Magnetic Resonance Imaging II - PowerPoint PPT Presentation

1 / 42
About This Presentation
Title:

Magnetic Resonance Imaging II

Description:

... planes can be obtained by energizing the appropriate gradient coils during image ... depends on a combination of two or more coils energized simultaneously ... – PowerPoint PPT presentation

Number of Views:62
Avg rating:3.0/5.0
Slides: 43
Provided by: michae354
Category:

less

Transcript and Presenter's Notes

Title: Magnetic Resonance Imaging II


1
Magnetic Resonance Imaging II
  • K-space data acquisition and image
    reconstruction

2
K-space matrix
  • MR data are initially stored in the k-space
    matrix, the frequency domain repository
  • The axes have units of cycles/unit distance
  • Each axis is symmetric about the center of
    k-space, ranging from fmax to fmax
  • Low-frequency signals are mapped around the
    origin of k-space and high-frequency signals are
    mapped further from the origin in the periphery

3
(No Transcript)
4
K-space matrix (cont.)
  • Frequency domain data are encoded in the kx
    direction by the frequency encode gradient (FEG),
    and in the ky direction by the phase encode
    gradient (PEG) in most image sequences
  • Lowest spatial frequency increment is the
    bandwidth across each pixel

5
(No Transcript)
6
Two-dimensional data acquisition
  • MR data acquired as a complex, composite
    frequency waveform
  • With methodical variations of the PEG during each
    acquisition, the k-space matrix is filled to
    produce the desired variations across the
    frequency and phase encode directions

7
(No Transcript)
8
  • Narrow band RF excitation pulse applied
    simultaneously with the slice select gradient
    (SSG)
  • Energy absorption dependent on amplitude and
    duration of the RF pulse at resonance
  • Longitudinal magnetization converted to
    transverse magnetization, extent of which depends
    on saturation of spins and angle of excitation
  • 90-degree flip angle produces maximal transverse
    magnetization

9
(No Transcript)
10
  • Phase encode gradient (PEG) applied for a brief
    duration to create phase difference among spins
    along the phase encode direction
  • Produces several views of the data along the ky
    axis, corresponding to the strength of the PEG
  • Refocusing 180-degree pulse delivered after
    selectable delay time, TE/2
  • Inverts direction of individual spins and
    reestablishes phase coherence of transverse
    magnetization with formation of echo at time TE

11
(No Transcript)
12
  • During echo formation and subsequent decay,
    frequency encode gradient (FEG) applied
    orthogonal to both slice select and phase encode
    gradient directions
  • Encodes precessional frequencies spatially along
    the readout gradient

13
(No Transcript)
14
  • Simultaneous to application of FEG and echo
    formation, computer acquires the time-domain
    signal using an analog-to-digital converter (ADC)
  • Sampling rate determined by excitation bandwidth
  • One-dimensional Fourier transform converts the
    digital data into discrete frequency values and
    corresponding amplitudes
  • Proton precessional frequencies determine
    position along the kx (readout) direction

15
(No Transcript)
16
  • Data deposited in k-space matrix in a row,
    specifically determined by the PEG strength
    applied during excitation
  • Incremental variation of the PEG throughout the
    acquisition fills the matrix one row at a time
  • Possible to acquire the phase encode data in
    nonsequential order to fill portions of k-space
    more pertinent to the requirements of the exam
    (e.g., in low-frequency, central area of k-space)
  • After matrix is filled, columns contain
    positionally dependent phase change variations
    along the ky (phase encode) direction

17
(No Transcript)
18
  • Two-dimensional inverse Fourier transform decodes
    the frequency domain information piecewise along
    the rows and then along the columns of k-space
  • Final image is a spatial representation of the
    proton density, T1, T2, and flow characteristics
    of the tissue using a gray-scale range
  • Each pixel represents a voxel thickness
    determined by slice select gradient strength and
    RF frequency bandwidth

19
K-space segmentation
  • Central area of k-space (lower spatial
    frequencies) represents the bulk of the
    anatomical information
  • Information near the center of k-space provides
    the large area contrast in the image
  • Outer areas in k-space contribute to the
    resolution and detail

20
(No Transcript)
21
2D multiplanar acquisition
  • Direct axial, coronal, sagittal, or oblique
    planes can be obtained by energizing the
    appropriate gradient coils during image
    acquisition
  • The slice select gradient determines the
    orientation of the slices
  • Axial uses the z-axis coils
  • Coronal uses the y-axis coils
  • Sagittal uses the x-axis coils
  • Oblique plane acquisition depends on a
    combination of two or more coils energized
    simultaneously

22
(No Transcript)
23
Acquisition time
  • Time required to acquire an image is equal to
  • Spin echo sequence for a 256 x 192 image matrix
    and two averages per phase encode step with TR
    600 msec yields an imaging time of 3.84 minutes
  • Where nonsquare pixels are used, the phase encode
    direction is typically placed along the small
    dimension of the image matrix

24
Multislice data acquisition
  • Average time per slice significantly reduced
    using multiple slice acquisition methods
  • Several slices within the tissue volume
    selectively excited during a TR interval to fully
    utilize the (dead) time waiting for longitudinal
    recovery in a specific slice
  • This requires cycling all of the gradients and
    tuning the RF excitation pulse many times during
    the TR interval

25
(No Transcript)
26
Multislice data acquisition (cont.)
  • Total number of slices is a function of TR, TE,
    and machine limitations
  • C is a constant dependent on MR equipment
    capabilities (computer speed, gradient
    capabilities, RF cycling, etc.)
  • Long TR acquisitions such as proton density and
    T2-weighted sequences provide greater number of
    slices than T1-weighted sequences with short TR

27
Data synthesis
  • Data synthesis takes advantage of the symmetry
    and redundant characteristics of the frequency
    domain signals in k-space
  • Acquisition of as little as one-half the data
    plus one row of k-space allows the mirroring of
    complex conjugate data to fill the remainder of
    the matrix
  • Reduces the acquisition time by nearly one-half
  • Penalty is a reduction in the SNR and potential
    for artifacts

28
(No Transcript)
29
Fast spin echo acquisition
  • FSE technique uses multiple phase encode steps in
    conjunction with multiple 180-degree refocusing
    RF pulses per TR interval to produce a train of
    up to 16 echoes
  • Each echo experiences differing amounts of phase
    encoding that correspond to different lines in
    k-space
  • Points near the center of k-space acquired from
    first echoes for best SNR

30
(No Transcript)
31
Inversion recovery acquisition
  • Phase and frequency encoding occur similarly to
    the spin echo pulse sequence
  • Since TR is typically long, several slices can be
    acquired within the volume of interest
  • STIR (short tau inversion recovery) and FLAIR
    (fluid attenuated inversion recovery) pulse
    sequences are commonly used

32
(No Transcript)
33
Gradient recalled echo acquisition
  • Similar to standard spin echo sequence with a
    readout gradient reversal substituting for the
    180-degree pulse
  • With small flip angles and gradient reversals,
    considerable reduction in TR and TE is possible
    for fast image acquisition ability to acquire
    multiple slices is compromised
  • PEG rewinder pulse of opposite polarity applied
    to maintain phase relationship from pulse to pulse

34
(No Transcript)
35
GRE acquisition (cont.)
  • Gradient echo image acquisition time given by
    same expression as for spin echo technique
  • Gradient-echo sequence for a 256 x 192 image
    matrix, two averages, and TR 30 msec, imaging
    time is approximately 15.5 seconds
  • Compromises include SNR losses, magnetic
    susceptibility artifacts, and less immunity from
    magnetic field inhomogeneities

36
Echo planar image acquisition
  • EPI acquisition provides extremely fast imaging
    time
  • For single-shot EPI, image acquisition typically
    begins with a standard 90-degree flip, then a
    PEG/FEG gradient application to initiate
    acquisition of data in periphery of k-space,
    followed by a 180-degree echo-producing pulse
  • Immediately after, an oscillating readout
    gradient and phase encode gradient blips are
    continuously applied to stimulate echo formation
    and rapidly fill k-space in a stepped pattern

37
(No Transcript)
38
EPI acquisition (cont.)
  • Echo planar images typically have poor SNR, low
    resolution, and many artifacts
  • Technique offers real-time snapshot image
    capability with 50 msec or less total acquisition
    time
  • Emerging as a clinical tool for studying
    time-dependent physiologic processes and
    functional imaging

39
My brain hurts
  • With apologies to Monty Python

40
(No Transcript)
41
(No Transcript)
42
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com