Title: Development of 3D microscopy of biological specimens using soft X-ray diffraction
1Development of 3D microscopy of biological
specimens using soft X-ray diffraction
- David Sayre, Chris Jacobsen, Janos Kirz
- David Shapiro, Tobias Beetz, Enju Lima
- Stony Brook University
- Malcolm Howells
- Advanced Light Source
2What is coherent X-ray diffraction imaging good
for?
3Atomic resolution imaging electrons or photons?
- 100 keV electrons
- About 2.5 inelastic scatters per elastic scatter
- About 45 eV deposited per inelastic scatter
- Therefore about 102 eV deposited per elastic
scatter - A hundred scattered electrons 102102 eV into (2
Å)3, or 2?1011 Gray
- 10 keV photons
- About 10 absorption events per elastic scatter
- About 10 keV deposited per absorption
- Therefore about 105 eV deposited per elastic
scatter - A hundred scattered photons 102 105 eV into (2
Å)3, or 2?1014 Gray
- Electrons are better than photons for atomic
resolution imaging - J. Breedlove and G. Trammel, Science 170,
1310 (1970) R. Henderson, Q. Rev. Biophys. 28,
171 (1995). - Crystallographys answer spread the dose out
over many identical unit cells
4Johnson Blundell, JSR 1999
- At DESY, a 500 GeV colliding-beam TESLA project
plans to produce a free-electron laser (FEL) of
breathtaking brilliance for 1 A X-rays. A
totally new science has to be explored. What
possibilities could this open for structural
biology? - With the extremely bright source it may be
possible to escape the benevolent tyranny of the
crystal and record molecular transforms from
individual molecules. Studies by Miao,
Charalambous, Kirz Sayre (private
communication) have shown that the soft-X-ray
molecular transform of a micrometre-size
non-crystalline specimen can be inverted to form
an image in which the phase problem is overcome
by over-sampling the diffraction pattern and use
of an iterative algorithm. In order to record a
molecular transform, would the molecule need to
be tethered in order to localize it sufficiently
or would a spray technique prove possible?
....Finally, will the biological molecule
withstand such a bright beam?
5Thick samples, lower resolutionphotons come out
ahead
X-rays better for thicker specimens. Sayre et
al., Science 196, 1339 (1977) Schmahl Rudolph
(1990). Electrons better for thin!
This plot Jacobsen, Medenwaldt, and Williams, in
X-ray Microscopy Spectromicroscopy (Springer,
1998)
6The role of coherent X-ray diffraction
- A technique for 3D imaging of 0.5 20 µm
isolated objects - Too thick for EM (0.5 µm is practical upper
limit) - Too thick for tomographic soft X-ray microscopy
(depth of focus lt 1 µm at 10 nm resolution for
soft X-rays even if lenses become available)
- Goals
- 10 nm resolution (3D) in 1 - 10µm size
biological specimens - (small frozen hydrated cell, organelle see
macromolecular aggregates) - Limitation radiation damage!
- lt4 nm resolution in less sensitive
nanostructures - (Inclusions, porosity, clusters, composite
nanostructures, aerosols) - eg molecular sieves, catalysts, crack
propagation
7Imaging beyond the lens limit?
- Working from diffraction patterns
- No optics-imposed resolution limits!
- Working from images
- The lens phases all the Fourier plane
information. - The lens limits resolution.
- Holography mix in a reference phase
8Where we really want to be
- Collect a high resolution 3D data set in an hour
or two - Reconstruct reliably in a comparable amount of
time
9Challenges 1/ getting sufficient coherent
photons
Currently we run at NSLS X1B. Designed for
spectroscopy. Poor use of brightness This fall
we go to the ALS, where 9.0.1 delivers 1000 times
more coherent photons over restricted energy
range. After upgrade both ALS and NSLS will
deliver at least 1000 times more yet!
(assuming custom beamline)
2/Choice of energy - Reconstruction of
real objects is easier. - Electron density
becomes more nearly real as energy increases. -
Exposure time scales as E4 . (Coherent flux for
fixed brightness scales as E-2, and cross
section scales as E-2) - Best compromise
specimen dependent 0.5 4 KeV
10ALS Brightness Now and After Upgrade
In-vacuum IDs
Superconducting ID
11NSLS II Design (J. Murphy, 6/20/03)
12Challenges 3/ recording the pattern
- Eliminate higher orders aperiodic undulator?
- Shielding detector from all but diffracted signal
- Aligning specimen with small beam-spot,
- Keeping it aligned as specimen is rotated
- Minimizing missing data
- (beam stop, large rotation angles, etc.)
- Dynamic range of detector
- Automation of data collection
- Mapping recorded data onto cubic array
13(No Transcript)
14(No Transcript)
15Challenges 4/reconstruction
- How to avoid stagnation local minima?
- The enantiomorph problem
- How to tell whether algorithm converged?
- (easy when object known)
- Multiple random starts
- How to make best use of the data?
- Of prior knowledge?
- How to optimize use of computer resources?
- Want 10243 DFT in 0.5 sec
- Much work remains to be done!
3D easier than 2D highly overdetermined
16Model calculations (E. Lima)
- Jacobsen model of biological cell (complex)
- Success in 3D using difference map
- (V. Elser), loose support
- Much easier for high contrast (especially binary)
specimens
17Challenges 5/ damage
- The ultimate limitation for radiation-sensitive
materials only - Dose fractionation
- (Hegerl and Hoppe 1976, McEwen 1995)
- M. Howells - tomorrow
18Preparation of Frozen-hydrated Fibroblasts
- Grids with live cells are
- Taken from culture medium and blotted
- Plunged into liquid ethane (cooled by liquid
nitrogen) - Loaded into cryo holder
19Radiation damage resistance in cryo
Maser et al., J. Microscopy 197, 68 (2000)
Left frozen hydrated image after exposing
several regions to 1010 Gray
Right after warmup in microscope (eventually
freeze-dried) holes indicate irradiated regions!
20PMMA at room, LN2 temperature
- T. Beetz, Stony Brook
- Repeated sequence dose (small step size, long
dwell time), spectrum (defocused beam) - Images dose region (small square) at end of
sequence
Room temperature mass loss immediately visible
LN2 temperature no mass loss immediately visible
After warm-up mass loss becomes visible
21PMMA at LN2, room temperatureXANES spectra
- T. Beetz, C. Jacobsen JSR May 2003
- Peak at 531.4 eV C0 bond
- Plateau at 540 eV total mass (plus some emphasis
on oxygen ? bonds)
C0 peak
Plateau
C0 peak
Plateau
Liquid nitrogen temperature
Room temperature
22Cryo protects PMMA against mass loss, but not
against chemical change!
23Near-term future
- Stony Brook/NSLS chamber
- Incorporates low resolution X-ray microscope
- Provision for frozen hydrated samples
- Precision positioning and rotation of specimen,
optics - Our Goals
- Develop technique (recording and reconstruction)
- Study dose vs resolution damage limits
- 3D reconstruction of frozen hydrated dwarf yeast
cell
24Freeze-dried yeast cells
25Diffraction pattern from freeze-dried yeast cell
- ln v I
- Combination of multiple exposures
- Edge of figure 45 nm
- pattern extends half way
26Azimuthally averaged power spectrum
27Soft X-ray 3D imaging using holographic
techniques
- Tobias Beetz
- Chris Jacobsen
28Why hologaphy?
- Encode phase for cases where it's difficult to
reconstruct - Hologram recorded interference between a
reference wave and a object wave - H (R S)(R S) RR SR RS SS
29In-line holography
- Resolution given by detector resolution (need to
record fine fringes) - photographic film/ photoresists (Howells et al.,
Science 238, 514 (1987) Jacobsen et al., JOSA
A7, 1847 (1990) Lindaas et al., JOSA A13, 1788
(1996)) - read out holograms, digitize gtgt not feasible for
3D data sets
30In-line holograms and reconstructions
- 1 µm diameter latex spheres
hologram
hologram
31Fourier transform holography
- resolution limited by the reference source size,
not by detector - use zone plate (Reuter and Mahr, J. Phys. E9, 746
(1976) McNulty et al., Science 256, 1009 (1992)) - better for radiation sensitive specimens (low
efficiency optics are in front of the specimen) - combine FTH with diffraction to help with phasing
the diffraction pattern
Experiments and computer simulations in progress!
32Conclusions
- Diffraction based imaging is the method of choice
for micron-size specimens - Damage will set limit on resolution for
radiation-sensitive specimens - Much progress on 2D problems, 3D just starting
- New, bright synchrotron sources will be superb
for this! - Surely an exciting prospect!
33- Phase and amplitude reconstruction in STXM using
a segmented detector - M. Feser and C. Jacobsen
34Detector developmentsegmented, current mode
M. Feser, B. Hornberger, C. Jacobsen (Stony
Brook) P. Rehak, G. de Geronimo (BNL
Instrumentation) L. Strüder, P. Holl (MPI
München)
- Silicon drift detector
- Simultaneous recording of bright field, dark
field, differential phase and interference
contrast (Polack Joyeux) - No significant upper limit to signal rate.
Acceptable dark noise (8 photons/msec
equivalent room temperature) - High quantum efficiency (gt90)
Assembly 40 mm across Active area 600 ?m
35Segmented silicon drift detector
- Corner of silicon nitride window silicon at 45
wall slope forms a prism - Refraction of x-ray beam in opposite direction
from visible light prisms
X-ray refractive index
All channels acquired simultaneously
36Amplitude And Phase Contrast Transfer Functions
- Concept developed by electron microscopy
community Waddel et al., Optik 54 (1979) - Description for x-ray microscopy Morrison et al.
1992, former XRM proceedings - New transfer functions for multisegmented
detectors with circular geometry (i.e., not CCD
pixels)
37Contrast Transfer Functions Example
- The transfer functions depend on the illumination
pupil function and the detector element geometry - Amplitude transfer has even symmetry
- Phase transfer has odd symmetry
- Difference images of opposing detector segments
have no amplitude transfer, only phase transfer
? Differential phase contrast
38Simultaneous Availability Of Contrast Modes
- Silica spheres 1 mm diameter or less
- Differential phase contrast filters out intensity
fluctuations of the source!
39Presenting DPC data
- Color code phase gradient in images
1mm
DPC
Incoherent BF
Nice, but can be difficult to interpret
40Reconstruction Filters
- They phase the detector images with respect to
each other
41Experiment Germanium Test Pattern
- Fabricated by S. Spector
- Production process of zone plates
- Quite absorbing, but b/d 0.346 _at_ 520eV (Henke,
Gullikson, et al.)
40 nmhalf-period
STEM image
42Unfiltered Incoherent Image
- Determined thickness of spokes to be ? 100 nm
from absorption profile - Expected phase shift is ? 40o in spokes
- Imaging zone plate 80mm diameter30nm outermost
zones
43Test Pattern Specimen Estimate
- Phase image shows finer details polymer resist
left from production - X-ray beam intensity fluctuations only transfer
to amplitude! - Measured ratio of d/b 0.3-0.4 compares well with
tabulated data
Phase
Amplitude
44Power Spectral Density
45Conclusions
- STXM amplitude and phase recovery with a
configured detector with good quantitative
agreement for high spatial frequencies - Spatial resolution at the diffraction limit of
the 30 nm zone plate (structures of 20 nm clearly
visible) - Coherence a prerequisite for full spatial
resolution! - Imaging without absorption by tickling
electrons below the carbon edge?
46Acknowledgements
- ALS Malcolm Howells
- DOE support
- NIH support
- David Sayre, Chris Jacobsen, Michael Feser,
Chi-Chang Kao, David Shapiro, Tobias Beetz, Onur
Mentez, Enju Lima