Title: ULTRAVIOLET-VISIBLE SPECTROSCOPY
1ULTRAVIOLET-VISIBLE SPECTROSCOPY
2In this lecture, you will learn
- Molecular species that absorb UV/VIS radiation
- Absorption process in UV/VIS region in terms of
its electronic transitions - Important terminologies in UV/VIS spectroscopy
3MOLECULAR SPECIES THAT ABSORB UV/VISIBLE
RADIATION
Inorganic species
Organic compounds
Charge transfer
4Definitions
- Organic compound
- Chemical compound whose molecule contain carbon.
- E.g. C6H6, C3H4
- Inorganic species
- Chemical compound that does not contain carbon.
- E.g. transition metal, lanthanide and actinide
elements - Cr, Co, Ni, etc..
- Charge transfer
- A complex where one species is an electron donor
and the other is an electron acceptor. - E.g. iron(III) thiocyanate complex
5NOTE Transition metals - groups IIIB through IB
6UV-VIS ABSORPTION
- In UV/VIS spectroscopy, the transitions which
result in the absorption of EM radiation in this
region are transitions btw electronic energy
levels.
7Molecular absorption
- In molecules, not only have electronic level
but also consist of vibrational and rotational
sub-levels. - This result in band spectra.
8Type of Transitions
- 3 types of electronic transitions
- s, ? and n electrons
- d and f electrons
- Charge transfer electrons
9What is s, ? and n electrons?
10Sigma (?)electron
- Electrons involved in single bonds such as those
between carbon and hydrogen in alkanes. - These bonds are called sigma (s) bonds.
- The amount of energy required to excite electrons
in s bond is more than UV photons of wavelength.
For this reason, alkanes and other saturated
compounds (compounds with only single bonds) do
not absorb UV radiation and therefore frequently
very useful as transparent solvents for the study
of other molecules. For example, hexane, C6H14.
11Pi (?) electron
- Electrons involved in double and triple bonds
(unsaturated). - These bonds involve a pi (?) bond.
- For example alkenes, alkynes, conjugated olefins
and aromatic compounds. - Electrons in ? bonds are excited relatively
easily these compounds commonly absorb in the UV
or visible region.
12- Examples of organic molecules containing ? bonds.
propyne
ethylbenzene
benzene
1,3-butadiene
13n electron
- Electrons that are not involved in bonding
between atoms are called n electrons. - Organic compounds containing nitrogen, oxygen,
sulfur or halogens frequently contain electrons
that are nonbonding. - Compounds that contain n electrons absorb UV/VIS
radiation.
14- Examples of organic molecules with non-bonding
electrons.
Carbonyl compound If R H aldehyde If R
CnHn ketone
aminobenzene
2-bromopropene
15Absorption by Organic Compounds
- UV/Vis absorption by organic compounds requires
that the energy absorbed corresponds to a jump
from occupied orbital unoccupied orbital. - Generally, the most probable transition is from
the highest occupied molecular orbital (HOMO) to
the lowest unoccupied molecular orbital (LUMO).
16Electronic energy levels diagram
Unoccupied levels
Occupied levels
17? ? transitions
- Never observed in normal UV/Vis work.
- The absorption maxima are lt 150 nm.
- The energy required to induce a s s
transition is too great (see the arrow in energy
level diagram) - This type of absorption corresponds to breaking
of C-C, C-H, C-O, C-X, .bonds
s s
vacuum UV region
18n ? transitions
- Saturated compounds must contain atoms with
unshared electron pairs. - Compounds containing O, S, N and halogens can
absorb via this type of transition. - Absorptions are typically in the 150 -250 nm
region and are not very intense. - e range 100 3000 Lcm-1mol-1
19Some examples of absorption due to n s
transitions
Compound ?max (nm) emax
H2O 167 1480
CH3OH 184 150
CH3Cl 173 200
CH3I 258 365
(CH3)2O 184 2520
CH3NH2 215 600
20n ? transitions
- Unsaturated compounds containing atoms with
unshared electron pairs - These result in some of the most intense
absorption in 200 700 nm region. - e range 10 100 Lcm-1mol-1
21? ? transitions
- Unsaturated compounds to provide the ? orbitals.
- These result in some of the most intense
absorption in 200 700 nm region. - e range 10oo 10,000 Lcm-1mol-1
22Some examples of absorption due to n ? and
? ? transitions
23CHROMOPHORE
- Unsaturated organic functional groups that absorb
in the UV/VIS region are known as chromophores.
24(No Transcript)
25AUXOCHROME
- Groups such as OH, -NH2 halogens that attached
to the doubley bonded atoms cause the normal
chromophoric absorption to occur at longer ? (red
shift). These groups are called auxochrome.
26Effect of Multichromophores on Absorption
- More chromophores in the same molecule cause
bathochromic effect ( shift to longer ?) and
hyperchromic effect(increase in intensity) - In the conjugated chromophores ? electrons are
delocalized over larger number of atoms causing a
decrease in the energy of ? ? transitions and
an increase in ? due to an increase in
probability for transition.
27Other Factor that Influenced Absorption
- Factors that influenced the ?
- i) Solvent effects (shift to shorter ? blue
shift) - ii) Structural details of the molecules
28Important terminologies
- hypsochromic shift (blue shift)
- - Absorption maximum shifted to shorter ?
- bathochromic shift (red shift)
- - Absorption maximum shifted to longer ?
29Terminology for Absorption Shifts
Nature of Shift Descriptive Term
To Longer Wavelength Bathochromic
To Shorter Wavelength Hypsochromic
To Greater Absorbance Hyperchromic
To Lower Absorbance Hypochromic
30Absorption by Inorganic Species
- Involving d and f electrons absorption
- 3d 4d electrons
- - 1st and 2nd transition metal series
- e.g. Cr, Co, Ni Cu
- - Absorb broad bands of VIS radiation
- - Absorption involved transitions btw filled and
unfilled d-orbitals with energies that depend on
the ligands, such as Cl-, H2O, NH3 or CN- which
are bonded to the metal ions.
31Absorption spectra of some transition-metal ions
32- 4f 5f electrons
- - Ions of lanthanide and actinide elements
- - Their spectra consists of narrow, well-defined
characteristic absorption peaks
33Typical absorption spectra for lanthanide ions
34Charge Transfer Absorption
- Absorption involved transfer of electron from the
donor to an orbital that is largely associated
with the acceptor. - an electron occupying in a s or ? orbital
(electron donor) in the ligand is transferred to
an unfilled orbital of the metal (electron
acceptor) and vice-versa. - e.g. red colour of the iron(III) thiocyanate
complex
35INSTRUMENTATION
36Important components in a UV-Vis spectrophotometer
1
2
5
3
4
Source lamp
Sample holder
Signal processor readout
? selector
Detector
UV region - Deuterium lamp H2 discharge tube
Phototube, PM tube, diode array
Quartz/fused silica
Prism/monochromator
Visible region - Tungsten lamp
Prism/monochromator
Glass/quartz
Phototube, PM tube, diode array
37UV-VIS INSTRUMENT
38Single beam instrument
39- Single beam instrument
- One radiation source
- Filter/monochromator (? selector)
- Cells
- Detector
- Readout device
40Single beam instrument
- Disadvantages
- Two separate readings has to be made on the
light. This results in some error because the
fluctuations in the intensity of the light do
occur in the line voltage, the power source and
in the light bulb btw measurements. - Changing of wavelength is accompanied by a change
in light intensity. Thus spectral scanning is not
possible.
41Double beam instrument
Double-beam instrument with beams separated in
space
42Double beam instrument
- Advantages
- 1. Compensate for all but most short-term
fluctuations in the radiant output of the source
as well as for drift in the transducer and
amplifier - 2. Compensate for wide variations in source
intensity with ? - 3. Continuous recording of transmittance or
absorbance spectra
43Quantitative Analysis
- The fundamental law on which absorption methods
are based on Beers law (Beer-Lambert law).
44Measuring absorbance
- You must always attempt to work at the wavelength
of maximum absorbance (?max) - This is the point of maximum response, so better
sensitivity and lower detection limits. - You will also have reduced error in your
measurement.
45(No Transcript)
46Quantitative Analysis
- Calibration curve method
- Standard addition method
47- Calibration curve method
- A general method for determining the
concentration of a substance in an unknown sample
by comparing the unknown to a set of std sample
of known concentration
48Standard Calibration Curve
Absorbance
Concentration, ppm
- How to measure the concentration of unknown?
- Practically, you have measure the absorbance of
your unknown. Once you know the absorbance value,
you can just read the corresponding concentration
from the graph .
49How to produce standard calibration curve?
A
- Prepare a series of standard solution with known
concentration. - Measure the absorbance of the standard solutions.
- Plot the graph A vs concentration of std.
- Find the best straight line by using
least-squares method.
50Finding the Least-Squares Line
Concentration xi Absorbance yi x2i y2i xiyi
5 0.201
10 0.420
15 0.654
20 0.862
25 1.084
? xi ? yi ? xi2 ? yi2 ? xiyi
N 5
N is the number of points used
51- The calculation of the slope and intercept is
simplified by defining three quantities Sxx, Syy
and Sxy. - Sxx ? (xi x)2 ? xi2 (? xi)2 (1)
- Syy ? (yi y)2 ? yi2 (? yi)2 (2)
- Sxy ? (xi x) (yi y)2 ? xiyi ? xi ? yi
(3)
N
N
N
52- The slope of the line, m
-
- m Sxy
- Sxx
- The intercept, b
- b y - mx
- Thus, the equation for the least-squares line is
- y mx b
53Concentration x y mx b
5
10
15
20
25
- From the least-squares line equation, you can
calculate the new y values by substituting the x
value. - Then plot the graph.
54- Most linear regression implementations have an
option to force the line through the origin,
which means forcing the intercept of the line
through the point (0,0). This might seem
reasonable, since a sample with no detectable
concentration should produce no response in a
detector, but must be used with care. - HOWEVER, forcing the plot through (0,0) is not
always recommended, since most curves are run
well above the instrumental limit of detection
(LOD). Randomly, adding a point (0,0) can skew
the curve because the instruments response near
the LOD is not predictable and is rarely linear.
As illustrated next page, forcing a curve through
the origin can, under some circumstances, bias
results.
55(No Transcript)
56- Standard addition method
- used to overcome matrix effect
- involves adding one or more increments of a
standard solution to sample aliquots of the same
size. - each solution is diluted to a fixed volume before
measuring its absorbance
57Standard Addition Plot
Absorbance
Concentration, ppm
58How to produce standard addition curve?
- Add same quantity of unknown sample to a series
of flasks - Add varying amounts of standard (made in solvent)
to each flask, e.g. 0,5,10,15mL) - Fill each flask to line, mix and measure
59Standard Addition Methods
Single-point standard addition method
Multiple additions method
60- Standard addition
- - if Beers law is obeyed,
- A ebVstdCstd ebVxCx
- Vt Vt
- kVstdCstd kVxCx
k is a constant equal to eb
Vt
61- Standard Addition
- - Plot a graph A vs Vstd
- A mVstd b
- where the slope m and intercept b are
- m kCstd b kVxCx
62- Cx can be obtained from the ratio of these two
quantities m and b - b kVxCx
- m kCstd
- Cx bCstd
- mVx
63- Standard Addition
- For single-point standard addition
Absorbance of diluted sample
A1 ebVxCx Vt
Eq. 1
Absorbance of diluted sample std
A2 ebVxCx Vt
ebVsCs Vt
Eq. 2
64- Standard Addition
- For single-point standard addition
- Dividing the 2nd equation by the first then
rearrange it will give - Cx A1 Cs Vs
- (A2 A1 ) Vx
65Example Standard Addition (single point addition)
Example 14-2 (page 376)
- A 2.00-mL urine specimen was treated with
reagent to generate a color with phosphate,
following which the sample was diluted to 100 mL.
To a second 2.00mL sample was added exactly
5.00mL of a phosphate solution containing 0.03 mg
phosphate /mL, which was treated in the same way
as the original sample. The absorbance of the
first solution was 0.428, while the second one
was 0.538. Calculate the concentration of
phosphate in milligrams per millimeter of the
specimen.
66- Solution
- Cx (0.428) (0.03 mg PO43-/mL) (5.00mL)
- (0.538 0.428)(2.00mL)
- 0.292 mg PO43- / mL sample
67Exercise
The concentration of an unknown chromium solution
was determined by pipetting 10.0mL of the unknown
into each of five 50.0 mL volumetric flasks.
Various volumes of a standard containing 12.2 ppm
chromium were added to the flasks and then the
solutions were diluted to the mark.
Standard, mL Absorbance
0.0 0.201
10.0 0.292
20.0 0.378
30.0 0.467
40.0 0.554
Determine the concentration of chromium (in ppm)
in the unknown.