Title: FDR Presentation Personal Health Monitoring Device


1
FDR PresentationPersonal Health Monitoring Device
  • Bassam Noaman
  • Dina El-Eissa
  • Mentor Prof Zhenyu Guo
  • Instructor Prof J Silverman
  • 4/12/00

2
Overview
  • Sensors
  • ECG
  • Blood Pressure
  • Temperature
  • Signal Conditioning
  • Filters, Amplifiers,Logic
  • PIC Microcontroller
  • program interface
  • Palm pilot
  • PC

3
Overall Block Diagram
4
Specification
  • Power Supply 9.0 VDC Battery
  • Operating time 2.8 hours(from Standard 9.0 VDC
    Duracell battery)
  • Inputs
  • Body Temperature Sensor
  • Detect Temperature in degree Celsius
  • Range between 25 C 50.5 C
  • Blood Pressure
  • Range between 0 5 Psi(259mmHg)
  • Accuracy 0.125 _at_ 25 C
  • Manual cuff inflation
  • Maximum systolic pressure 250 mmHg
  • ECG
  • Bandwidth 0.3Hz 150 Hz
  • Detect Heart Rate in bpm
  • One lead ECG

5
Specification
  • Monitor Performance Specification
  • Palm Pilot screen record and plays the patients
    data file on the screen
  • PC screen displays the patient data file
  • Recorded body temperature
  • Systolic and diastolic blood pressure
  • ECG waveform
  • Heart rate
  • Email this data file
  • Interfaces
  • Cables to interface the human body to the device
  • Serial interface from the device to the Palm
  • Serial interface from the palm to PC

6
Specification
  • Indicators
  • LED power status
  • LED blood pressure visual detector
  • Memory Storage
  • 1 MB of Palm is allocated for health monitoring
  • 55.6 minutes of display range

7
Block Diagram
8
Temperature signal conditioner simulation
9
Output voltage of the body Temperature sensor
10
Output voltage of the signal conditioner of the
body temperature sensor
11
Blood pressure measurement
  • Blood pressure
  • is defined as the intra-arterial pressure
    generated by the pulsatic flow of blood through
    an artery.
  • The blood pressure is indicated from two numbers.
  • Systolicrepresents the pressure while the heart
    is beating. maximum value for blood pressure
  • Diastolic represents the pressure when the
    heart is resting between beats.

12
Flowchart for blood pressure measurement

13
Signal Conditioner of Capacitive Microphone
14
ECG Signal Conditioner
15
Simulation of the frequency domain of ECG
16
Microcontroller circuit
17
Software design block diagram
18
Actual Physical Layout
19
Physical layout of the program interface
20
Demonstration
  • Measure current temperature and blood pressure
    and display it on the palm.
  • Display temperature,blood pressure and ECG wave
    in Hyper terminal, then take the data and save it
    in a file. Finally display it on java applet.

21
Conclusion
  • The Palm pilot is not only an organizer, but also
    a device to monitor data(Blood pressure,
    Temperature)

22