production)
  • Likewise they can serve as thin dosimeters, or to
    measure LET distributions of charged-particle
    fields

40
Lithium-Drifted Detectors (cont.)
  • Ge(Li) detectors are preferred over Si(Li) for x-
    or ?-ray spectrometry above 50 keV, or for
    energy-fluence measurements, because the higher Z
    (32) of Ge gives it a greater photoelectric cross
    section than Si (Z 14), so that Ge stops the
    beam more efficiently
  • Si(Li) detectors are preferred for lower-energy x
    rays and for ?-ray dosimetry because their
    backscattering is much less

41
Lithium-Drifted Detectors (cont.)
  • One disadvantage of Ge(Li) and Si(Li) detectors
    is that, to maintain their energy resolution for
    spectrometry, they must be maintained and
    operated at liquid-N2 temperature
  • Allowing Ge(Li) detectors ever to warm up to room
    temperature deteriorates them by allowing the Li
    ions to migrate, thus disturbing donor-acceptor
    compensation
  • Si(Li) detectors usually may be allowed to reach
    room temperature without damage, because of lower
    Li-ion mobility

42
Use of Si(Li) as an Ion-Chamber Substitute
  • The density of Si is about 2.3 g/cm3, or about
    1800 times that of air
  • Thus, considering also the W difference, a
    Si(Li) detector will produce about 18,000 times
    as much charge as an ion chamber of the same
    volume, in the same x-ray field, at energies (gt
    100 keV) where the photoelectric effect is
    unimportant

43
Use of Si(Li) Junctions with Reverse Bias as
Counting Dose-Rate Meters
  • Si(Li) detectors 1 mm thick have been used as
    probes for measuring the depth dose due to heavy
    charged particles, including pions
  • The pulse height was found to be proportional to
    the energy spent by the particle in the sensitive
    volume of the detector
  • Dose vs. LET results have been consistent with
    those of a Rossi proportional counter

44
Fast-Neutron Dosimetry
  • Silicon detectors are damaged by very high doses
    (gt 104 Gy) of electrons or x rays, but are much
    more sensitive to damage by fast neutrons
  • Doses of 0.1 to 10 Gy (tissue) cause permanent
    defects in the Si crystal lattice, which act as
    traps for charge carriers
  • As a result the resistance of the detector is
    effectively increased
  • The voltage drop across the detector when a
    constant test current is passed through it in a
    forward direction increases gradually vs. dose
Write a Comment
User Comments (0)
About PowerShow.com