Title: Integrating Dosimeters III
1Integrating Dosimeters III
2Calorimetric Dosimetry
- The measurement of the temperature rise in
calorimetric dosimeters comes closest of any
method to providing a direct measurement of the
full energy imparted to matter by radiation - Only relatively small corrections for thermal
leakage and for chemical reactions are necessary
3Temperature Measurement
- In principle any kind of thermometer can be
applied in a calorimeter if the temperature
change is large enough to measure with sufficient
accuracy and precision - In practice only thermocouples and thermistors
are sufficiently sensitive and small thermistors
are usually preferable because of their greater
sensitivity
4Temperature Measurement (cont.)
- The temperature increase per unit of absorbed
dose to the material in the calorimeters
sensitive volume depends on its thermal capacity,
which is usually expressed in cal/g C or J/kg C - The exact value of the calorie (i.e, the energy
required to raise 1 g of water 1 C) depends upon
the temperature of the water to which it refers - Usually thermal-capacity (or specific-heat)
tables assume the value of the calorie for water
at 15 C hence 1 cal 4.185 J, and 1 cal/g C
4185 J/kg C
5Temperature Measurement (cont.)
- For a sensitive volume containing a material of
thermal capacity h (J/kg ?C), mass m (kg), and
thermal defect ?, and that absorbs E joules of
energy, the temperature increase is given by - where D is the average absorbed dose (Gy) in
the sensitive volume
6Temperature Measurement (cont.)
- Thus a measurement of D does not require
explicit knowledge of m if h is known - The thermal defect ? is the fraction of E that
does not appear as heat, due to competing
chemical reactions, if any - ? is negative for exothermic reactions
- A few typical values of h are given in the
following table
7Thermal capacity of several calorimetric materials
8Temperature Measurement (cont.)
- For example, in Al a dose of 1 Gy causes a
temperature increase of 1.12 ? 10-3 ?C - To measure this temperature rise with 1
precision would require a thermometer capable of
detecting temperature changes of the order of 10
??C
9Temperature Measurement Thermocouples
- Thermocouples typically have temperature
coefficients of 40 70 ?V/?C - A temperature change of 10 ??C would then give a
potential change of (4 7) ? 10-10 V - This is too small to detect with available
instruments, such as a nanovoltmeter - Increasing the dose to 100 Gy would cause a
temperature rise of 0.112 ?C, requiring detection
of (4 7) ? 10-8 V for 1 precision, which can
be accomplished with a nanovoltmeter
10Thermocouples (cont.)
- Thermocouples are generally found to be most
useful in calorimeters where large doses (gt 10
Gy) are given, usually in a short enough time
period for thermal leakage to be negligible
(i.e., under adiabatic conditions) - Thermocouple sensitivity can be multiplied by
constructing a thermopile, consisting of a number
of thermocouples in series, but this is usually
not practical for calorimetric dosimetry because
of the increase in perturbation of the medium and
the number of thermal leakage paths
11Temperature Measurement Thermistors
- Thermistors can be obtained in sized comparable
to thermocouples - They are semiconductors made of metallic oxides
and other constituents that are usually not
specified by the manufacturer - They exhibit negative temperature coefficients of
the order of several percent per C at room
temperatures, increasing in negative coefficient
with decreasing temperature, as shown in the
following diagram
12A typical resistance-vs.-temperature curve for a
thermistor. The slope is the temperature
coefficient of resistance at any temperature.
For example, the slope dR/dT at 20 C, shown by
the dashed line, is ?-120 ?/?C or 3.7/?C.
13Thermistors (cont.)
- The resistance of a thermistor at room
temperature is typically 103 105 ?, which can
be conveniently measured with great precision and
accuracy by a Wheatstone bridge as shown in the
following diagram - The bridge null detector must be sensitive enough
so that the power dissipated in the thermistor is
negligible compared to the radiation heating
14A Wheatstone bridge circuit for measuring the
resistance of a thermistor in the sensitive
volume (or core) of a calorimetric dosimeter.
When Rx is set to produce a null current
reading,Rc/Rx R1/RJ, from which Rc can be
determined.
15Calorimeter Design
- We will consider three general types of
radiometric calorimeter designs, depending on
whether absorbed dose in a reference medium,
energy fluence in a radiation beam, or power
output of a radioactive source is to be measured
16Calorimeter Design Absorbed-Dose Calorimeters
- An absorbed-dose calorimeter must have a
sensitive volume that is small compared to the
penetrating ability of the radiation and is
thermally insulated from its surroundings to the
extent necessary to attain acceptably small
levels of thermal leakage - The sensitive volume (often called the core) is
made of a thermally conductive material identical
to, or simulating, a medium of dosimetric
interest (e.g., graphite, tissue-equivalent
plastic, or silicon), and contains a temperature
sensor of negligible mass, usually a thermistor
17Absorbed-Dose Calorimeters (cont.)
- The core is surrounded by a shell (jacket or
mantle) of the same material to provide
charged-particle as well as thermal equilibrium - If the thermal capacity and thermal defect of the
core material are known, if the temperature
sensor is correct, and if the thermal leakage is
negligible, then the calorimeter can be operated
adiabatically, without any energy calibration, to
measure the average absorbed dose in the core by
application of - The mass of the core need not be known in that
case
18Absorbed-Dose Calorimeters (cont.)
- An example of such a calorimetric dosimeter that
does not require calibration is shown in the
following diagram - This was designed to measure the dose deposited
in silicon by intense single pulses of
penetrating x rays - The core was a pea-sized sphere of high-purity
silicon of known thermal capacity, enclosed in an
equilibrium-thickness shell of the same material,
which was insulated from the surrounding air by a
shell of styrofoam
19An absolute adiabatic silicon calorimetric
dosimeter of simple spherical design for
measuring the absorbed dose to silicon from
intense single pulses of penetrating x-rays
20Photograph of the complete dosimeter, with scale
in inches
21Absorbed-Dose Calorimeters (cont.)
- The core, containing a very small calibrated
thermistor, was centered in a spherical void by
four conical points of silicon projecting from
the inside of the shell - Since the radiation pulse raises the temperature
of the shell almost instantaneously ( 10-7 s) by
the same amount as the core, the heat-loss rate
from the core was found to be initially
negligible even with evacuating the surrounding
gap, thus permitting the very simple design shown
22Absorbed-Dose Calorimeters (cont.)
- Arguably, to avoid the complication of including
an ohmic heater in the core and the need to know
the mass of the core, one should build the core
and surrounding shell from a material that has an
accurately known thermal capacity - Classical methods are available for measuring the
thermal capacity of any given sample of material
before it is incorporated into a calorimetric
dosimeter
23Absorbed-Dose Calorimeters (cont.)
- The conventional practice has been to design an
ohmic heater into the core, which can then be
made of a material (or materials) for which the
thermal capacity is only approximately known in
advance - Electrical energy E results in a temperature rise
?T in the core, thus allowing the average value
of h for the materials making up the core to be
determined if its total mass is known
24Absorbed-Dose Calorimeters (cont.)
- The following diagram is a schematic
representation of this kind of absorbed dose
calorimeter - The core is surrounded by two or more thermally
insulated layers of the same material as the
core, and the entire assembly is usually
surrounded by a constant-temperature environment - Each layer may contain a thermistor and/or an
ohmic heater, thus allowing measurement and
control of the temperature environment of the core
25Schematic arrangement of a typical absorbed-dose
calorimeter containing several concentric thermal
bodies. Each may include thermistors and/or
ohmic heaters for temperature measurement and
control
26Absorbed-Dose Calorimeters (cont.)
- Dosimetry by means of such an apparatus is
generally complicated and time-consuming - Some simplifications have been devised to shorten
the time necessary to reach thermal equilibrium
before an exposure run and automatically correct
for heat leakage from the core to the jacket - A modification in the Wheatstone bridge results
in the measurement of the ?T that would have
occurred in the core due to electrical heating if
none of the heat had leaked into the jacket
27Modified Wheatstone bridge for measuring the core
jacket temperature rise, assuming that the two
thermistors are identical in characteristics, and
that the jacket has the same mass and is made of
the same material as the core
28Absorbed-Dose Calorimeters (cont.)
- This requires that the jacket have the same mass
and be made of the same material as the core, and
that their two thermistors must be virtually
identical with respect to dR/dT as well as
resistance - When the bridge is initially balanced,
- where RJ is the resistance of the thermistor
in the jacket, and the other resistances are
identified as in the earlier Wheatstone bridge
schematic
29Absorbed-Dose Calorimeters (cont.)
- After electrical heating of the core and
readjusting Rx by the small amount ?Rx necessary
to balance the bridge again - Assuming that R1 Rx RJ Rc, we have
- in which for small changes ?RJ and ?Rc, the
last term is vanishingly small
30Absorbed-Dose Calorimeters (cont.)
- This equation means that ?Rx has the same value
whether all the electrical energy stays in the
core or some leaks to the surrounding jacket - Thus heat leakage out of the core during the
electrical calibration is automatically corrected
for, provided that thermal leakage from the
jacket into the shield is negligible - This can be assured by feedback control of the
shield to maintain it at the same temperature as
the jacket
31Absorbed-Dose Calibration (cont.)
- Before exposing the calorimeter to radiation, the
Wheatstone bridge circuit is to be switched back
to the original circuit, with RJ RJ - If the radiation is sufficiently penetrating, the
core, jacket, and shield will be heated equally
by the uniform dose - Supplementary electrical energy can be supplied
to the shield to compensate for its thermal
losses - Heat losses from the core can again be made
negligible, so that the ?T measured in the core
can be correctly interpreted in terms of the core
jacket ?T that was observed during the
electrical calibration
32Calorimeter Design Energy-Fluence Calorimeters
- An energy-fluence calorimeter contains a core,
usually consisting of a cylindrical piece of
dense material such as lead or gold, large enough
to stop an incident beam of radiation - The geometry is shown schematically in the
following diagram
33Schematic arrangement of an energy-fluence
calorimeter
34Energy-Fluence Calorimeters (cont.)
- The core is suspended by nylon strings in an
insulated vacuum chamber, sometimes adjacent to a
twin core that serves as a control to determine
thermal leakage - Because of the size of the core, more than one
thermistor may be necessary to sample the
temperature adequately, and the heater should be
designed to distribute the heat uniformly - The high-Z core may require a significant
backscattering correction
35Energy-Fluence Calorimeters (cont.)
- h can be determined through electrical
calibration as discussed before - The energy fluence of the radiation beam passing
through the aperture of area A is given
(neglecting ?) by
36Calorimeter Design Power Output of a Radioactive
Source
- A power-output calorimeter has a cup-shaped core
into which a radioactive source can be inserted
for measurement - The walls of the core are made thick enough to
stop all the radiation to be measured from
escaping - The usual method for electrical calibration can
be used to determine h
37Power Output (cont.)
- Then the power output (W) is given by
- where temperature rise ?T (?C) occurs in the
core of mass m (kg) during the time interval ?t
(s), and ? has been neglected
38Advantages of Calorimeters
- They can be made absolute, either intrinsically
or by means of electrical-heating calibration. - The measurement of temperature rise comes closest
of any dosimetric technique to being a direct
measurement of the energy involved in the
absorbed dose. Only relatively small exothermic
or endothermic chemical reactions, and thermal
leakage, must be corrected for, and these are
often negligible.
39Advantages (cont.)
- Almost any absorbing material, solid or liquid,
can be employed in the calorimeter sensitive
volume, so long as it is reasonably conductive
thermally and has a known thermal defect - Calorimeters are inherently dose-rate-independent
under adiabatic conditions, and become more
convenient to use as the dose-rate increases
because thermal leakage during dose delivery
becomes negligible. At high dose rates, where
other dosimeters show saturation effects,
calorimeters are at their best.
40Advantages (cont.)
- Calorimeters add up the energy contributions in
the sensitive volume from different types of
radiations (e.g., neutrons and ? rays) with
weighting factors of unity, neglecting
differences in thermal defect. - Calorimeters have no LET dependence (neglecting
minor differences in thermal defect, if any),
since ionic recombination is irrelevant to the
temperature rise.
41Advantages (cont.)
- Calorimeters are relatively stable against
radiation damage at high doses the thermistor
(if used as the temperature sensor) is usually
the limiting factor in this respect.
42Disadvantages of Calorimeters
- Temperature rises to be measured are typically
very small, usually only a minute fraction of a
degree, which limits calorimetry to relatively
large doses. - Thermal insulation, and instrumentation for
thermal control and measurement, often make the
calorimeter apparatus bulky and difficult to
transport and set up. This limits the kinds of
situations to which calorimetry is usually
applied to calibration of other dosimeters.
43Disadvantages (cont.)
- For low dose rates, thermal leakage in and out of
the calorimetric sensitive volume limits the
accuracy and precision available. - Some materials undergo radiation-induced
endothermic or exothermic reactions which cause a
difference between the integral dose and the
energy available to heat the sensitive volume.
In A150-type TE plastic about 4 of the absorbed
dose goes to an endothermic reaction instead of
heat.
44Conclusions
- The present brief introduction to calorimetric
dosimetry deliberately errs on the side of
simplicity, for two reasons - In the simplest applications at high dose or
fluence rates, especially for intense single
pulses, adiabatic operation is well approximated,
?T is large, and calorimetry becomes the method
of choice - In the difficult applications of calorimetry
(i.e., at low dose rates), the problems of heat
leakage and temperature drift are too complicated
to be dealt with adequately in an introductory
course