Analog to Digital Conversion - PowerPoint PPT Presentation

1 / 11
About This Presentation
Title:

Analog to Digital Conversion

Description:

Analog to Digital Conversion 12 bit vs 16 bit A/D Card Nyquist Theorem Applied to EMG Sampling The middle graph shows that all of the EMG data are below 500 Hz ... – PowerPoint PPT presentation

Number of Views:327
Avg rating:3.0/5.0
Slides: 12
Provided by: MarkR99
Category:

less

Transcript and Presenter's Notes

Title: Analog to Digital Conversion


1
Analog to Digital Conversion
2
12 bit vs 16 bit A/D Card
Input Volts A/D 12 bit 212 4096 10
Volts 4095 0 Volts 2048 -10 Volts
0
Input Volts A/D 16 bit 216 65536 10
Volts 65536 0 Volts 32768 -10 Volts
0
16 bit cards have much higher resolution.
3
Nyquist Sampling Theorem
  • If you sample the signal to the left at too slow
    of a rate you will not know if the point off the
    curve is valid or noise.
  • According to the Nyquist Sampling Theorem you
    must sample at least twice as fast as the highest
    frequency of the signal.

4
Nyquist Theorem Applied to EMG Sampling
  • The middle graph shows that all of the EMG data
    are below 500 Hz, therefore you should sample EMG
    at 1000 Hz twice the highest frequency in your
    signal.

5
Analog Signals
  • You can understand why CDs have such high
    fidelity if you understand the analog-to-digital
    conversion process better.
  • Let's say you have a sound wave, and you wish to
    sample it with an ADC.
  • Here is a typical wave (assume here that each
    tick on the horizontal axis represents
    one-thousandth of a second)
  • When you sample the wave with an
    analog-to-digital converter, you have control
    over two variables
  • The sampling rate - Controls how many samples are
    taken per second
  • The sampling precision - Controls how many
    different gradations (quantization levels) are
    possible when taking the sample

6
Slow Sampling Rate
  • In the following figure, let's assume that the
    sampling rate is 1,000 per second and the
    precision is 10
  • The green rectangles represent samples.
  • Every 1000th of a second, the ADC looks at the
    wave and picks the closest number between 0 and
    9.
  • The number chosen is shown along the bottom of
    the figure.
  • These numbers are a digital representation of the
    original wave.
  • When the DAC recreates the wave from these
    numbers, you get the blue line shown in the
    following figure

7
Higher Sampling Rates
  • You can see that the blue line lost quite a bit
    of the detail originally found in the red line,
    and that means the fidelity of the reproduced
    wave is not very good.
  • This is the sampling error.
  • You reduce sampling error by increasing both the
    sampling rate and the precision. In the following
    figure, both the rate and the precision have been
    improved by a factor of 2 (20 gradations at a
    rate of 2,000 samples/sec)
  • In the following figure, the rate and the
    precision have been doubled again (40 gradations
    at 4,000 samples/sec)
  • You can see that as the rate and precision
    increase, the fidelity (the similarity between
    the original wave and the DAC's output) improves.
  • In the case of CD sound, fidelity is an important
    goal, so the sampling rate is 44,100 samples per
    second and the number of gradations is 65,536. At
    this level, the output of the DAC so closely
    matches the original waveform that the sound is
    essentially "perfect" to most human ears.

8
National Instruments PCI-6224 New Dell Computer
  • 16 bit resolution
  • 32 single-ended inputs
  • Max Sampling rate of 250 K Samples/sec

9
National Instruments PCI-6251 old Dell
  • 16 bit resolution
  • 16 channels
  • 1.25 M Samples/s
  • 7 different voltage ranges
  • Voltage triggered sampling

10
National Instruments PCI-6229
  • 16 bit resolution
  • 250 KHz samples/s
  • Input Range
  • 10 V
  • 5 V
  • 1 V
  • 0.2 V

11
Keithley Metrabyte 1802 HC
  • Key Features and Benefits
  • Maximum sample rate of up to 333kS/s
  • 12-bit inputs 64 single-ended or 32 differential
  • 4 digital inputs, 8 digital outputs
  • 2 analog outputs
  • High and low programmable gains
  • Extensive triggering options
  • DriverLINX and TestPoint software drivers
  • LabVIEW VIs
  • The KPCI-1802HC features low-gain inputs of 1, 2,
    4, and 8.
  • These boards feature continuous, high-speed,
    gap-free data acquisition. Sample any single
    channel at any gain up to 333kS/s. Multiple
    channels can be sampled at aggregate rates of up
    to 312.5kS/s.
  • The KPCI-1801HC/1802HC boards feature flexible
    clocking, triggering, and gating modes and
    provide 4 data transition methods Bus mastering,
    Interrupt mode, Target mode, and Programmable
    Burst mode.
Write a Comment
User Comments (0)
About PowerShow.com