Title: Ionization and Mass Analyzers
1Ionization and Mass Analyzers
- Ionizations
- Electron Ionization - hard
- Chemical Ionization - soft
- MALDI - desorption
- FAB - desorption
- Laser Post Ionization - increase signal!
- Ion Bombardment (SIMS) - soft or hard
- Depends on ion fluence
- Mass Analyzers
- Analyze all mass/charge
- Time-of-flight
- Pre- Select Ions
- Quadropole
- Magnetic or Electric Sectors
2Electron Ionization
- Molecule e-
molecule 2e- - Four electromagnetic poles
- Rapidly oscillating field
- If mass of ion and frequency of oscillation are
comparable then the ion will oscillate towards
the detector and be measured - A certain mass range can be selected depending on
the pole diameter
Quadropole Mass Analyzer
3Electrostatic and Magnetic Sectors
B strength of magnetic field V ion
accelerating voltage r radius of ion
curvature m mass of ion z charge of ion
M 1-50 (change by changing B)
http//www.cea.com/cai/simsinst/m_anal.htm
4Secondary Ion Mass Spectrometry Ionization
Ga
Static SIMS lt 1 x 1013 ions/cm2
5Time-of-Flight Analyzer
e charge on an electron 1.6 x 10-19 C t time
for ion to reach detector m mass v velocity
V stage voltage d length of field free
region V voltage of detector
6TOF-SIMS Instrument
7Laser Postionization (LPI)
Laser Beam
Postionization parameters 800 nm,
150 fs laser pulses, 8.5 x 1012 W/cm2
power density
8TOF-SIMS Molecule-Specific Imaging
Si
Cu
copper grid affixed to silicon substrate with
silver paste
9TOF-SIMS Identification of Plasma Membrane Lipids
DPPE dipalmitoylphosphatidylethanolamine
m/z 682 (MH)
10Are there Lipid Domains in the Inner Leaflet of
the Membrane?
Total Ion DPPE/Cholesterol
250 µm x 250 µm
Intensity
Au
-
A. G. Sostarecz C. M. McQuaw et. al. J. Am.
Chem. Soc. 2004.
C. M. McQuaw A. G. Sostarecz et. al. Langmuir
2005.
11Multi-Isotope Imaging Mass Spectrometry
Atomic Mass Images
14N
nucleus
lamellipodia
endothelial cell
Image from Claude Lechene Harvard Medical School
12MIMS is a Secondary Ion Mass Spectrometry
Technique
B strength of magnetic field V ion
accelerating voltage r radius of ion
curvature m mass of ion z charge of ion
M 1-50 (change by changing B)
http//www.cea.com/cai/simsinst/m_anal.htm
13MIMS uses an Electrostatic and Magnetic Sector
Mass Analyzer
mass separate
focus
14N
ion 1
84
16
Image from Claude Lechene
MIMS analysis is performed as a collaborative
project with Dr. Claude Lechene at the
National Resource for Imaging Mass Spectrometry
at Harvard Medical School