Title: Pacemaker Implantation


1
Pacemaker Implantation

2
Pacemaker Implantation
  • Pre-implantation
  • Clinical assessment of patient
  • Age
  • Symptons
  • Underlying cardiac rhythm
  • Underlying medical state
  • Size
  • Evaluation of daily living activities
  • Future needs

3
Pacemaker Implantation
  • Temporary pacing
  • Access - Subclavian
  • - External / Internal jugulars
  • - Femoral
  • - Brachial
  • - Surgical (epicardial)
  • Emergency Pacing - Oesophageal
  • - Transthoracic
  • - Transcutaneous

4
Pacemaker Implantation
  • Permanent pacing
  • Access - Subclavian
  • - Cephalic
  • Lead / Implantable Pulse Generator (IPG)
    placement site
  • Dominant side / patient activities
  • Anatomy / venous access
  • ? Previous intervention / infection
  • Concommitent therapy ... surgery / ICD
  • Experience of operator

5
Access - Blood Vessels
1 - Anatomy 5
6
Access - Cardiac Chambers
1 - Anatomy 6
7
Pacemaker Implantation
  • Implantation analysis
  • 3 lead ECG monitor / recorder
  • chest lead - unipolar Electrogram (EGM)
  • Electrophysiology amp - bipolar EGM
  • Pacing system analyzer (PSA)
  • Documentation
  • patient demographics
  • medical history
  • IPG data
  • other implant data

Cont/.....
8
Pacemaker Implantation
  • Implantation analysis cont/.....
  • Sterile connector cables
  • IPG programmer software
  • Magnet
  • Sterile accessories
  • Spare batteries, connectors etc.

9
Pacemaker Implantation
Signal Amplitude / Slew Rate
10
Pacemaker Implantation
Signal Frequency
11
Pacemaker Implantation
  • Measurements
  • 1.Sensing threshold
  • Amplitude (Minimum P - 2mV , R - 5mV)
  • Slew rate
  • Frequency content
  • Duration

Cont/.....
12
Pacemaker Implantation
  • Measurements
  • 2.Pacing threshold
  • (Measured amplitude at various pulse widths)
  • Paced rate gt 20 ppm above spontaneous rate
  • Decrement variable output Threshold lt 1V _at_
    0.5ms
  • Impedance measurement 300 - 1000 ohms
  • .... calculated / displayed
  • Intracardiac electrogram
  • .... PSA
  • .... ECG machine
  • .... Monitoring system

Cont/.....
13
Pacemaker Implantation
  • Measurements
  • 3.Lead stability
  • 1V output, or 2X diastolic threshold (min. 1V)
  • 10V output for diaphragm stimulation test
  • Lead sutured in position
  • Re-measure sensing / pacing characteristics
  • Dual-chamber system
  • - Measure Retrograde Conduction Time
  • - Adjust PSA mode
  • Interrogate / Optimise IPG
  • Confirm labelling information

Cont/.....
14
Pacemaker Implantation
  • Measurements
  • 4.Lead Connection
  • Lead attached to IPG, setscrew tightened and
    sealed with cap
  • IPG inserted
  • Ascertain correct sensing / pacing function of
    system. Apply magnet if required.
  • Complete all documentation
  • - pacemaker clinic / hospital records
  • - manufacturers file / central database

15