Title: Dosimetry by PulseMode Detectors I
1Dosimetry by Pulse-Mode Detectors I
- Geiger-Müller and Proportional Counters
2Introduction
- This chapter deals with dosimetry by means of gas
proportional counters, Geiger-Müller counters,
scintillators, and semiconductor detectors - The objective of the present chapter is to
discuss the characteristics of these devices that
make them useful for dosimetry, and how their
output signals can be interpreted in relation to
absorbed dose - Principles of operation will be introduced only
to the extent necessary to achieve that goal
3Gas Multiplication
- Any ionization chamber with sufficiently good
electrical insulation can in principle be
operated at an applied potential great enough to
cause gas multiplication, also called gas
amplification or gas gain - This is a condition in which free electrons from
ionizing events can derive enough kinetic energy
from the applied electric field, within a
distance equal to the electrons mean free path
?e to ionize other gas molecules with which they
collide
4Gas Multiplication (cont.)
- Thus a single electron can give rise to an
avalanche, as the number of free electrons
doubles repeatedly in their flight toward the
anode - At atmospheric pressure the minimum field
strength required for the onset of gas
multiplication is 103 V/mm
5Gas Multiplication (cont.)
- Cylindrical counter geometry, with a thin axial
wire serving as the anode and a cylindrical shell
as the cathode, is often employed - This provides a sheathlike gas volume immediately
surrounding the wire, in which the electric field
strength ? is much larger than the average value
obtained by dividing the applied potential P by
the cathode-anode separation
6Gas Multiplication (cont.)
- The electrical field strength ?(r) at radius r
from the cylindrical axis is given by ?(r) P/r
ln (a/b), where b is the radius of the wire anode
and a that of the coaxial cylindrical cathode - Thus the maximum electric field, ?(r)max, occurs
at the surface of the wire, where it reaches a
value of ?(r) P/b ln (a/b) - ?(b) is approximately proportional to the
reciprocal of the wire radius for constant a and P
7Gas Multiplication (cont.)
- The gain factor G for cylindrical geometry is
given approximately by - where G is the number of electrons that
arrive at the wire anode per electron released by
ionizing radiation in the gas volume outside of
the gas-multiplication sheath surrounding the
wire
8Gas Multiplication (cont.)
- ?V is the average potential difference (eV)
through which an electron moves between
successive ionizing events, which is greater than
the ionization potential Vi because of energy
wasted in atomic excitations - K is the minimum value of the electric field
strength per atmosphere of gas pressure, below
which multiplication cannot occur in a given gas - p is the gas pressure in atmospheres
- P, a, and b are defined as before
9Gas Multiplication (cont.)
- Some typical values of K and ?V in gases that are
often employed to achieve useful gas
multiplication in proportional counters are given
in the following table - For example, a cylindrical proportional counter
with a 1 cm, b 10-3 cm, P 1000 V, and
containing P-10 gas at 1 atm would have a gain
factor of about 100, while reducing the gas
pressure to 0.5 atm would increase G to ?2000
10Characteristics of typical proportional-counting
gases
11Gas Multiplication (cont.)
- Although the equation predicts that very large
gas gain factors are attainable for some
combinations of parameters, the upper G limit for
proportional gas multiplication is 104 - Above that value, space-charge effects cause G to
be less for large groups of initiating electrons
traveling together than for the few initial
electrons that might result from the passage of a
low-LET particle
12Gas Multiplication (cont.)
- It is important in proportional counters that the
gain factor be the same for all sizes of primary
ionizing events, so that the gas-amplified pulse
size will properly represent the relative
contributions of such events to the absorbed dose
in the gas
13Gas Multiplication (cont.)
- To obtain useful levels of gas gain, a
nonelectronegative gas or gas mixture must be
used, so that the free electrons do not become
attached to atoms - A few percent of a polyatomic gas such as methane
or isobutane is added to the noble gases for
proportional counting to absorb secondary UV
photons that are emitted from excited gas atoms - The energy of such photons is thus dissipated in
vibrational and rotational motion, instead of
causing new ionizing events
14Gas Multiplication (cont.)
- In G-M counters the gaseous UV absorber is
omitted because these photons are essential to
the process of propagating the discharge
throughout the tube - Certain quenching gases (the halogens Cl or Br,
or organics like ethyl alcohol) are added (5 10
) to the filling gas instead, to prevent
repeated or continuous gas discharge from
occurring
15Gas Multiplication (cont.)
- When a positive gas ion arrives at the cathode,
it is neutralized by an electron taken from that
surface - If the ionization potential of the gas is more
than twice the surface work function, there is a
chance that two electrons instead of one may be
released - Since the second electron is then free, it will
be drawn to the anode and hence will trigger
another G-M discharge - A quenching-gas molecule has a small enough
ionization potential that it can serve as the
positive charge charge carrier without releasing
more than the one electron needed to neutralize
its own charge
16Gas Multiplication (cont.)
- It should be obvious that the central wire must
serve as the anode, since otherwise the free
electrons produced by radiation in the counter
filling gas would travel outward, away from the
high-field sheath around the wire where the gas
multiplication occurs - This sheath is very thin, as can be seen from the
following considerations
17Gas Multiplication (cont.)
- At a pressure p (atm) the field strength ?(r)
must equal or exceed pK for gas multiplication to
occur - Hence the radius rs of the outer boundary of the
amplifying sheath region is -
18Gas Multiplication (cont.)
- For example in a cylindrical counter containing
P-10 gas at 1 atm, with b 10-3 cm, a 1 cm,
and P 1000 V, one has rs 3 ? 10-3 cm - The sheath thickness is therefore equal to 2 ?
10-3 cm or 20 ?m, occupying only about 0.001 of
the chambers gas volume - The probability that the radiation field will
produce primary ion pairs within the sheath
volume, thus giving rise to electron avalanches
of lesser gain, is nil
19Proportional Counters Operation
- A proportional counter is just an amplifying ion
chamber with its output measured in terms of
numbers and amplitudes of individual pulses,
instead of the charge collected - The electrometer circuit is usually replaced by a
preamplifier, a linear amplifier, and a
pulse-height analyzer, although specific
requirements of an experiment may also call for
coincidence circuits, pulse-shape discriminators,
and other pulse-processing electronics
20Operation (cont.)
- An ionizing event in the present context
includes all of the ionization that is produced
in the counter gas by the passage of a single
charged particle and its ?-rays - All of the resulting free electrons reach the
anode wire within 1 ?s - The measured electrical pulse, however, is
primarily due to the motion of positive ions away
from the wire, since they move greater distances
within the amplifying sheath than the electrons
do, on average
21Operation (cont.)
- Although the positive ions are much slower than
the electrons, they virtually all originate
simultaneously within the amplifying sheath and
move outward in unison - Since the electric field near the wire is so
strong, the positive-ion motion there gives rise
to a sharply defined fast-rising electrical pulse
that can be clipped electronically to eliminate
the later slow component contributed as the ions
progress outward toward the cathode
22Operation (cont.)
- The cloud of positive ions is small in volume,
and does not interfere with the ions resulting
from other ionizing events taking place elsewhere
in the counter - The amount of positive charge in a given ion
cloud is proportional to the number of electrons
in the associated avalanche, which in turn is
proportional to the number of ion pairs created
in the original ionizing event - Thus the size (i.e., height) of the electrical
pulse generated by the positive ions is
proportional to the energy imparted to the gas in
the initial event, provided that W/e is constant
23Operation (cont.)
- Proportional counters can operate with pulse
resolving times of about a microsecond where only
gross pulse counting is required - If pulse heights are to be measured also, the
average interval between pulses should be
greater, approaching the transit times for the
positive ions (100 ?s) for greatest accuracy - Reducing the gas gain by lowering the applied
voltage and replacing that gain by an adjustment
of the linear amplifier can provide a check on
whether the pulse-height spectrum is being
distorted by the proportional counter
24Operation (cont.)
- The following diagram shows how the pulse height
from a proportional counter (or the charge output
from an ion chamber) increases as the applied
potential P is raised - Two curves are shown, representing initial
ionizing events releasing 10 and 103 electrons - Both curves rise steeply at low voltages to reach
the ion-chamber region, in which the voltage is
great enough to closely approach complete
collection of charge without causing gas
multiplication
25Pulse height from a proportional counter as a
function of applied potential
26Operation (cont.)
- At still higher voltages the gas-multiplication
threshold is passed and the proportional-counting
region begins - The factor-of-100 difference between the two
curves in the ion-chamber region extends
throughout the proportional-counting region as
well, while G rises from 1 to 104
27Operation (cont.)
- Further increase in the applied potential results
in gain factors that are so large that
space-charge effects limit the growth of the
larger pulses, and strict proportionality of
pulse height with the number of original
electrons no longer holds - This is the region of limited proportionality
28Operation (cont.)
- Finally, at still higher voltages the two curves
merge, indicating that initiating events of
different sizes produce equal output pulses - This is the G-M region
- Increasing the voltage beyond the G-M region
results in spontaneously repeated or continuous
electrical discharge in the gas
29Proportional Counters Use with Pulse-Height
Analysis
- If the amplified output from a proportional
counter is connected to a multichannel analyzer,
the number of pulses of each height (i.e., in
each channel) can be counted to obtain a
differential distribution of counts per channel
vs. channel number, as shown in the following
diagram
30(No Transcript)
31Pulse-Height Analysis (cont.)
- To facilitate the calibration of the pulse height
h in terms of absorbed dose to the counter gas,
some proportional counters are equipped with a
small ?-particle source with a gravity-controlled
shutter - This source can send a narrow beam of ?-particles
through the counter along a known chord length ?x
32Pulse-Height Analysis (cont.)
- The expectation value of the dose contributed to
the gas by each ?-particle can be written as - where dT/?dx mass collision stopping power
of the gas for
?-particles, - ? gas density in the
counter, - ?x chord length, and
- m mass of the gas
33Pulse-Height Analysis (cont.)
- This value of dose is to be associated with the
pulse height h? at which the ?-particle peak
appears - Since the h-scale is linear vs. event size, the
dose contributions by a pulse of any size is
thereby known, assuming W/e to the the same for
all event sizes
34Pulse-Height Analysis (cont.)
- The total dose Dg in the gas, represented by the
distribution in the figure, can then be obtained
by summing over all the counts, each weighted by
its pulse height h expressed as dose - where N(h) is the distribution of counts per
channel vs. channel number h
35Pulse-Height Analysis (cont.)
- Evidently such a proportional counter can be used
as an absolute dosimeter, by virtue of the
built-in ?-source - Dg can of course be related to the dose in the
counter wall by cavity theory - The most important example of proportional
counters that are used with pulse-height
analyzers in dosimetry applications is the Rossi
counter, a commercial model of which is
illustrated in the following diagram
3613-mm-I.D. tissue-equivalent proportional counter
37Pulse-Height Analysis (cont.)
- These counters are usually made with spherical
walls of A-150 tissue-equivalent plastic, and are
operated while flowing a tissue-equivalent
counting gas through at reduced pressure,
typically 10-2 atm - Proper adjustment of the gas pressure allows
simulation of biological target objects such as
individual cells, in terms of the energy lost by
a charged particle in crossing it - This is the primary experimental instrument used
in microdosimetry
38Proportional Counters Applications without Pulse
Height Analysis
- Proportional counters of various designs are also
used for many applications in which pulse-height
analysis is not used - The main advantages of proportional counters over
G-M counters in this connection are - their short pulse length (1 ?s) with practically
no additional dead time, accommodating high count
rates, and - the capability of discriminating by simple means
against counting small pulses that might result,
for example, from background noise, or ?-ray
interactions in a mixed ? neutron field
39Geiger-Müller Counters Operation
- The gas-amplified pulses in a G-M counter come
out approximately the same, regardless of the
size of the initiating event - If the resulting pulse size is larger than the
counter-circuit threshold ht, then the pulses
will be counted if they are too small, they will
not - As a result, since the pulse size gradually
increases as a function of applied potential, one
would expect to see a step function in the
count-rate-vs.-voltage curve where the pulse
height begins to exceed ht
40Operation (cont.)
- The step is actually S-shaped, due to the
Gaussian distribution of pulse sizes produced in
the counter even under ideal G-M conditions - At the applied potential P1 no counts are
obtained - At P2 the pulses in the ideal G-M counter are all
larger than ht, as shown by the solid curve inset
in the following figure - In that case a flat plateau would be observed, as
indicated by the solid curve in the graph of
count rate vs. P
41The counting plateau in a G-M tube. The solid
curve is an ideal G-M plateau that would be
seen for a narrow distribution of pulse heights.
The dashed curve has a residual slope within the
G-M region because of the presence of a
low-amplitude tail on the pulse-height
distribution (see inset).
42Operation (cont.)
- In actual G-M tubes there is a residual slope in
the plateau region, as shown by the dashed curve
in the diagram - This is caused by a small-pulse tail on the
Gaussian distribution of pulse heights, as
indicated by the dashed curve in the inset of the
diagram - These small pulses are mostly produced by the
ionizing events that occur during the period
before the G-M tube has fully recovered from the
preceding discharge
43Geiger-Müller CountersDead Time
- Immediately after a discharge the positive space
charge so weakens the electric field near the
wire that gas multiplication cannot occur - Thus the tube does not respond to radiation at
all until the positive-ion cloud starts arriving
at the cathode and the electric field strength
gradually builds up again - As that takes place, the tube becomes capable of
responding to an ionizing event with a discharge
of less than full size
44Dead Time (cont.)
- The true dead time is the time from the start of
the preceding pulse until the tube recovers to
the point where a minimum-sized pulse can be
generated - The recovery time is the time until a full-sized
pulse is again possible, as shown in the
following diagram - The threshold peak size ht necessary for counting
by a G-M counter circuit is considerably less
than the average pulse size that would be
generated by a fully recovered G-M tube when it
is being normally operated at a potential in the
middle of the G-M plateau
45Dead time and recovery time of a G-M tube
46Dead Time (cont.)
- The minimum time between detectable pulses will
be less than the recovery time - This is the pulse resolving time, but is more
commonly referred to as the dead time in place
of the narrower definition above
47Dead Time (cont.)
- If an ionizing event occurs during the true dead
time, it causes no electron avalanche and hence
has no effect on the tube - This is called nonparalyzable dead-time behavior
- If an ionizing event occurs after the end of the
true dead time, but before the resulting pulse is
large enough to be counted (i.e., gt ht), not only
will that event go uncounted but a new dead-time
period will begin - This is called paralyzable dead-time behavior
48Dead Time (cont.)
- Obviously, a G-M counter exhibits dead-time
behavior that is intermediate, being a mixture of
the paralyzable and nonparalyzable cases - Reducing the detectable pulse-height threshold ht
tends to decrease the paralyzable component
49Dead Time (cont.)
- If m is the observed count rate, n is the true
count rate, and ? is the pulse resolving time,
then for the nonparalyzable case the correction
for dead-time counting losses is - and for the paralyzable case
- which must be solved for n iteratively
- In the limiting case of small dead-time counting
losses, (n ltlt 1/?), both types reduce to -
50Dead Time (cont.)
- At high values of the true count rate n, the
value of m approaches 1/? asymptotically in a
nonparalyzable counter - However, in the paralyzable case m reaches a flat
maximum at n 1/?, then gradually decreases with
further increases in n because of overlapping
chains of dead-time periods - Thus a low reading of a G-M counter may result
from a strong radiation field if the dead-time
behavior is predominantly paralyzable
51Geiger-Müller Counters Applications
- Since G-M counters are only triggered by ionizing
events, producing discharge pulses of more or
less the same size regardless of the initiating
event, the observed output conveys little
information about the dose to the counter gas - Nevertheless G-M counters are used in some
dosimetry applications because they offer several
advantages
52Applications (cont.)
- They require little if any further amplification,
since pulses of 1 10 V can be obtained directly - They are also inexpensive and versatile in their
construction and geometry - Thus they are often used in radiation survey
meters to measure x- and ?-ray fields in
radiation protection applications - When equipped with a thin ( 1 mg/cm2) window
they can also be used to detect ?-rays
53Applications (cont.)
- Scale calibrations of G-M counters, if given in
terms other than the count rate, should always be
suspect because of the lack of dose response - Because most G-M tubes are constructed of
materials that are higher in atomic number than
tissue or air, they exhibit strong
photoelectric-effect response below 100 keV - Enclosing the G-M tube in a suitable high-Z
filter tends to flatten the overresponse at low
energies
54Typical energy-dependence curves of the response
relative to tissue dose for survey meters
containing G-M tubes of the following types (a)
uncompensated, and (b) compensated with metallic
filters to produce flatter response.