Status of Oxford Setup - PowerPoint PPT Presentation

1 / 16
About This Presentation
Title:

Status of Oxford Setup

Description:

N.B. Picture taken straight from the oscilloscope. 6. Measurements: X-ray signal ... Illuminated CPC2 with Fe55 X-ray source (5.9 keV) Signal : 15-20 mV useful ... – PowerPoint PPT presentation

Number of Views:13
Avg rating:3.0/5.0
Slides: 17
Provided by: Phys157
Category:
Tags: oxford | setup | status

less

Transcript and Presenter's Notes

Title: Status of Oxford Setup


1
Status of Oxford Setup
  • Matthew Chalk, Erik Devetak, Johan Fopma, Brian
    Hawes, Ben Jeffery, Nikhil Kundu, Andrei
    Nomerotski
  • University of Oxford ( 18 August 2006 )

2
Oxford Test Setup
  • Goals
  • Fully test and exercise produced electronics,
    should improve understanding, quality and
    feedback
  • Analogue readout of CPC2
  • Perform additional CPC2 measurements, build up
    CCD expertise
  • Experiment with power provisions for the clock
    driver with real sensor and super-capacitors

3
Oxford Test Setup
Light version of the RAL setup
4
Status
  • Minimal configuration is functioning
  • Sequencer based on BVM2
  • Agilent LV PS
  • Oxford Bias PS controlled by BVM1
  • Agilent mixed-signal Oscilloscope used as ADC
  • LabView code was mostly inspired by RAL examples
  • Many thanks to Konstantin for help
  • Main parameters
  • Sequencer Master Clock 1-10MHz
  • Integration time (time between triggers) 0.003 -
    0.5 sec
  • Graph RG, CCD Clock and CCD analogue readout

5
Measurements X-ray signal
  • Standard procedure to enhance probability to
    have a Fe55 signal need to integrate long time ?
    need to operate at low temperature
  • Our setup is running so far at room temperature ?
    accumulate substantial noise from leakage current
  • Reduced integration time to minimum (3 msec) and
    used accumulation feature of the scope

Background noise
Signal (approx. 15mV below the noise )
N.B. Picture taken straight from the oscilloscope
6
Measurements X-ray signal
  • RMS of Raw Noise 2-4 mV
  • Illuminated CPC2 with Fe55 X-ray source (5.9 keV)
  • Signal 15-20 mV ? useful calibration
  • 18 mV expected based on expected amplification
    etc
  • 1600 e from Fe55 ? 100 e / mV
  • Noise 300 e (to compare to 50-100 e at RAL
    ways to go!)
  • Noise performance is poor as expected at room
    temperature and long integration time
  • Had a setback using the ICS-554 PCI ADC
  • Double PCI-PMC carrier does not fit in our PC ? ?
    ?
  • Will exchange for a single carrier, here in a
    month
  • Potentially ICS-554 will allow fast readout
    (faster than VME), it also has onboard large FPGA
    for pre-processing so further speed-up is
    possible (for ex setting a threshold and passing
    on to the PCI bus only events above the
    threshold)
  • In the meantime tried to detect signal using
    scope trace readout to PC but this was too slow
  • Concentrated on noise measurements

7
Noise Analysis
  • Examples on noise graphs
  • Different pixels have different offsets
    (pedestals)
  • Software subtracts pedestals so many rows can be
    added up (right bottom graph)
  • Almost no tails in noise distribution ? good fit
    by single Gaussian

8
Noise vs Integration Time
  • Naively expect that leakage current is main noise
    contributor
  • Ncarriers Tintegration
  • Noise sqrt( Tintegration)
  • Trigger rate effectively determines integration
    time
  • See some dependence above 100 msec, no dependence
    below
  • It looks that at smaller integration time noise
    is not caused by the leakage current ? room for
    further improvement?

9
Noise vs Master Clock Frequency
  • Higher frequency effectively decreases
    integration time so this qualitatively agrees
    with previous result

10
Noise analysis
  • To explain better what we measure
  • Scope traces include 50 consecutive pixels (so
    50 rows of CCD)
  • These 50 rows can be sampled at any location of
    CPC2
  • Simultaneously 3 CCD columns are measured, see
    graph
  • Amplitude for each pixel is measured at a certain
    delay wrt trigger
  • Can study noise correlations between CCD rows and
    columns

11
Noise correlations
  • Dnoise standard deviation of amplitude
    difference for two channels
  • If no correlation expect Dnoise to increase by
    sqrt(2) wrt noise of single channel
  • If absolute correlation expect Dnoise to be 0
  • If absolute anti-correlation expect Dnoise to be
    doubled
  • Can study
  • Column Dnoise expect correlation as readings
    are taken simultaneously
  • Row Dnoise expect less or no correlation as
    rows are read out at different time

12
Column Dnoise
  • Took data for 12 locations 4x3 columns with
    analogue outputs times 3 row regions
  • Adjacent column Dnoise Aij Ai1j
  • Noise 2 mV
  • Assuming no correlation between pixels ?
  • 2mV/1.4100e/mV 140 e
  • Uniform column Dnoise across CCD

13
Column Dnoise
  • Correlation of far away columns
  • Far away column Dnoise Aij AiNj
  • Less correlation with far away columns

14
Row Dnoise
  • Adjacent row Dnoise Aij Aij1

15
Row Dnoise
  • Far away row Dnoise Aij AijN
  • Limited by 50 pixel readings (scope)
  • Results suggest anti-correlation
  • Need further study a bit controversial
    measurement

16
Summary
  • Oxford setup is up and running
  • So far studied CPC2 noise at room temperature
  • Setup was very useful for MB4.3 debugging
    hopefully goes to RAL in better shape than
    previous versions
  • Plans
  • Lower temperature
  • Set up proper ADC readout
  • Study signal response ? gain calibration etc
  • Set up charge injection mode
Write a Comment
User Comments (0)
About PowerShow.com