Title: The BRITE Instrument
1The BRITE Instrument Optics, Filters, Detector
Stefan Mochnacki Dept. of Astronomy
Astrophysics University of Toronto (Canada Day,
2008)?
2Performance RequirementsFilters and Passbands
- Two Austrian and two Canadian BRITES two with
Blue filter, two with Red filter. - Passbands chosen to allow for mode separation and
equal average weighting of signal (hence B filter
narrower than R). - Interference filters chosen for well-defined
passbands. - Direct measurement of delivered filters being
done with DDO spectrograph (Pribulla).
3(relative efficiency)?
4Performance RequirementsPhotometric Accuracy
- Stars with V3.5 will be observed in
differential mode. - Each measurement (defined as up to 15 minutes of
integration within a single orbit) shall be
better than 0.001 mag. - The error amplitude spectrum in photometric
measurements for periods longer than a month
shall be less than 20 ppm.
5S/N, Exposure time Calculations
- Simple spreadsheet used to compute S/N given
basic parameters and assumptions, at 5500 A for
1500 A bandpass. - Only noise sources source photon noise (vN),
sky background 18 mag/sq), dark current (5 e- _at_
20 C), readout noise (13 e-). - Gaussian PSF (6 pixels), integrating aperture
dia. 2 x FWHM. Approx. 110 pixels in
integrating aperture. - Plate scale 26.56/pixel. PSF 3', ap. dia.
6'. - Bright star limit set by saturation of central
pixel. - 400 exps. co-added (0.5 1.4 )s x 400 13
min. - No real penalty on brighter stars due to
multi-exp.
6(More accurate estimates have been computed by
A.Kaiser)?
7Telescope
- We chose a 30mm aperture, 70mm focal length,
which with a 35mm-format sensor 9µ pixels has an
image scale of 26.6 arcsec/pix. - The field of view is between 22 and 25 degrees.
- The lens design was driven by the need for
adequately sampled images and good baffling,
which also yielded an image-space telecentric
telescope.
8Optical Design (Blue)?
9Red Brite spot matrix
10Blue BRITE Point Spread Function
11 Blue BRITE design PSF simulation by Rainer
Kuschnig- approximate simulation - Zemax PSF
data resampled/binned to the CCD frame
BRITE Blue PSF _at_ 8.5 deg
Zemax PSF
Desired Gaussian PSF
12BRITE Telescope Side View
13BRITE Optical Cell
14Instrument Baffle and Pupil Stop
15Angled View (Materials)?
Aluminum
Magnesium
Aluminum
Stainless Steel
16Telescope and Startracker Assembly
CCD Header Tray and Enclosure
17BRITE Instrument
- Photometry
- Differential photometry with 0.1 precision.
- Error amplitude spectrum lt20 ppm, gt 1 month.
- Timing
- Exposure times 0.1-100s, known to 0.01
- Absolute time accuracy better than 0.1s.
- Optics
- Point Spread Function Gaussian, 5-7p FWHM.
- No vignetting, telecentric, minimum ghosting.
- Filter either 390-460nm or 550-700nm.
- 3 cm aperture telescope, 24 degrees FOV.
- Detector
- Detector temperature low, measured to 0.1ºC.
- SNR 1000 per 100s exposure at V3.5
- Design out sun stare risk, no shutter or door.
- Stray Light Baffle and light-tight instrument.
18Front and Side Views of Instrument in the BRITE
Satellite
19The BRITE Satellite
20Choice of Detector
- After a CMOS effort, we chose the Kodak KAI-11002
CCD Interline transfer, microlenses. - Very low dark current at room temperature (no
active cooling on BRITE), avg. 5e-/s/pix. - Low readout noise 13e-/pix possible, BRITE
prototype yields 25e-/pix. - Good quantum efficiency 50
- Fast readout 12 MHz
21CCD Sensor
Kodak KAI-11002
22Quantum Efficiency Curve for BRITE CCD
23- Issues affecting sampling of KAI-11002M CCD
- It is an INTERLINE TRANSFER device, which is
essential for a CCD without a mechanical shutter. - It uses microlenses to compensate for the 70
dead space. The peak quantum efficiency goes
from 16 to 50 when microlenses are fitted,
suggesting there should be little effective dead
space. - This means that microlenses probably eliminate
most of the dead space as far as its
contribution to undersampling is concerned. Even
back-illuminated CCDs have intra-pixel
sensitivity variation. - The angular response is quite good lt 12 degrees.
24Variation of Quantum Efficiency with Angle of
Incidence in KAI-11002M (Kodak data).
25- Prototype camera using KAI-11002 chip has been
built and is used extensively for testing. - Most important are tests on real star fields and
on simulated star fields to measure PSFs and
sampling. - The dark current distribution is very promising
if longer exposures are desired. - We have adjustable bias level and gain. At
present bias 100 ADU, gain 3e-/ADU. The gain
is shown in photon transfer curves
26Saturation effect
Slope 0.33 gt gain3 e/ADU
27Another transfer curve (unsaturated)?
28Typical light distribution of target imaged for
transfer curve measurements
2960 second dark, bias subtracted. Note spikes,
typical of Kodak CCDs.
30Dark current from 10 second exposure.
31Dark current for 60 sec. integration
32A blow-up of lower dark current levels (60 second
integration).
33Same thing, linear count scale. 10 ADU bins.
34Breakdown of imaging for T120 s.
35Conclusions re Dark Current
- In a 60 second exposure, we see less than 5
e-/p/s in 99 of pixels. The hot pixels appear
always to be the same ones. This will be studied
in greater detail. - The device begins to break down at exposure times
gt 60 seconds at 20C. - Temperature dependence of dark current and
breakdown could constrain temperature regulation
by heating.
36Real and Simulated Star Fields
- Real star fields observed without tracking, 1
second exposures. These were made to focus the
prototype and examine PSFs. - Some star images at different positions are shown
on the next slide. - Artificial star fields in the lab used to compare
photometry when small movements of the camera are
made. - All testing is now being done with the BRITE
prototype lens (vibration test model) and
electronics.
37Prototype Blue BRITE and electronics
38(No Transcript)
39Real star images
40Central image, -0.125 mm out of focus.
41Artificial star field, in focus
42Artificial field, 0.5mm out of focus
43Test results
- Simulated star field tests show some
undersampling errors when the BRITE lens is
sharply focused, requiring remedial action (Alex
Kaiser will discuss). Fully sampled images have
shown no aliasing in previous testing. - A wide range of focus positions is being tested.
- Real star field tests show the variation of the
PSF over the whole field, and test the instrument
sensitivity.