Title: Introduction to Magnetic Resonance Imaging
1Introduction to Magnetic Resonance Imaging
- Benjamin M Ellingson, MS
- Marquette University
- 21 February 2007
2Quantum Theory of Magnetic Resonance
3Quantum Theory of Magnetic Resonance
Angular Momentum
Magnetic Moment, M0
Low Energy Parallel
High Energy Antiparallel
4Quantum Theory of Magnetic Resonance
5Quantum Theory of Magnetic Resonance
- So, B0 (static magnetic field) causes some
particles to align antiparallel, but most align
parallel - Classical View Vector sum of all magnetization
is parallel to B0
M0
B0
Classical View is easier to conceptualizehowever
some quantum restraints
6Theory of Magnetic Resonance
- Because of the Uncertainty Principle, spins
cannot completely align with B0 because the
momentum of the particle cannot be defined
completely, instead they precess or wobble around
B0 at the Larmor Frequency.
Magnetic Field
Larmor Frequency
Gyromagnetic Ratio (specific to atom 1H ? 42.6
MHz/T
7Theory of Magnetic Resonance
Laboratory Frame of Reference See m rotating
about B0 with net magnetization in z-direction,
Mz. The time average value of Mxy is 0.
Mz
m
Mxy
Rotating Frame of Reference Observer is rotating
at the precession frequency, such that m is not
moving. All we see all the components of m. In
rotating frame of reference we will call this M0.
So, placing many 1H atoms in a static magnetic
field ? M0 Mz.
8Perturb Magnetic Equilibrium
- By applying a horizontal oscillating field at
Larmor frequency (B1) produces a torque on the
magnetization vector, M0. - Since B1 ltlt B0 the net field is still in
z-dir - Causes M0 to tip into xy-plane.
9RF Excitation
Laboratory Frame of Reference
Rotating Frame of Reference
10RF Excitation Effect of Frequency
B1(t) B1 cos (0.5wt)
Static B1
B1(t) B1 cos (wt)
B1(t) B1 cos (1.5wt)
B1(t) B1 cos (2wt)
11Relaxation
- After excitation, if B1 field is turned off the
spins undergo relaxation in both transverse and
longitudinal directions at different rates. - Transverse Relaxation T2-relaxation Spin-Spin
Relaxation - Corresponds to dephasing of neighboring spins
- Causes decrease in Mxy
- Longitudinal Relaxation T1-relaxation Lattice
Relaxation - Causes increase in Mz after excitation
12MR Signal
- If we have a lot of 1H excited such that they are
spinning in phase in the xy-plane (i.e. changing
magnetic field) we can detect this with an
antenna due to Faradays Law of Induction
Antenna
Total Magnetization
13MR Signal Free Induction Decay
- As T2 relaxation occurs (Mxy decreasing),
sinusoidal signal at antenna decays with T2
envelope ? Free Induction Decay (FID)
14Localization via Magnetic Field Gradients
- In a static magnetic field, we have no way of
knowing where MR signal is coming from (i.e. all
1H are precessing at same frequency)
15Localization via Magnetic Field Gradients
- To solve this problem we introduce a GRADIENT
FIELD - Gradient magnetic fields add to or subtract from
the main magnetic field in a controlled and
predictable pattern so the field is no longer
homogeneous.
16Localization via Magnetic Field Gradients
17Localization via Magnetic Field Gradients
FREQUENCY ENCODE
18Localization via Magnetic Field Gradients
- Frequency Encoding causes 1-D localization but
what about other dimensions? - Use field gradients to Phase Encode signal
- By pulsing a gradient in another direction we can
speed up or slow down spins
19Localization via Magnetic Field Gradients
Gradient Turned On
Gradient Turned Off
Spins in Static Magnetic Field
Same frequency but different phase!
202-D Spatial Frequency Domain k-space
21The FID Echo
- We get maximum signal when all spins are in phase
and no signal when spins are dephased. - Just as we used a pulsed gradient to phase
encode, we can use pulsed gradients to rephase
after dephasing has occurred. - The process of rephasing spins causes a symmetric
FID with maximum at time when spins are
completely rephased.
22Slice Selection
- The previous RF excitation was applied to all
1H-spins in the body because they were all at the
Larmor Frequency (w0 gB0). - If we apply a gradient, Gss, while applying RF
excitation at a very specific frequency we can
excite an infinitely thin layer of spins. - Practically, we want to excite a slab of spins
so we have high signal, therefore we envelope the
RF excitation in a sinc function.
23The MRI Pulse SequenceIdeal Gradient Recalled
Echo (GRE)
K-space
24MRI Pros Cons
- Pros
- Non-ionizing radiation
- Limitless Contrast Possibilities (based on Pulse
Sequence Design) - Can image in any plane (vs. Axial only for CT)
- Exquisite Resolution Soft Tissue Contrast
- Cons
- Relatively Slow (changing due to better hardware
and sequence design such as EPI) - No metal (although most implants are now MR
compatible) - Claustrophobia Loud
25Clinical Applications
- Too Numerous to list them all
- Angiography
- Diffusion
- fMRI (BOLD ASL)
- Cardiac
26Medical Imaging Computing
- Making information accessible
- CAD, 3D Visualization, Modality Registration
- Reconstruction Processing Algorithms
- Novel Pulse Sequence Image Reconstruction
- Real-time Image Reconstruction
- Code optimization for fast imaging sequences
- Archival Storage
- DICOM, PACs, Image Compression
27Additional Info References
- Additional Information
- http//www.ellingsonbiomedical.com/MRI/Lectures/In
tro_to_MRI.htm - Medical College of Wisconsin Biophysics
- http//www.mcw.edu
- NIH Image Processing Interest Group
- http//image.nih.gov
- Johns Hopkins Biophysics Group
- http//biophysics.jhu.edu
- Stanford Magnetic Resonance Laboratory
- http//smrl.stanford.edu