Title: Image Artifacts


1
Image Artifacts
  • Chapter 8
  • Bushong

2
Image Artifacts
  • Voluntary and involuntary patient motion can
    result in a motion artifact
  • The motion artifact appears as streaks or
    step-like patterns at high contrast edges
  • Respiratory motion artifacts in computed
    tomography angiography can simulate vascular
    stenosis or aneurysm
  • Metal artifacts occur because the x-ray
    absorption results in incomplete projection
    profiles

3
Image Artifacts
  • Metal in tissue gives rise to streak and star
    shaped artifacts
  • With penetration of cranial bone, the x-ray beam
    is selectively filtered and hardened
  • The hardened x-ray beam presents a false low
    linear attenuation coefficient and false low CT
    number

4
Image Artifact
  • The bean hardening artifact appears as a dark
    ring inside cranial bone and cupping at the
    center of the image
  • When an object (calcification) is not fully
    within a slice thickness, the CT number
    representing that object will be false
  • Partial volume artifacts can be reduced by
    overlapping scans but that increases the patient
    dose

5
Image Artifacts
  • Partial volume artifacts can be reduced by using
    thinner slice thickness but at the expense of a
    higher image noise and/or patient dose
  • Partial volume artifacts can be reduced in spiral
    CT by moving the plane of reconstruction
  • Multiple reconstruction along the z-axis during
    spiral CT reduces partial volume artifacts
  • Ring artifacts can occur in third-generation CT
    imagers because of detector malfunction

6
Image Artifacts
  • Partial volume artifacts can be reduced by
    combining several thin slice reconstructed images
  • Pulsation artifact is observed in CTA
  • Pulsation artifact can simulate vascular stenosis
  • Pulsation artifact can be reduced by using 360
    degree interpolation

7
Image Artifacts
  • The stair-step artifact is associated with spiral
    CT
  • The stair-step artifact is most apparent on
    inclined vessels during spiral CTA
  • The height of a stair step artifact is
    proportional to couch increment
  • The height of a stair-step artifact is
    independent of collimation or reconstruction
    interval