Acute Respiratory
Distress (COPD, Asthma) ?1. Initiate O2 via
Mask or Cannula according to distress 2. ALS
Crews initiate CPAP 15-25 L/min (5-10 cm H2O)
according to distress. 3. Administer
Albuterol up to 5.0 mg via nebulizer at 8-10 L/
minute. 4. Monitor EKG and Vital
signs. 5. Initiate IV with saline lock or NS
at KVO rate. 6. Contact Medical Control
Pediatric Doses (ages 2-12) Albuterol up
to 5.0 mg via Nebulizer at 8-10 l/min.
7
(No Transcript)
8
Boussignac CPAP System
  • Select the appropriate size face mask for your
    patient
  • - Child Mask (Size 3)
  • - Small Adult Mask (Size 4)
  • - Medium Adult Mask (Size 5)
  • - Large Adult Mask (Size 6)

9
Boussignac CPAP System
  • Set Oxygen flow to deliver CPAP in cmH2O of water
    pressure
  • - 15 liters 5cmH2O
  • - 20 liters 7.5 cmH2O
  • - 25 liters 10 cmH2O

10
Boussignac CPAP System
  • Connect funnel end of Green Oxygen Tubing to an
    O2 source capable of delivering flow up to 25
    liters/min.
  • This provides the full range of CPAP needed for
    clinical use (5-10cmH2O).
  • Typically a D size cylinder in the field.

11
Boussignac CPAP System
  • Insert white end of the Boussignac CPAP into the
    face mask.

12
Boussignac CPAP System
  • If required, insert a pressure manometer between
    the Boussignac CPAP and the face mask.

13
Boussignac CPAP System
  • Explain to the patient how the Boussignac CPAP
    will help their breathing.
  • Gently hold the mask to the patients face
    insuring a good face/mask seal.
  • Turn the flow control device to the desired
    liters/min, generally 15 l/min, to begin the
    CPAP.
  • Gradually adjust the flow to achieve the desired
    level of CPAP.

14
Boussignac CPAP System
  • Secure the Boussignac CPAP System to the patient
    with the head strap.
  • Check around the mask for any leaks.
  • Adjust the mask and/or head strap accordingly.

15
Boussignac CPAP System
  • If the patient requires suctioning of the oral
    cavity, insert French size suction catheter
    through the open end of the Boussignac CPAP
    System.
  • CPAP pressure will not be affected.

16
Boussignac CPAP System
  • CO2 can be monitored with either a nasal cannula
    or an in-line CO2 adapter.

17
Boussignac CPAP System
  • When transporting the patient in the ambulance,
    connect the Green Oxygen delivery tube to the
    wall Oxygen in the vehicle.

18
(No Transcript)
19
Boussignac CPAP System Nebulizer
  • Fill the nebulizer with the prescribed
    medication.

20
Boussignac CPAP System Nebulizer
  • Insert the 22 mm male end of the nebulizer into
    the face mask.
  • Insert the 22 male end (white) of Boussignac CPAP
    into the nebulizer.

21
Boussignac CPAP System Nebulizer
  • Connect the funnel end of the nebulizers gas
    delivery tube to a gas source.
  • Turn on gas source to 6 liters/min. to power the
    nebulizer.
  • Maintain O2 flow to Boussignac CPAP System at 15
    l/min. (5cmH2O).

22