Title: Radiation Protection in Radiotherapy
1Radiation Protection inRadiotherapy
IAEA Training Material on Radiation Protection in
Radiotherapy
- Part 10
- Good Practice in EBT
- Lecture 1 (cont.) Equipment design
22. Features of safe design in practice
- A General considerations
- B Kilovoltage radiation units
- C Telecurie units
- D Megavoltage units
- E Other irradiation units
3A. General Safety Requirements
- Radiation Protection Measures include
- Protection of the patient during treatment
- Equipment shielding
- Collimation system
- Patient comfort and control
- Protection of others
- Room shielding (this was covered in part 7)
4Equipment shielding
- Part of dose reduction strategy for patients
- Dose to patient other than target due to scatter
and leakage
5Equipment shielding
- X Ray equipment - only needed when machine is on
- protects the patient during treatment
- Telecobalt units - shielding needed all the time
- protects patient and staff during set-up
General design limit - leakage should be
less than 0.1 of the primary radiation
6Testing of shielding integrity of a linac head
using film
About 2t of lead
7Collimation
- Creates outlines of the radiation field which
should conform to the target - Can be done by a variety of different measures
depending on the treatment unit type - Always includes some leakage through the
collimation - typically lt2 of the primary beam
8Collimation
Customized blocks or prefabricated blocks in
geometric shapes
- Aim to limit field to the target only
9Collimation
- Applicators
- electron beams
- superficial beams
- Movable jaws
- Lead blocks
- fixed shapes
- customized
- Multileaf collimator
10Custom shielding may reduce the dose to critical
organs
- e.g. scrotal shields to reduce dose to scrotum
due to scattered radiation
11Patient comfort and control
- The best collimation does not help if the patient
is not stable - need good immobilization devices
- need to put patient in a reasonably comfortable
position (this is often difficult with very sick
patients) - need to make them feel comfortable
12Immobilization/set-up devices
- There are innumerable systems - many of them home
built and designed - A good mould room is essential - they are
responsible for both, - immobilization and
- block making
13Immobilization/set-up devices
14Head and Neck Immobilization
Head rests to fit
Prone head rest
All MedTec
15Lateral Head position
16Immobilization/set-up devices
- The more accuracy is required, the more effort
one must make e.g. - Stereotactic head frame with repositioning
accuracy better than 2mm
17Immobilization/set-up devices
- Immobilization shells for head
- Vacuum bag for body immobilization
18Various body immobilisation devices
Body fix with external markers for set-up
All MedTec
19Belly board for prone position
- Allows belly to move into space
- Some of the bowel can be moved out of the field
20Vacuum bags
Customized for every patient
All MedTec
21Immobilization/set-up devices
- Board for set-up of breast patients
Arm rest to get arm out of the treatment field
Head rest
Slope to straighten sternum in order to minimize
lung dose
Leg rest
22 sometimes movement is difficult to control...
- e.g. rectal and bladder filling in prostate
treatment - determine location of the prostate prior to each
treatment fraction using ultrasound
23 sometimes movement is difficult to control...
- e.g. lung motion due to breathing
- determine motion and gate radiation beam
External markers on the patient which can be
tracked by a video system
24Low cost solutions
- Ask patients to
- hold still
- have reproducible bladder filling (e.g. always
full or always empty) - provide dietary advise
- breath shallow
- Make patients feel comfortable and secure
25A note on intercom systems
- Need to be able to see the patient - is he/she
comfortable? Is she/he moving? - Need to be able to talk to the patient
- Need to be able to hear if the patient is in
distress
26B. Kilovoltage Equipment (10 - 150 kV)
- Dose rate is approximately proportional to the
nth power of the accelerating potential as kVn
where 2 lt n lt 3 - Dose rate is approximately proportional to
current (mA) - Therefore important that kV and mA are stable.
- It is obviously important that the timer is
accurate and stable
27Kilovoltage Equipment (10 150 kV)
- Dose control is achieved by a dual timer system
as it is usually not practical to use a
transmission ionization chamber - Interlocks should be present to prevent incorrect
combinations of kV, mA, and filtration
28Quick Question
- What are the fluctuations of the mains voltage in
your hospital? What would be the consequence in
dose if these would not be filtered out before
generating the high voltage for the X Ray tube?
29Answer
- A /- 10 voltage variation is not uncommon due
to loading of the net at different times of the
day or heavy occasional uses on the same mains
(e.g. a lift) - This translates into 40 dose variation which is
unacceptable - Mains stabilization is a MUST
30Kilovoltage Equipment (10 - 150 kV)
- Leakage from the tube housing, the Air Kerma Rate
(AKR) shall not exceed - 10 mGy h-1 at 1 metre from focus
- 300 mGy h-1 at 5 cm from housing or accessory
equipment - if the tube is designed to operate in the range
10 - 50 kV then a special housing is required
with a maximum leakage of 1 mGy h-1 - Testing for hot spots should be carried out using
film-wrap techniques
31Patient shielding
- May be done on the skin using lead sheets cut
into customized shapes - Special shields may be used - e.g. eye shields
32Kilovoltage Equipment (150 - 400 kV)Orthovoltage
irradiation units
- It is practical to use a transmission ionization
chamber with this equipment and the primary dose
control system should be an integrating
dosemeter.
- The backup (secondary) dose control system can be
either an independent integrating dosemeter or a
timer
33Kilovoltage Equipment (150 - 400 kV)
- Leakage from the tube housing, the Air Kerma Rate
(AKR) shall not exceed - 10 mGy h-1 at 1 metre from focus
- 300 mGy h-1 at 5 cm from housing or accessory
equipment (including the beam collimation system
such as cones) - Testing for hot spots should be carried out using
film-wrap techniques
34C. Telecurie units
- 137-Cs or more importantly 60-Co
- High activity in treatment head
- Termination of exposure is usually by dual
independent timers
35Timers
- Need two completely independent timers
- One should count time up, one down
36Gamma-ray equipment
- The source should be sealed such that the
container can withstand temperatures likely to be
obtained in building fires. - Wipe tests should be carried out initially at
installation and at regular intervals to check
for surface contamination. This test need not be
carried out directly on the source surface and
can be carried out on a surface which comes into
contact with the source during normal operation
of the equipment.
37Cobalt unit designs
38Gamma-ray equipment
- At commissioning, cross-sectional drawings of the
head should be examined to identify possible
locations where radiation leakage could be a
problem. - Film wrap techniques can be used to identify
positions of hot spots. - Accurate integrated ionization chamber readings
should be made at the location of any hot spots
and also in a regular pattern around the head.
39Gamma-ray equipment
- Leakage from the head with the source in the Off
position the Air Kerma Rate (AKR) shall not
exceed - 10 ?Gy h-1 at 1 metre from source
- 200 ?Gy h-1 at 5 cm from housing or accessory
equipment
40Gamma-ray equipment
- Leakage from the head with the source in the On
position the Air Kerma Rate (AKR) shall not
exceed - 10 mGy h-1 at 1 metre from source or
- 0.1 of the useful beam AKR
- whichever is the greater
41Gamma-ray equipment
- The beam control mechanism shall be of the fail
to safety type and will return to the Off
position in the event of - end of normal exposure
- any breakdown situation
- interruption of the force holding the beam
control mechanism in the On position, for example
failure of electrical power or compressed air
supply
42Gamma-ray equipment
Mechanical source position indicator
- In case of failure of the automatic source return
section of the beam control mechanism, it shall
be possible to interrupt the exposure by other
means, for example, a manual return system - It shall be possible to unload or repair the
treatment head without exceeding the dose limit
for occupational exposure recommended by
regulation
43Gamma-ray equipment
- Collimation, patient immobilisation and blocking
as described in first section of part 10 and the
case of linacs. - Two particularities
- No commercial MLC available (but several home
built systems) - Due to large source size and wide penumbra
penumbra trimmers (collimation close to the
patient can be employed)
44Specific design for Co units
- Penumbra trimmers - collimation close to patient
reduces penumbra width
45Beam stopper
- Metal disk at the exit side
- reduces primary beam shielding requirements
- may make set-up of patients more cumbersome
46D Megavoltage units
- Electron linear accelerators - linacs
- Capable of X Ray (4 to 25MV) and electron (4 to
25MeV) irradiation
47Linacs
- Radiation exposure is usually controlled by two
independent integrating transmission ionization
chamber systems. - One of these is designated as the primary system
and should terminate the exposure at the correct
number of monitor units - The other system is termed the secondary system
and is usually set to terminate the exposure
after an additional dose, typically set around
0.25 Gy - Most modern accelerators also have a timer which
will terminate the exposure if both ionization
chamber systems fail
48Linacs
- Modern accelerators have a lot of treatment
options as discussed in part 6, for example - X Rays or electrons (dual mode)
- multiple energies
- 2 X Ray energies
- 5 or more electron energies
- wedges
- 3 or more fixed wedges
- auto-wedge
- dynamic wedge
49Linacs
- With such a large number of possible settings it
is essential that interlocks be provided to
prevent inappropriate combinations from being
selected - It is also essential that the control console
provide a clear indication of what functions have
been set
50A linac control example
Active selection
Parameter display
Varian
51Linacs
- Verification systems
- All accelerator manufacturers now produce
computer controlled verification systems which
provide an additional check that the settings on
the accelerator console are correct for - proper accelerator function and
- correspond exactly with the parameters determined
for the individual patient during the treatment
planning process
52Linacs - a note on MLCs
- X Ray Collimators may be
- rectangular (conventional)
- Multi-Leaf collimators (MLC)
- the transmission through the collimators should
be less than 2 of the primary (open) beam - The transmission between the leaves should be
checked to ensure that it is less than the
manufacturers specification - this can be done
using radiographic film
53Linacs - electrons
- Electron applicators, these may be
- open sided for modern accelerators using double
scattering foils or scanned beams - enclosed for older accelerators using single
scattering foils - should be checked for leakage
- adjacent to the open beam
- on the sides of the applicators
Cut-out at the end of the applicator
54Electron collimation
- Done at the end of the applicator using
customized cut-outs
Pour LMA around it
Cut-out foam where field should be
55IEC 601.2
- Limit values at different locations around the
useful field
56Electron Accelerators
- Head leakage
- the Air Kerma Rate (AKR) due to leakage radiation
at any point outside the maximum useful beam, but
inside a plane circular area with a radius of 2
metres centred around, and perpendicular to, the
central axis of the beam at the normal distance
of treatment shall not exceed 0.2 of the AKR at
the central axis of the open beam. The
measurement shall be done with a thick shielding
block covering the open beam.
57Electron Accelerators
- Head leakage
- Except in the area defined in the previous slide
the Air Kerma Rate (AKR) due to leakage radiation
(excluding neutrons) at any point 1 metre from
the path of the electrons between their origin
and the target or electron window shall not
exceed 0.5
58IEC standard 601.2
- Leakage in from linac head particularly of
concern if the radiation can reach the patient
59Guidance on leakage levelsin different parts of
the field
60Also consider
- Treatment in different patient positions e.g.
sitting or standing next to the linac for
treatment of a hand
61Linacs - a note on neutrons
- Neutrons will only be a problem if the X Ray
energy is greater than 10 MV - in practice
consideration MUST be given to neutrons if the
energy is greater than or equal to 15MV - The rate of equivalent dose of the neutrons
should not exceed 1 of the dose-equivalent rate
of the X Rays - measured in sievert - The radiation weighting factor for the neutrons
should be taken as 20. The above limit means that
the neutron absorbed dose rate is always less
than the X Ray absorbed dose rate
62Accidents due to equipment design...
- An operator of an accelerator quickly selected X
Ray mode and quickly changed to electron mode
before the machine was able to complete the first
request (to operate in X Ray mode) and it
operated with hybrid instructions. The same
accident occurred in six different hospitals and
two patients died due to doses as high as about
160-180 Gy
63This should not have happened...
- Contributing factors
- The computer controlled accelerators were not
tested for the extreme conditions that occurred
in practice at the hospitals. - It took too long for the manufacturer to identify
the problem and disseminate the information and
by then six hospitals had experienced the same
failure and two patients had died from radiation
64E Other irradiation units
- Diagnostic units in radiotherapy
- CT scanner
- Simulator
- Other therapy irradiation units
- heavy particles
65Diagnostic units in radiotherapy
- Essential and often integral part of a modern
radiotherapy department - Essential for adequate target definition -
therefore important also for optimization of
medical exposure from a radiation protection
point of view - Includes not only X Ray equipment but may be MRI,
ultrasound and nuclear medicine - Beyond the scope of this course - however,
covered in separate courses on diagnostic
radiology and nuclear medicine
66A note on simulators
- The simulator should be capable of reproducing
all motions and X Ray exposure types (not
radiation energy and dose though) of the
treatment units
67Simulator control
Patient clearly visible through large lead glass
window
Fluoroscopy screen
Varian Medical Systems
Control screen similar to linac
68Simulators and other diagnostic equipment
- Often the most important aspect of design is to
ensure that the simulator patient set-up can be
transferred without any modifications to the
treatment unit. This includes imperfections of
the systems such as couch sag under patients
weight.
69Heavy Particles
- These could include
- Neutrons
- Protons
- Helium nuclei (alpha particles)
- Other heavy nuclei (Carbon nuclei)
- Negative pi mesons
- Protons are most common and increasing in their
use
70Heavy Particles treatment facilities
- These are very specialized installations
- shielding with high neutron fluxes can be
extensive and complex - neutrons require hydrogen rich materials for good
energy absorption for example wood and or
plastics - many neutron interactions produce high energy
gamma rays requiring large thicknesses of
concrete , or steel to absorb them
71Additional note on heavy particles
- Many of the points covered for electron
accelerators are also applicable for these
installations - Specialized systems for positioning patients may
be required - The charged particle accelerators are often
multipurpose facilities which also serve research
objectives (e.g. material research). These
applications may require entirely different beam
parameters (e.g. high particle flux) than medical
treatment. More care has to be taken to ensure
that only the correct beam can reach the patient. - There may be several treatment rooms for one
accelerator