Title: Hemoglobin Testing Using the HemoCue
1Hemoglobin TestingUsing the HemoCue
2Instrument calibration
- At the beginning of each survey day, check the
instrumentaccuracy using the calibrator
cuvette. Record readings. - If readings are in question, clean the cuvette
holder and calibrator cuvette with a dry Kim
wipe. - If readings continue to be outside the correct
range or reads ERROR, do not use the instrument.
It should be serviced or replaced. - If the calibrator cuvette reads within the
specified range listed on its container, report
the result and perform daily quality control (QC)
on the instrument.
3Daily quality control
- Perform daily QC (morning and evening) by
measuring and recording the results for each of
the low, normal, and high range control vials. - Report results on the HemoCue Hemoglobin Quality
Control Form.
4Morning quality control
HemoCue Serial Number Low control lot number
CDC Number Normal control lot number
Control cuvette number High control lot number
Quality control for blood hemoglobin results
(grams per deciliter - g/dL)
Morning Morning Morning Morning Morning Morning
Date Control Cuvette specific for the Instrument Low Target 8.0 0.4 (7.6 - 8.4) Normal Target 11.9 0.6 (11.3-12.5) High Target 15.9 0.8 (15.1 - 16.7) Initials
5Procedure from a Microtainer or Vacutainer
- Assemble all the materials needed prior to
testing each subject. Reseal the cuvette
container immediately after taking out one
cuvette for use. - Remove the cap from the well-mixed Microtainer or
Vacutainer tube.
6Procedure from a Microtainer or Vacutainer
- Fill the HemoCue cuvette by holding the
Microtainer or Vacutainer tube in a horizontal
position and carefully tilt the blood forward to
the edge of the tube. Place the pointed tip
(with the cuvettes groove facing upward) of the
HemoCue cuvette into the blood drop for a count
of three. The cuvette will fill automatically by
capillary action.
7Procedure from a Microtainer or Vacutainer
- Never try to top off the cuvette after the
initial filling. If the cuvette is not filled
with the first attempt, discard it in the sharps
container and use a second cuvette. - Wipe off any excess blood from the cuvette with a
Kim wipe, being careful not to touch the outer
curved edge. This may be done by wiping the
edges as you would a butter knife. Ensure that
no blood is sucked out of the cuvette when
wiping it.
8Procedure from a Microtainer or Vacutainer
- Place the cuvette in its holder and gently push
the holder into the photometer. The results will
be displayed in approximately 15-45 seconds and
will remain displayed for 4 minutes or until the
slide arm is pulled out for removal of the
cuvette.
9Procedure from a Microtainer or Vacutainer
- Record your results, dispose of the cuvette in
the sharps container, and dispose of all other
materials in the biohazard bag.
10Evening quality control
- Measure and record the results for each of the
low, normal, and high range control vials. - Report results on the HemoCue Hemoglobin Quality
Control Form.
Evening Evening Evening Evening Evening
Date Low Target 8.0 0.4 (7.6 - 8.4) Normal Target 11.9 0.6 (11.3-12.5) High Target 15.9 0.8 (15.1 - 16.7) Initials
11HemoCue cleaning instructions
- Once or twice a day, clean the cuvette holder
with an alcohol pad. - Allow to completely dry before testing.
- Periodically, clean the reading chamber with dry
gauze. - More detailed cleaning instructions are provided
in the manual.