Title: PowerPoint Presentation Templates
1A new method to measure 10B uptake in lung
adenocarcinoma in hospital BNCT
E. Capelli, P. Chiari
F. Basilico, P. Mauri
A. Zanini
D. Bolognini, E.M. Donegani, A. Mattera, M.
Prest, S. Hasan
G. Giannini, E. Vallazza
P. Cappelletti, M. Frigerio, S. Gelosa, A. F.
Monti
F. Borasio
2Outline
- BNCT
- The INFN PhoNeS project
- An innovative approach to Boron Imaging
- Neutron Autoradiography
- Measurements and results
- Conclusion
3Boron Neutron Capture Therapy
10B (n,a,?) 7Li
binary technique
high LET particles
- BNCT could become useful for
- highly-fatal tumours
- Extended (e.g. liver, lung)
- Radioresistant (melanoma)
- Localized near vital organs (glioblastoma)
4Difficulties for BNCT applications
Low toxicity High specificity Non-persistence
in the bloodstream
High intensity thermal-epithermal neutron
beam (5108 n/cm2 s1) lt 10 keV energy range
Gamma contamination lt 2 10-13 Gy cm2 / n
BPA
BSH
5Neutron Autoradiography
- Nuclear reactor
- Solid state track detector (CR 39)
- Sometimes cryogenic temperature during
- the irradiation
- Optic microscope for identification and count
- Hospital-based neutron source
- Real time detector
6 PhoNeS (PhotoNeutronSource)
Neutron flux in the treatment cavity F (1.3
0.1) x 107 n cm-2 s-1
?-contamination reduced of 97
7 Real-time Neutron Autoradiography
0.5 cm CERROBEND photonconversion
6 cm PMMA thermalization
- IRON BOX
- Samples
- Detector
Neutron flux F (1.5 0.1) x 105 n cm-2 s-1
8The imaging detector
Silicon Dimensions (9.5 x 9.5) cm2
Hamamatsu Microstrip Silicon Detector 768 strips
Readout pitch 242 µm
Read-out by 3 self-triggering TAA1 ASICs
Counts vs position
9The Data Acquisition System
- UNDEPLETED silicon detector
to detect a-particles from boronated samples down
to 25 ng of 10B
18 MeV electrons Bunches of 5 µs Frequency 100 -
300 Hz
10The samples
SOLUTION HOMOGENEITY
SPECIMEN SUPPORT
to prevent crystallization and self-absorption
of the a-particles by the sample itself
- plastic support (liquid solution)
- evaporation
- glass slide (blood samples)
- coagulation process
KNOWN VOLUME
Drops of 5 µl
11Calibrating with BSH
Integrated Spot Content (AU) vs Concentration
(ppm)
500 MU/min 18 MV photons Varian CLINAC 2100
C/D, SantAnna Hospital (Como)
12The first ever kinetic curves
Blood and urine of a patient administered BSH,
taken at different times with respect to the
administration.
URINE
BLOOD
rescaled at the max value obtained by
Si-autoradiography
Special thanks to Dr.Sauerwein
13Ex-vivo lung perfusion
in collaboration with S.Luigi Hospital in
Orbassano (TO), Italy
- Human lung affected by adenocarcinoma
- therapetic resection
- perfusion of blood and BPA
- Blood samples taken at different times
- to estimate 10B uptake
14Kinetic curves
- Blood samples doped with BPA
- Comparison between real-time
autoradiography to MS analysis
- The absorption takes place only in the first
90mins
15CR 39
10B concentration in tumour and healthy tissues
of resected lobe
High-resolution images to reproduce the features
of the tissues
Unhealthy tissue
CT / CH 1.5 - 2
Healthy tissue
Special thanks to Dr. Giannini
16Conclusions
- Neutron Autoradiography in a hospital environment
successfully performed - Linac - Development of an hospital-base neutron source
to perform Boron Imaging with a 1D spatial
resolution - Real-time detector to perform pharmacokinetic
studies
results during the irradiation
- Blood samples from a perfused lung affected by
adenocarcinoma kinetic curves - 10BPA down to
concentrations of 5ppm - Evaluation of 10B concentration in tumour and
healthy tissues
Si-autoradiography CR 39 - Looking forward to make BNCT possible in every
radiotherapic unit
17A new method to measure 10B uptake in lung
adenocarcinoma in hospital BNCT
E. Capelli, P. Chiari
A. Zanini
F. Basilico, P. Mauri
Thank you!
G. Giannini, E. Vallazza
F. Borasio
D. Bolognini, E.M. Donegani, A. Mattera, M.
Prest, S. Hasan
P. Cappelletti, M. Frigerio, S. Gelosa, A. F.
Monti