Title: Testing of New Ozone Photometers


1
Testing of New Ozone Photometers
  • Ozone Transport Commission/MANEVU
  • Fall Meeting
  • November 14, 2013
  • Washington, DC
  • Will Ollison

2
Testing of New O3 Photometers
  • Do we know what the O3 monitors really
    measuring?
  • EPA currently certifies FEM photometers with up
    to 60 ppb interference bias, 24-hour zero/span
    drifts of 20 ppb, LODs of 10 ppb, and noise
    levels of 5 ppb but notes in the 2013 ISA (EPA
    600/R-10/076F) that these criteria should be
    revised to more accurately reflect the necessary
    performance requirements for O3 monitors used to
    support the current NAAQS.
  • NIST staff notes (Norris et al. 2013 JAWMA 63
    565-574) that regular audits currently provide
    reliable measures of field monitor precision and
    bias in response to challenges of O3 in zero air,
    although not of sample matrix effects arising
    from ambient humidity and interfering species
    (Spicer et al. 2010 JAWMA 60 1353-1364).
  • New monitor testing suggests actual NAAQS
    attainment is better than reported and affordable
    photometer scrubber upgrades will improve O3
    measurement accuracy Ollison et al. 2013 JAWMA
    63 855-863 Johnson et al. 2013 JAWMA (in
    press).

3
Ozone Monitor Types
  • EPA is currently evaluating the NO-scrubbed 2B
    Technologies 211 photometer and the Teledyne-API
    265E NO-O3 chemiluminescence monitor as potential
    new ozone Federal Reference Methods

4
Percent O3 Monitor Share U.S. Network (1990-2011)
5
  • TE49C-211 Hourly Difference Collocated NO VOCs

Ollison et al. (2013) JAWMA 63 855-863.
6
211 ME Factor (211F) 205-211 Difference (205D)
Home Activity
  • Johnson et al. (2013) JAWMA (in press)

7
GPT Scrubberless Module
  • Drop-in NO-scrubbers are commercially available
    from 2B Technologies to convert Teledyne-API
    Thermo photometers to interference-free (H2O,
    Hg, aromatic VOCs) O3 monitors.