Equipment Care and Maintenance - PowerPoint PPT Presentation

1 / 148
About This Presentation
Title:

Equipment Care and Maintenance

Description:

Oak Environmental Inc. Equipment Care and Maintenance Wesley Hutchen Oak Environmental Inc. * * * * Before we get into too much detail about the different hyperlink ... – PowerPoint PPT presentation

Number of Views:1348
Avg rating:5.0/5.0
Slides: 149
Provided by: betterear
Category:

less

Transcript and Presenter's Notes

Title: Equipment Care and Maintenance


1
Equipment Care and Maintenance
Oak Environmental Inc.
  • Wesley Hutchen

2
The Basics
  • Overview Becoming familiar with equipment use
    and maintenance
  • Troubleshooting
  • Basic servicing

3
Gastech 1238 ME

4
Gastech 1238 ME
  • Requires 5 minute warm up period
  • Flow rate must be 2 5 CFH
  • Needle must be zeroed

5
Gastech 1238 ME
  • Buttons must have free movement
  • Calibration must be done weekly, or after a high
    concentration exposure
  • Water filter must be used

6
Diagnosing a Problem
  • Check battery condition
  • Check flow rate
  • Check sensor chamber and hoses for leaks
  • Inspect wiring
  • Is the unit warmed up?

7
Pump Servicing and Cleaning
8
Pump Service and Cleaning
  • Take note of where parts go!
  • Use
  • Diagrams
  • Camera phone

9
Pump Servicing
  • Turn unit on, check for adequate suction at inlet

10
Pump Servicing
  • Remove retaining screw from bottom of unit

11
Pump Servicing
  • Remove retaining bracket

12
Pump Servicing
  • Remove pump assembly

13
Pump Servicing
  • Disconnect inlet hose from pump

14
Pump Servicing
  • Remove 4 pump head screws

15
Pump Servicing
  • Remove pump head

16
Pump Servicing
  • Remove and inspect pump o-rings, seals and valves

17
Pump Servicing
  • Remove 4 retaining screws from pump body

18
Pump Servicing
  • Carefully separate pump body from pump motor
    bracket

19
Pump Servicing
  • Remove screw and retainer from pump diaphragm

20
Pump Servicing
  • Remove diaphragm from pump body assembly

21
Clean parts with distilled water
Pump Servicing
22
Pump Servicing
  • Check sensor hose for debris

23
Remember!
Pump Servicing
  • Check all parts for damage or wear!

24
Pump Servicing
  • Re-install pump diaphragm and retainer rib on
    diaphragm faces up

25
Pump Servicing
  • Center diaphragm and replace pump body

26
Pump Servicing
  • Pump marker should face towards motor assembly

27
Pump Servicing
  • Replace 4 retaining screws on pump body do not
    over tighten!

28
Pump Servicing
  • Replace o-rings and Teflon seals in correct
    sequence

29
Pump Servicing
  • Replace pump head and tighten 4 retaining screws
    do not over tighten!

30
Pump Servicing
  • Re-attach inlet hose to pump ensure correct
    orientation!

31
Pump Servicing
  • Re-install pump into Gastech

32
Pump Servicing
  • Re-install pump retaining bracket

33
Pump Servicing
  • Turn unit on and recheck suction

34
Pump Servicing
  • Allow unit to fully warm up before use

35
Things to Check
  • Is the sensor tight?
  • Are there any broken wires?
  • Are there any hoses kinked or disconnected?
  • Is the water filter plugged?

36
Questions?

37
Calibration
  • Gastech 1238 ME

38
Gastech 1238 ME
  • Typical Gas
  • Hexane 400 PPM
  • Hexane 40 LEL

39
Gastech 1238 ME
  • Circuit board Layout

40
Gastech 1238 ME
  • Find center position for zero knob
  • Zero instrument in LEL mode using coarse zero
    potentiometer

41
Gastech 1238 ME
  • Measure flow rate using flow meter

42
Gastech 1238 ME
  • Apply gas at same flow rate or use a demand
    regulator
  • Use humidifier!

43
Gastech 1238 ME
  • Adjust needle to match gas concentration on bottle

44
Gastech 1238 ME
  • Sensor span can be low in PPM mode apply LEL
    gas in FULL GAS MODE for 30 seconds. This can
    restore sensor span
  • Sensor lead wires must be soldered to ensure a
    clean connection
  • Use water trap filter and maintain cotton in probe

45
Questions?

46
RKI Eagle/Thermo Innova
  • Introduction

47
RKI Eagle/Thermo Innova
  • Replacements for the Gastech 1238 ME

48
RKI Eagle/Thermo Innova
  • Benefits and Features
  • LCD display
  • Internal hydrophobic filter
  • Low flow alarm
  • User programmable alarm limits
  • No potentiometers!
  • Simple to use and calibrate

49
RKI Eagle/Thermo Innova
  • Operates in LEL or PPM mode
  • Has methane elimination
  • Uses standard D cell batteries

50
RKI Eagle/Thermo Innova
  • Has optional dilution fitting for low O2
    environments

51
Questions?

52
  • Photo Ionization Detector
  • (PID)

53
PIDs
  • MiniRae 2000/3000
  • Photo Ionization detector

54
Field Operation
  • Requires 5 second warm up period
  • Flow rate between .5 1 CFH
  • Probe must not be obstructed
  • Use humidity filter for headspace sampling

55
Field Operation
  • Warm up/self test sequence not to be interrupted
  • Calibration must be done frequently
  • Water filter must be used
  • Moisture filter must be used for headspace
    sampling

56
Diagnosing a Problem
  • Check battery condition
  • Check flow rate
  • Check inlet probe for tightness
  • Inspect filter
  • Are gas settings correct?
  • Inspect hoses and fittings

57
  • Servicing and Cleaning

58
Servicing and Cleaning
  • Remove inlet probe

59
Servicing and Cleaning
  • Check for suction with probe removed

60
Servicing and Cleaning
  • Unscrew cap, check filter and flame
  • arrestor

61
Other Problems
  • Return unit to Oak Environmental for
  • service

62
Questions?

63
Interface Meters

64
Interface Meters
  • Cleaning and Servicing

65
Interface Meters
  • Check battery
  • Condition

66
Interface Meters
  • Inspect prism for
  • damage

67
Interface Meters
  • Inspect tape for
  • damage

68
Interface Meters
  • Cleaning

69
Interface Meters
  • Wipe tape assembly with towel and Alconox

70
Interface Meters
  • Carefully clean probe and prism

71
Interface Meters
  • Clean wiper assembly

72
Other Problems
  • Return to Oak Environmental for service

73
Water Level Meters
  • Check battery
  • Check tape for damage
  • Do not test in DI water
  • May read sporadically in cascading water

74
  • pH/Conductivity Meters
  • Care and Maintenance

75
Questions?

76
pH/Conductivity Meters
  • pH probes are to be stored in solution
  • Probes must not be frozen
  • Clean with soft cloth or Q tip

77
pH/Conductivity Meters
  • Check battery
  • condition

78
pH/Conductivity Meters
  • Inspect probes for
  • damage

79
pH/Conductivity Meters
  • Inspect wiring for
  • damage

80
pH/Conductivity Meters
  • Calibrate instrument according to
  • manufacturers instructions

81
Other Issues
  • Dried out probe must be cleaned and calibrated
    before use
  • Allow time to stabilize before calibration
  • Return to Oak Environmental for further servicing

82
DO Meters
  • E7 error often means loose membrane
  • Probe must be moving when sampling

83
Questions?

84
YSI Meters
  • YSI 556

85
The Basics

86
YSI 556
  • Multi parameter instrument which can measure
  • pH
  • Conductivity
  • ORP
  • DO
  • Temperature
  • TDS
  • Salinity

87
YSI 556
  • Probe guard must be used

88
YSI 556
  • Stores data either in a snapshot or datalogs at
    user defined intervals

89
YSI 556
  • Simple to operate and calibrate

90
YSI 556
  • Servicing/Calibration

91
YSI 556
  • Typically a two point pH calibration is performed

92
YSI 556
  • Wait for readings to stabilize before confirming
    calibration

93
YSI 556
  • If parameters are too far off, instrument will
    ask you if you want to proceed
  • Often required several times

94
YSI 556
  • Keep probe stored in either storage solution or
    distilled water

95
YSI 556
  • Check battery condition, and have spares!

96
YSI 556
  • Ideal instrument for low flow sampling

97
Questions?

98
  • GT Series Gastech

99
Diagnosing a Problem
  • Check battery condition
  • Check flow rate
  • Fresh air zero instrument

100
Diagnosing a Problem
  • Error messages?
  • Warmed up?
  • Inspect hoses and fittings

101
GT Series Gastech
  • Calibrate according to manufacturers
  • instructions

102
Other Issues
  • Return to Oak Environmental for servicing

103
Questions?

104
4 Gas Personal Monitors
  • 4 Gas Personal Monitors

105
Common Issues
  • Batteries
  • Calibration due
  • Zero drifting

106
Zero Procedure for BW Monitors
  • Press and hold both bottom buttons
  • (Fresh air atmosphere)
  • Wait for unit to zero
  • Check for cleared sensors

107
Crowcon Tetra 3 Meters
  • Follow same procedures for BW personal monitors

108
Crowcon Tetra 3 Meters
  • Zero is performed using one button
  • Double click to access menu
  • Single click to move cursor
  • Double click to activate zero sequence

109
  • Remember!
  • Properly maintaining equipment saves time and
    money!

110
Questions?

111
Equipment Quiz

112
Question One
  • Which of the following does a Photo Ionization
    Detector NOT detect?
  • Perchlorethylene
  • Diesel fuel
  • Methanol

113
Question One Answer
  • C. Methanol (Trick question though. You can use
    an 11.7ev lamp, and it will!)

114
Question Two
  • Why must you use a water knockout filter when
    headspace sampling with a PID?
  • The unit will rust
  • Moisture will condense on the lamp, rendering the
    unit useless

115
Question Two Answer
  • B. Moisture will condense on the lamp

116
Question Three
  • What is the reason for using a Photo Ionization
    Detector for specific gases?
  1. It automatically adjusts when it senses a gas
  2. You can set a correction factor in the unit to
    give exact readings for specific compounds
  3. It is prettier than a 1238 ME

117
Question Three Answer
B. You can set a correction factor in the unit to
give exact readings for specific compounds
118
Question Four
  • You can turn off a GT series Gastech while there
    is a value remaining on the display for CO or
    Combustibles
  • True or False?

119
Question Four Answer
  • False.
  • When you turn the unit back on, you will see a
    negative reading due to the unit retaining a
    memory of the previous reading

120
Question Five
  • Which one of the following is true about a
    Gastech 1238 ME?

A. It calibrates using methane B. You can
switch between ME mode and full gas without
recalibrating C. It requires a five minute warm
up period
121
Question Five Answer
  • C. It requires a five minute warm up period

122
Question Six
  • What is ME referring to?
  • Methane elimination
  • Methane enabled

123
Question Six Answer
  • A. Methane elimination

124
Question Seven
  • Which part should you NOT use the gastech 1238 ME
    without?
  1. The water filter
  2. The cotton particulate filter in the probe
    assembly
  3. The Dust cap which covers the charger plug
  4. All of the above

125
Question Seven Answer
D. All of the above
126
Question Eight
  • When should you charge the battery in a Gastech
    1238 ME?
  • When the low battery alarm sounds
  • Weekly
  • Nightly

127
Question Eight Answer
  • C. Nightly

128
Question Nine
  • How often should a DO meter be calibrated?
  • Weekly
  • Daily
  • After a change of membrane

129
Question Nine Answer
  • Both B and C.
  • That way the unit can adjust for temperature,
    electrolyte changes, altitude and membrane
    contamination

130
Question Ten
  • When calibrating a Gastech 1238 ME, which one of
    the following is correct?
  • Matching the gas flow rate to the flow of the
    pump
  • Turning the course zero potentiometer on the
    motherboard to gain span
  • Turning off the alarm during calibration

131
Question Ten Answer
  • A. Matching the gas flow rate to the flow of the
    pump

132
Question Eleven
  • Most conductivity meters have built in
    compensation for what parameter?
  • Altitude
  • Temperature
  • Turbidity

133
Question Eleven Answer
  • Temperature

134
Question Twelve
  • RKI Eagles and Innova FV instruments calibrate
    using LEL gas only
  • WHY?

135
Question Twelve Answer
  • PPM calibration is not accurate,
  • nor stable in a digital instrument

136
Question Thirteen
  • YSI DO parameter should be calibrated in the
    sealed chamber it is stored in
  • True or False?

137
Question Thirteen Answer
  • False.
  • Atmospheric compensation will not be possible,
    and the calibration will not be accurate

138
Question Fourteen
  • You can test a depth sounder or interface probe
    by dipping the probe into distilled water
  • True or false?

139
Question Fourteen Answer
  • False.
  • You need conductivity in the sample to be able to
    activate the instrument

140
Question Fifteen
  • You can test an oxygen sensor by blowing on it
  • True or False?

141
Question Fifteen Answer
  • True.
  • The reading will drop dramatically just from your
    breath

142
Gastech Reminders
  • Calibrate frequently
  • Check flow rate daily
  • Report any problems or malfunctions immediately
  • Have unit serviced quarterly by a certified
    technician

143
Four Gas Monitor Reminders
  • Zero unit at the beginning of the day
  • Bump test daily
  • Report any malfunctions
  • Do not use unit if bump test fails

144
pH/Conductivity Reminders
  • Keep probes clean
  • Do not allow probes
  • to touch the ground

145
Questions?

146
  • Email
  • wesley_at_oakenviro.com
  • sara_at_oakenviro.com
  • frank_at_oakenviro.com
  • terry_at_oakenviro.com

147
  • Website address
  • www.oakenviro.com

148
  • Thank You!
  • Oak Environmental Inc.
Write a Comment
User Comments (0)
About PowerShow.com