Title: AIHA-LAP, LLC Webinar
1 AIHA-LAP, LLC Webinar
- Measurement Uncertainty and
- Traceability Policies Review
- Friday, April 30, 2010
2Webinar Outline
- Introduction of Presenters/Moderators
- Ground Rules
- Background
- Whats New
- Basis of the new Policies
- Policy Review - Section by Section
- Discussion of guidance and examples
- What to Expect Beginning April 1
- Questions and Answers
3Webinar Speakers
- Speakers
- Margie Breida
- Ronald Peters (MU)
- Maureen Hamilton (Traceability)
- Moderator
- Sage Morgante
4Ground Rules
- All phone lines will be muted during the
presentation. - Speakers will present information in the slides.
- Questions can be be typed in the chat box during
the presentation. Participants should type in
their name/laboratory and a question. - When the speaker thinks it is appropriate to take
questions, he/she will pause and the moderator
will read the questions. - We will answer as many questions as we can, but
not all questions will be answered.
5Ground Rules (contd)
- If you experience technical problems or you
cannot connect to the call, please contact us at
(703) 846-0739.
6Background
- AIHA-LAP, LLC evaluation for international
recognition (APLAC/IAAC) - MU and traceability changes needed so that we are
more aligned with other accreditation bodies - New policies, guidance and examples replace
previous guidance. - New MU and traceability documents posted on web
site on February 18, 2010. - Became effective April 1, 2010.
- All assessments are now being conducted against
these new policies.
7Whats New?
- Follow links to the 2010 AIHA-LAP, LLC Policy
Modules - Appendix G
- Policy on the Estimation of Uncertainty of
Measurement (Rev. 0, effect. April 1, 2010) - Guidance on the Estimation of Uncertainty of
Measurement (Rev. 1, 2/17/10) - Guidance on Statistical Analysis of EMLAP QC
(Rev. 1, 2/17/10)
8Whats New?
- Appendix H
- Policy on Traceability of Measurement (Rev. 0,
effect. April 1, 2010) - Guidance on Traceability of Measurement (Rev. 1,
2/17/10)
9Whats New?
- Example Excel Workbooks
- CALA Example Internal Calibration Workbook
- Example Chemistry Measurement Uncertainty
Calculations - Example Microbiology Measurement Uncertainty
Calculations
10Important to Remember!
- Policy Documents Include Requirements
- Guidance Documents Include guidance and
recommendations These are NOT Requirements but
include acceptable approaches
11Policy on the Estimation of Uncertainty of
Measurement
- (MU Policy)
- 1. Scope
- 2. References
- 3. Terms and Definitions
- 4. Background
- 5. AIHA-LAP, LLC Uncertainty Policy
- 6. Assessment for Accreditation
- 7. Guidance and Examples
12Basis of MU Policy/Guidance
- References
- ISO/IEC 170252005 - General Requirements for
the Competence of Testing and Calibration
Laboratories - Guide to the Uncertainty of Measurement (GUM)
published by ISO, IEC, CIPM, BIPM, Eurachem, etc - Quantifying Uncertainty in Analytical
Measurement, 2nd Edition, 2000, Eurachem/CITAC - APLAC TC 005 (2006) Interpretation and Guidance
on the Estimation of Uncertainty of Measurement
in Testing, Asia-Pacific Laboratory Cooperation - ILAC Guide 17 Introducing the Concept of
Uncertainty of Measurement in Testing in
Association with the Application of the Standard
ISO/IEC 17025. 2002, ILAC Rhodes, NSW,
Australia. - CALA P19 CALA Policy on the Estimation of
Uncertainty of Measurement in Environmental
Testing.
13Basis of MU Policy
- REQUIREMENTS
- ISO/IEC 17025 Section 5.4.6.2 Testing
laboratories shall have and shall apply
procedures for estimating uncertainty of
measurement. - ISO/IEC 17025 Section 5.4.6.3 When estimating
the uncertainty of measurement, all uncertainty
components which are of importance in the given
situation shall be taken into account using
appropriate methods of analysis.
14Basis of MU Policy/Guidance
- Key Terms and Definitions
- uncertainty of measurement (VIM 2.26 JCGM
2002008) non-negative parameter characterizing
the dispersion of the quantity values being
attributed to a measurand, based on the
information used - NOTE 1 Measurement uncertainty includes
components arising from systematic effects, such
as components associated with corrections and the
assigned quantity values of measurement
standards, as well as the definitional
uncertainty. Sometimes estimated systematic
effects are not corrected for but, instead,
associated measurement uncertainty components are
incorporated. - NOTE 2 The parameter may be, for example, a
standard deviation called standard measurement
uncertainty (or a specified multiple of it), or
the half-width of an interval having a stated
coverage probability.
15Basis of MU Policy/Guidance
- Key Terms and Definitions
- uncertainty of measurement (VIM 2.26 JCGM
2002008) - NOTE 3 Measurement uncertainty comprises, in
general, many components. Some of these may be
evaluated by Type A evaluation of measurement
uncertainty from the statistical distribution of
the quantity values from series of measurements
and can be characterized by standard deviations.
The other components, which may be evaluated by
Type B evaluation of measurement uncertainty, can
also be characterized by standard deviations,
evaluated from probability density functions
based on experience or other information. - NOTE 4 In general, for a given set of
information, it is understood that the
measurement uncertainty is associated with a
stated quantity value attributed to the
measurand. A modification of this value results
in a modification of the associated uncertainty.
16Basis of MU Policy
- REQUIREMENTS
- ISO/IEC 17025 Section 5.10.3.1 In addition to
the requirements listed in 5.10.2, test reports
shall, where necessary for the interpretation of
the test results, include the following - c) where applicable, a statement on the
estimated uncertainty of measurement
information on uncertainty is needed in test
reports when it is relevant to the validity or
application of the test results, when a
customer's instruction so requires, or when the
uncertainty affects compliance to a
specification limit
175.0 AIHA-LAP, LLC Uncertainty Policy
- The requirement which underlies this policy is
given in ISO/IEC 17025, Clauses 5.4.6 and
5.10.3.1 c). - Laboratories accredited under the AIHA-LAP, LLC
Accreditation Program shall fulfil the following
requirements with respect to the estimation of
uncertainty of measurement for tests associated
with their scope of accreditation
18AIHA-LAP, LLC Uncertainty Policy
- 5.1 Laboratories shall be able to demonstrate
their ability to estimate measurement uncertainty
for all accredited quantitative test methods. In
those cases where a rigorous estimation is not
possible, the laboratory must make a reasonable
attempt to estimate the uncertainty of test
results. All approaches that provide a reasonable
and valid estimation of uncertainty are equally
acceptable.
19AIHA-LAP, LLC Uncertainty Policy
- 5.2 Laboratories shall make independent
estimations of uncertainty for tests performed on
samples with significantly different matrices.
For example, estimations made for filter samples
cannot be applied to bulk samples. - 5.3 Estimations of measurement uncertainty are
not needed where the reported test results are
qualitative. Laboratories are, however, expected
to have an understanding of the contributors to
variability of test results. Examples of such
tests are those that report only organism
identifications or presence/absence.
20AIHA-LAP, LLC Uncertainty Policy
- 5.4 Laboratories shall have a written procedure
describing the process used to estimate
measurement uncertainty, including at a minimum - 5.4.1 Definition of the measurand.
21Basis of MU Policy/Guidance
- Key Terms and Definitions
- measurand (VIM 2.3 JCGM 2002008) quantity
intended to be measured - NOTE The specification of a measurand requires
knowledge of the kind of quantity, description of
the state of the phenomenon, body, or substance
carrying the quantity, including any relevant
component, and the chemical entities involved. - NOTE In chemistry, analyte, or the name of a
substance or compound, are terms sometimes used
for measurand. This usage is erroneous because
these terms do not refer to quantities
22AIHA-LAP, LLC Uncertainty Policy
- 5.4 Laboratories shall have a written procedure
describing the process used to estimate
measurement uncertainty, including at a minimum - 5.4.1 Definition of the measurand.
- 5.4.2 Identification of the contributors to
uncertainly.
23MU Guidance Document
- Consider the following categories or processes
as listed by CALA and ILAC Guide 17 - Sampling or sub-sampling in-house sub-sampling
typically applies only to bulk samples tested in
AIHA-LAP, LLC laboratories. Note that field
sampling is most often outside the
responsibilities of the laboratory. In such
cases, the laboratory should clearly state that
any estimates of uncertainty reported with
samples relate only to analytical uncertainty. - Transportation, storage and handling of samples
- Preparation of samples -all steps of the sample
procedure prior to analysis. This can include
variations in drying, grinding, filtering,
weighings, dispensing of materials, extractant
backgrounds, etc.
24MU Guidance Document
- Consider the following categories or processes
as listed by CALA and ILAC Guide 17 - Environmental and measurement conditions - those
conditions that can impact some test results
(e.g., gravimetry, microbiology) when they vary
(e.g., temperature or humidity of the balance
room, seasonal changes in microbiological
background of micro labs, etc.). - The personnel carrying out the tests -This is
especially important with subjective tests such
as microscopy and organism identification. - Variations in the test procedure - for example,
different recoveries for different batches of
media, impurities in reagent lots, the effect of
different extraction, digestion or incubation
times and temperatures, percentages of
microscopic samples read, etc.
25AIHA-LAP, LLC Uncertainty Policy
- Consider the following categories or processes
as listed by CALA and ILAC Guide 17 - The measuring instruments - variations in
baseline drift, day to day calibration
differences, carry over effects, interferences
specific to the test method, microscope
magnification used, etc. - Calibration standards or reference materials -
uncertainty related to reference materials or due
to preparation differences, etc. Uncertainty
estimates may come from certificates of analysis
or estimations based on provider claims. - Methods of generating test results - uncertainty
due to data interpretation (e.g., peak
integration, baseline manipulation, etc.), blank
corrections, differences in how the software was
used, other data manipulations, etc. - Corrections for systematic effects - if test
results are corrected for bias, include the
uncertainty of the correction.
26AIHA-LAP, LLC Uncertainty Policy
- 5.4 Laboratories shall have a written procedure
describing the process used to estimate
measurement uncertainty, including at a minimum - 5.4.1 Definition of the measurand.
- 5.4.2 Identification of the contributors to
uncertainly. - 5.4.3 Details of the approaches used for
estimating measurement uncertainty, such as Type
A and/or Type B.
27Basis of MU Policy/Guidance
- Key Terms and Definitions
- type A evaluation of measurement uncertainty (VIM
2.28 JCGM 2002008) evaluation of a component of
measurement uncertainty by a statistical analysis
of measured quantity values obtained under
defined measurement conditions - NOTE 1 For various types of measurement
conditions, see repeatability condition of
measurement, intermediate precision condition of
measurement, and reproducibility condition of
measurement. - type B evaluation of measurement uncertainty (VIM
2.29 JCGM 2002008) evaluation of a component of
measurement uncertainty determined by means other
than a Type A evaluation of measurement
uncertainty - EXAMPLES Evaluation based on information
- associated with authoritative published
quantity values, - associated with the quantity value of a
certified reference material, - obtained from a calibration certificate,
- about drift,
- obtained from the accuracy class of a verified
measuring instrument, - obtained from limits deduced through personal
experience.
28AIHA-LAP, LLC Uncertainty Policy
- When using the Type A approach, laboratories
shall utilize one or more of the following
options. These options are generally considered
from 1) most suitable, to 4) least suitable - 1) Uncertainty specified within a standard
method. In those cases where a well-recognized
test method (such as a peer-reviewed AOAC, NIOSH,
OSHA, ASTM, etc. method), specifies limits to the
values of the major sources of uncertainty of
measurement and specifies the form of
presentation of calculated results, laboratories
need not do anything more than follow the
reporting instructions as long as they can
demonstrate they follow the reference method
without modification and can meet the specified
reliability.
29AIHA-LAP, LLC Uncertainty Policy
- 2) Laboratory Control Samples (LCS) and Matrix
Spikes. In cases where matrix specific LCS (CRM
or media spikes) and/or matrix spike data are
available, include uncertainty estimated from the
standard deviation of long term data collected
from routine sample runs for existing test
methods or from the standard deviation of the LCS
or matrix spike data for method
validation/verification studies for new test
methods.
30AIHA-LAP, LLC Uncertainty Policy
- 3) Duplicate Data. In cases where sub-sampling
occurs and there are data over the reporting
limit, include uncertainty estimated from long
term duplicate data collected from routine sample
runs for existing test methods or method
validation/verification studies for new test
methods. - 4) Proficiency Testing (PT) Sample Data. In cases
where the previous options are not available and
where PT samples are analyzed with sufficient
data above the reporting limit, pooled PT sample
data can be used to estimate uncertainty.
31AIHA-LAP, LLC Uncertainty Policy
- 5.4 Laboratories shall have a written procedure
describing the process used to estimate
measurement uncertainty, including at a minimum -
- 5.4.4 Identification of the contributors of
variability for qualitative test methods. - 5.4.5 All calculations used to estimate
measurement uncertainty and bias. - 5.4.6 The reporting procedure.
32AIHA-LAP, LLC Uncertainty Policy
- 5.5 Laboratories are required to re-estimate
measurement uncertainty when changes to their
operations are made that may affect sources of
uncertainty. - 5.6 Laboratories shall report the expanded
measurement uncertainty, along with the reported
analyte concentration, in the same units as
analyte concentration, when - it is relevant to the validity or application
of the test results, or - a customer's instructions so requires, or
- the uncertainty affects compliance to a
specification limit.
33AIHA-LAP, LLC Uncertainty Policy
- 5.7 When reporting measurement uncertainty, the
test report shall include the coverage factor and
confidence level used in the estimations
(typically k approximately 2 at the 95
confidence level). - 5.8 When the test method has a known and
uncorrected systematic bias, it shall be reported
separately from the test result and uncertainty
estimation, as a probable bias value
34Basis of MU Policy/Guidance
- Key Terms and Definitions
- Expanded uncertainty (VIM 2.35 JCGM 2002008)
product of a combined standard measurement
uncertainty and a factor larger than the number
one - NOTE 1 The factor depends upon the type of
probability distribution of the output quantity
in a measurement model and on the selected
coverage probability. - NOTE 2 The term factor in this definition
refers to a coverage factor. - NOTE 3 Expanded measurement uncertainty is termed
overall uncertainty in paragraph 5 of
Recommendation INC-1 (1980) (see the GUM) and
simply uncertainty in IEC documents. - combined standard uncertainty (VIM 2.31 JCGM
2002008) standard measurement uncertainty that
is obtained using the individual standard
measurement uncertainties associated with the
input quantities in a measurement model
35Basis of MU Policy/Guidance
- Key Terms and Definitions
- coverage factor (VIM 2.38 JCGM 2002008) number
larger than one by which a combined standard
measurement uncertainty is multiplied to obtain
an expanded measurement uncertainty - NOTE A coverage factor, k, is typically in the
range of 2 to 3. - standard uncertainty (VIM 2.30 JCGM 2002008)
measurement uncertainty expressed as a standard
deviation
36Summary of Guidance Document Steps
- Review and identify the contributors
- Determine if contributors are accounted for with
existing QC data - Compile the applicable QC data and any other
contributors and perform calculation of combined
uncertainty -
37Summary of Guidance Document Steps
- Combined uncertainty
- SDc v SD12 SD22 SDn2
- It may be beneficial to use RSD instead of SD as
it allows for the concentration dependence of SD.
- NOTE Sources that have an SD of less than 1/3
of the largest SD can be eliminated
38Summary of Guidance Document Steps
- Calculate the expanded uncertainty
- Apply the appropriate coverage factor 'k'.
Calculate the expanded uncertainty by multiplying
the combined standard uncertainty by the
appropriate coverage factor (k) to give an
expanded uncertainty with the desired confidence
level. The factor k is the confidence interval
Student distribution t-factor for n-1 degrees of
freedom. For a confidence level of 95, k is
approximately 2 for a data set of 30 points or
more, for normally distributed data sets. - Expanded measurement uncertainty k x SDc
39Summary of Guidance Document Steps
- Reporting test results with the expanded
measurement uncertainty - Total benzene concentration of 88 ug/sample 11
ug/sample at the 95 confidence level (k2). - Where bias is present, report it along with the
uncertainty as a probable bias in a manner such
as the following example - Total lead concentration of 78 ug/sample 12
ug/filter at the 95 confidence level (k2). This
method has an average recovery of 94, or at this
level, a probable bias of -5 ug/filter. - Alternate forms of reporting uncertainty and bias
are acceptable as long as required information is
clearly presented.
406. ASSESSMENT FOR ACCREDITATION
- During assessment and surveillance of a
laboratory, the assessor will evaluate the
capability of the laboratory to estimate the
measurement uncertainty for test methods included
in the laboratorys scope of accreditation. The
assessor will verify that the methods of
estimation applied are valid, all significant
contributors to uncertainty have been considered,
and all the criteria of the AIHA-LAP, LLC policy
are met.
41Guidance and Examples
- Refer to the AIHA-LAP, LLC Guidance on the
Estimation of Uncertainty of Measurement for
suggestions and examples for implementing the
policies and helpful references. - Example Excel spreadsheets also on the website
42Guidance and Examples
- Example Chemistry Measurement Uncertainty
Calculations workbook includes - Common Contributors to Measurement Uncertainty
- Example contributors and calculations for
Chromatography - Example contributors and calculations for Lead in
Paint
43Guidance and Examples
- Example Microbiology Measurement Uncertainty
Calculations workbook includes - Example contributors and calculations for Direct
Examination Air - Example contributors and calculations for
culturable analyses (fungal swabs)
44Summary of Requirements
- Section 5.1 requires laboratories to be able to
calculate an estimated measurement uncertainty,
when requested, for all quantitative test methods
covered by their AIHA LAP, LLC accreditation. - Section 5.2 requires laboratories to treat each
significantly different matrix (e.g., filters vs.
bulks) separately when estimating measurement
uncertainty.
45Summary of Requirements
- Section 5.3 addresses qualitative test methods.
For these methods, laboratories are only required
to understand the contributors to variability of
test results and attempt to minimize them.
46Summary of Requirements
- Section 5.4 requires laboratories to have a
written procedure describing how they estimate
measurement uncertainty. - Procedures must define the measurand, identify
contributors to uncertainty, detail the approach
that will be used to make the measurement
uncertainty estimation, identify the calculations
to be used to estimate measurement uncertainty
and bias, and define how the information will be
reported when required. -
- Types of QC data that may be used in making the
estimates are identified.
47Summary of Requirements
- Section 5.5 identifies conditions that will
require laboratories to re-estimate measurement
uncertainty of a test method. - Section 5.6 defines when and how measurement
uncertainty, as expanded measurement uncertainty,
shall be reported.
48Summary of Requirements
- Section 5.7 requires the coverage factor used and
confidence level to be included when reporting
measurement uncertainty. - Section 5.8 identifies requirements for reporting
bias when it is known and a correction for it has
not been applied to the reported result.
49Beginning April 1, 2010
- Your procedures must include all elements in
Section 5 of the AIHA-LAP,LLC measurement
uncertainty policy. - The procedures must be implemented.
- All contributors to MU must be considered and
dealt with in a defensible manner. - The calculations must make sense.
- Your lab must have the ability to report MU, even
if the customer is currently not requesting it.
50Questions?
51Traceability Documents
- Appendix H
- Policy on Traceability of Measurement (Rev. 0,
effect. April 1, 2010) - Guidance on Traceability of Measurement (Rev. 1,
2/17/10) - CALA Example Internal Calibration Workbook
52Policy on the Traceability of Measurement
- (Traceability Policy)
- 1. Scope
- 2. References
- 3. Terms and Definitions
- 4. Background
- 5. Policy
- 6. Guidance on Implementing this Policy
53Basis of Traceability Policy
- References
- ISO/IEC 170252005 - General Requirements for
the Competence of Testing and Calibration
Laboratories - ILAC-P10 Policy on Traceability of Measurement
Results - ILAC-G24 Guidelines for the determination of
calibration intervals of measuring instruments - CALA A61 CALA Traceability Policy (CALA)
544. BACKGROUND
- Traceability is characterized (in ILAC documents
and the VIM) by 6 basic requirements - (a) an unbroken chain of comparisons going back
to stated references acceptable to the parties,
usually a national or international standard - (b) uncertainty of measurement the uncertainty
of measurement for each step in the traceability
chain must be calculated or estimated according
to agreed methods and must be stated so that an
overall uncertainty for the whole chain may be
calculated or estimated
554. BACKGROUND
- Traceability is characterized (in ILAC documents
and the VIM) by - (c) documentation each step in the traceability
chain must be performed according to documented
and generally acknowledged procedures the
results must be recorded - (d) competence the laboratories or bodies
performing one or more steps in the traceability
chain must supply evidence for their technical
competence (e.g. by demonstrating that they are
accredited for that activity)
564. BACKGROUND
- Traceability is characterized (in ILAC documents
and the VIM) by - (e) reference to SI units the chain of
comparisons must, where possible, end at primary
standards for the realization of the SI units - (f) calibration intervals calibrations must be
repeated at appropriate intervals the length of
these intervals will depend on a number of
variables (e.g. uncertainty required, frequency
of use, way of use, stability of the equipment).
574. BACKGROUND
- AIHA-LAP, LLC accredited laboratories must
understand the following simple relationship. All
three of these components must exist at every
level in the traceability chain in order for the
final test result to be traceable. - Calibration (1) with Uncertainty (2) produces a
measurement result that is Traceable (3)
58Basis of Traceability Policy/Guidance
- Requirements
- ISO/IEC 17025, Clause 5.6
- 5.6.1 All equipment used for tests and/or
calibrations, including equipment for subsidiary
measurements (e.g. for environmental conditions)
having a significant effect on the accuracy or
validity of the result of the test, calibration
or sampling shall be calibrated before being put
into service. The laboratory shall have an
established programme and procedure for the
calibration of its equipment. - 5.6.2 specific requirements for calibration and
testing - 5.6.3 reference standards and reference
materials
595. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- The requirement which underlies this policy is
given in ISO/IEC 17025, Clause 5.6. - 5.1 Laboratories accredited by AIHA-LAP, LLC
shall demonstrate, when possible, that
calibrations of critical equipment and hence the
measurement results generated by that equipment,
relevant to their scope of accreditation, are
traceable to the SI through an unbroken chain of
calibrations.
605. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- 5.2 External calibration services shall, wherever
possible, be obtained from providers accredited
to ISO/IEC 17025 by an ILAC recognized signatory.
Calibration certificates shall be endorsed by a
recognized accreditation body symbol.
Certificates shall indicate traceability to the
SI or reference standard and include the
measurement result with the associated
uncertainty of measurement.
615. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- 5.3 Where traceability to the SI is not
technically possible or reasonable, the
laboratory shall use certified reference
materials provided by a competent supplier (refer
to ISO/IEC 17025 4.6.4), or use specified methods
and/or consensus standards that are clearly
described and agreed to by all parties concerned.
A competent supplier is an NMI or an accredited
reference material provider (RMP) that conforms
with ISO Guide 34 in combination with ISO/IEC
17025, or ILAC Guidelines for the Competence of
Reference Material Producers, ILAC G12.
Conformance is demonstrated through accreditation
by an ILAC recognized signatory.
625. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- NOTE There are many gaps in the measurement
traceability of the calibration infrastructure in
the world and there are a relatively small number
of accredited reference material providers. In
recognition of this situation, AIHA-LAP, LLC will
not require the use of accredited reference
material providers, where applicable, until
January 2012. AIHA-LAP, LLC assessors will, at
present, note any nonconformity with this
requirement of Section 5.3 of this policy as a
suggestion for improvement.
635. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- 5.4 Reference materials shall have a certificate
of analysis that documents traceability to a
primary standard or certified reference material
and associated uncertainty, when possible. When
applicable, the certificate must document the
specific NIST SRM or NMI certified reference
material used for traceability.
645. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- 5.5 Calibrations performed in-house shall be
documented in a manner that demonstrates
traceability via an unbroken chain of
calibrations regarding the reference
standard/material used, allowing for an overall
uncertainty to be estimated for the in-house
calibration. - 5.6 Calibrations shall be repeated at appropriate
intervals, the length of which can be dependant
on the uncertainty required, the frequency of use
and verification, the manner of use, stability of
the equipment, and risk of failure
considerations. Table 5-1 provides the minimum
frequencies that are required.
655. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- 5.7 Periodic verifications shall be performed to
demonstrate the continued validity of the
calibration at specified intervals between
calibrations. The frequency of verifications can
be dependent on the uncertainty required, the
frequency of use, the manner of use, stability of
the equipment, and risk of failure
considerations. Table 5-1 provides the minimum
frequencies that are required.
665. AIHA-LAP, LLC TRACEABILITY OF MEASUREMENT
POLICY
- 5.8 The laboratory shall have procedures
describing their external and internal
calibration and verification activities and
frequencies, and the actions to follow if the
equipment is found to be out of acceptable
specification. - 5.9 Laboratory staff performing in-house
calibrations and verifications shall have
received documented training.
67Table 5-1 Minimum Calibration/Verification
Frequency Requirements for Common Reference
Standards and Support Equipment
Reference Standard / Equipment Calibration Frequency Verification Frequency
Reference Thermometer Initial and every 5 years Not applicable
Working Thermometer Initial and when verification fails Annually
Reference Masses Initial and every 5 years Not applicable
Working Masses NA Initial and then annually
Stage Micrometer Initial and if damaged Not applicable
Balance Initial and following service/repair or when verification fails Each day of use
Mechanical Pipettes Initial and when verification fails Annual
Volumetric Containers for critical functions (non-Class A) Not applicable Each lot prior to use
68Table 5-1 Minimum Calibration/Verification
Frequency Requirements for Common Reference
Standards and Support Equipment
- NOTE It is imperative laboratories understand
that this table is not a list of recommended
frequencies. Rather, they are the minimum
frequencies that will be accepted by AIHA-LAP,
LLC assessors. It is the laboratorys
responsibility to establish a suitable schedule.
69Guidance Document and Examples
- Refer to the AIHA-LAP, LLC Guidance on the
Traceability of Measurement for suggestions and
examples for implementing the policies and
helpful references.
70Guidance Document Includes
- Information to help determine what equipment
needs calibration or verification and the
frequency at which these are needed. - Important considerations when purchasing
reference standards and calibration services,
including content of calibration certificates.
71Guidance Document Includes
- Considerations when performing in-house
calibrations and verifications of various types
of analytical instruments and support equipment. - Tables listing Uncertainty Contributors when
calibrating balances, thermometers and pipettes.
72Guidance Document and Examples
Uncertainty Contribution Table for Thermometers Uncertainty Contribution Table for Thermometers Uncertainty Contribution Table for Thermometers Uncertainty Contribution Table for Thermometers
Contribution (nomenclature) Contribution (nomenclature) Distribution Estimated Value
ur Standard uncertainty of the nominal values of the reference thermometer. Normal Expanded uncertainty on the calibration certificate of the reference thermometer divided by 2 (coverage factor k)
sp Standard deviation of the set of calibration readings Normal Standard deviation of the set of calibration measurements.
u1 Standard uncertainty of the readability and resolution of the working thermometer Uniform (Square) Smallest gradation of the working thermometer divided by v3. Use ONLY if Sp 0
73Guidance Document and Examples
Uncertainty Contribution Table for Pipettes Uncertainty Contribution Table for Pipettes Uncertainty Contribution Table for Pipettes Uncertainty Contribution Table for Pipettes
Contribution (nomenclature) Contribution (nomenclature) Distribution Estimated Value
ur Standard uncertainty of the nominal values of the reference balance. Normal Expanded uncertainty on the calibration certificate of the reference balance divided by 2 (coverage factor k)
Sp Standard deviation of the set of calibration readings Normal Standard deviation of the set of calibration measurements.
ST Standard deviation of corrections caused by temperature (?T) when the temperature differs from standard temperature (20oC). The thermal coefficient of expansion of water is 0.00021 per 1 Celsius at 20 Celsius. Uniform (Square) Relative Standard Deviation (?T x 0.0002) / (v3) in millilitres per millilitre
74Guidance Document and Examples
Uncertainty Contribution Table for Pipettes Uncertainty Contribution Table for Pipettes Uncertainty Contribution Table for Pipettes Uncertainty Contribution Table for Pipettes
Contribution (nomenclature) Contribution (nomenclature) Distribution Estimated Value
u1 Standard uncertainty of the readability and resolution of the working volumetric instrument Uniform (Square) Smallest gradation of the working volumetric instrument divided by v3. Use ONLY if Sp 0
75Additional Guidance Document Content
- Guidance on how to find an ILAC recognized
accredited calibration laboratory. - Guidance on selecting reference material
providers, including sources for identifying
those that are ILAC recognized.
76Guidance Document and Examples
- Example Excel calibration worksheets are also
found on the AIHA website. - Key concept to remember traceable calibration
must also include uncertainty. - Calculate the combined uncertainty
- SDc v SD12 SD22 SDn2
- Calculate the expanded uncertainty
- Apply a coverage factor 2'.
77Summary of Traceability Policy
- Section 5.1 requires laboratories to demonstrate,
when possible, that their analytical results are
traceable to the SI (International System of
Units) through an unbroken chain of calibrations
within the measuring system. This requirement can
be met for weights (masses), balances,
thermometers, volumetric ware (e.g., mechanical
pipettes) and stage micrometers. - Section 5.2 expands upon the required information
that must be included in calibration certificates
received from accredited external calibration
services.
78Summary of Traceability Policy
- Section 5.3 details requirements for providing
traceability of analytical results when
traceability to the SI is not possible (most
chemical and microbiological analyses). Included
are requirements for providers of reference
materials. Because there are currently only a few
accredited reference material producers, use of
these producers will be encouraged where
available but not mandated until 2012.
Traceability options using specified methods or
consensus standards are also provided.
79Summary of Traceability Policy
- Section 5.4 requires certificates of analysis for
reference materials and defines information that
is required on the certificates. -
- Section 5.5 addresses documentation required for
in-house calibrations and requires these
calibrations to include an estimation of the
overall measurement uncertainty. -
- Section 5.6 requires laboratories to establish
the frequency at which calibrations will be
repeated. A table of minimum calibration
frequencies for reference standards and support
equipment is included.
80Summary of Traceability Policy
- Section 5.7 requires laboratories to also
establish a frequency for and perform calibration
verifications to demonstrate the continued
validity of the calibrations. A table of minimum
verification frequencies is provided - Section 5.8 requires laboratories to have
procedures that describe internal and external
calibration and verification activities and their
frequencies. These procedures must also describe
the actions to be taken if equipment is found to
be performing outside of acceptable
specifications. - Section 5.9 requires laboratories to maintain
documentation demonstrating that personnel who
perform in-house calibrations and verifications
have been trained to perform these activities.
81As of April 1, 2010
- Traceability procedure(s) must include all
elements from the AIHA-LAP, LLC policy. - The procedure(s) must be implemented.
- All equipment must be considered and those with a
significant impact on test results should be
dealt with in a defensible manner. - The calculations used follow the MU policy
82Next Steps
- .25 CM points were granted for the March 11 MU
and Traceability Webinar - Will also apply for CM points this Webinar
- We are collecting questions from both Webinars
with the goal of developing a FAQ to post on the
AIHA-LAP, LLC website in late June. - Thank you for participating!
83Questions?