Title: Comparative Analysis of Microvessel Density in Xenograft Tumors: Manual and Automated Methods for Qu
1Comparative Analysis of Microvessel Density in
Xenograft TumorsManual and Automated Methods
for Quantitating Changes in Tumor
VasculatureBelinda Cancilla, PhD
2Outline
- Spectrum Setup
- Organization of digital slides in research
environment - Aperio MVD Algorithm Optimization
- Standard CD31 xenograft images
- Overstained CD31 xenograft images
- Understained CD31 xenograft images
- MVD Analysis of Xenograft Studies
- Manual MVD analysis
- Semi-automated analysis using ACIS
- Semi-automated analysis using Aperio
3Spectrum Setup
4Spectrum Setup Projects
5Spectrum Setup Specimens
6Spectrum Setup Digital Slides
7Project Structure
8Specimen Structure
9Spectrum Setup TMAs
10TMA Specimen
11Aperio MVD Algorithm
12Aperio MVD Algorithm Optimization
Tuning vessel stain
Analysis vessels
Analysis vessels (show excluded vessels)
Tuning vessel thresholding
Analysis lumen and vascular cells
Analysis lumen and vascular cells Include
incomplete vessels
Segmented vessel
18 Vessels (include incomplete vessels)
7 Vessels (exclude incomplete vessels)
13Aperio MVD Algorithm
- Overstained CD31 in MCF7 Xenograft
14Aperio MVD Algorithm Optimization
Tuning vessel stain
Tuning vessel thresholding
Tuning vessel thresholding
Dark stain threshold 160 Light stain threshold
215
Dark stain threshold 150 Light stain threshold
185
Analysis lumen and vascular cells Include
incomplete vessels
Segmented vessel
Analysis lumen and vascular cells Include
incomplete vessels
Segmented vessel
Smoothing 2 20 Vessels
Smoothing 5 9 Vessels
15Aperio MVD Algorithm
Understained CD31 in A549 Xenograft
16Aperio MVD Algorithm Optimization
Tuning vessel stain
Tuning vessel thresholding
Tuning vessel thresholding
Dark stain threshold 160 Light stain threshold
210
Dark stain threshold 210 Light stain threshold
235
Analysis lumen and vascular cells Include
incomplete vessels
Analysis lumen and vascular cells Include
incomplete vessels
Enlarged lumen
Smoother vessel
8 Vessels
Adjusted vessel parameters
17MDA-MB-231 Xenograft StudyCD31 Manual Counts
18MDA-MB-231 Xenograft StudyCD31 ACIS Analysis
19MDA-MB-231 Xenograft StudyCD31 Aperio Analysis
20Comparison of Results
21Alternative to MVD counts
CD31 in Angiosarcoma
Analysis lumen and vascular cells Include
incomplete vessels
289 Vessels
Tuning vessel thresholding
Positive Pixel Algorithm
Total Analysis Area 230081.6 um2 Total Stain
Area 51012.1 um2 CD31 staining ( area) 22
CD31 staining ( area) 17
22Summary Aperio MVD Algorithm
- Pros
- Identifies CD31 stained vessels in well stained
xenograft images - Identifies CD31 stained vessels in overstained
xenograft images - Identifies CD31 stained vessels in understained
xenograft images - Analysis allows semi-automated batch analysis
- Analysis is comparable to manual and other
semi-automated methods - Cons
- Uses lumen to define vessels not all vessels
have nice round lumens - Over-segmenting vessels based on lumen parameters
- Undercounting some samples need to re-adjust
parameters for some samples within a batch - MVD unit conversion need data in mm2 not µm2