Title: SERVO-i WITH HELIOX OPTION
1SERVO-i WITH HELIOX OPTION
2GROWING WITH THE CUSTOMERS NEEDS
SERVO-i System Version 4.0
- Heliox option
- Low MV alarm audio off option
- Possibility to switch off back-up ventilation
- Other usability improvements
-
3SERVO-i WITH HELIOX OPTION
- Latest addition to a wide range of options
4HELIOX GAS PROPERTIES
- Heliox HeO2 is a mixture of helium (He) and
oxygen (O2) (e.g. 80/20 70/30). - Helium is an inert, low density gas (1/7 the
density of air and oxygen). - Lower density leads to a more laminar flow (less
turbulence) and thus reduces resistance in the
airways.
5HELIOX GAS PROPERTIES
- Laminar gas flow (A) is more efficient than
turbulent flow (B).
6HELIOX EFFECTS ON VENTILATION
- Effects of Heliox therapy according to the
literature - Established positive effects on ventilation
- Reduces work of breathing
- Increases CO2 diffusion
- Potential positive effects on ventilation
- Reduces the peak pressure needed to maintain flow
- Increases CO2 elimination and improves gas
exchange - Increases expiratory flow
- Reduces respiratory rate and auto PEEP
- Increases drug deposition in distal parts of the
airways.
7HELIOX USES AND APPLICATIONS
For which patients is Heliox most commonly used?
- Upper airway obstruction
- Lower airway obstruction
8HELIOX POTENTIAL BENEFITS
Potential benefits of Heliox therapy according to
the literature
- Avoiding endotracheal intubation
- Protecting the lungs
- Improving patient outcome
- Enhancing patient comfort
9SERVO-i WITH HELIOX OPTION
General
- Available as an option for the SERVO-i
- Field upgrade
- Factory installed
- The option consists of
- Heliox SW
- Heliox adapter kit
10SERVO-i WITH HELIOX OPTION
Functionality
- Can be used with all pediatric and adult
patients, as well as neonates - Compatible with all invasive/non-invasive
ventilation modes - Supporting mobility from emergency room to MR
environment - Connection of 80/20 79/21 78/22 mixtures
possible - Adjustable O2 concentration (21 to 100).
11SERVO-i WITH HELIOX OPTION
Key benefits of using SERVO-i for Heliox therapy
- Unique solution for automatic gas identification
- Automatic compensation of flow delivery
- Automatic compensation of CO2 and volume
monitoring
12SERVO-i WITH HELIOX OPTION
Key benefits of using SERVO-i for Heliox therapy
- Quick and easy start of therapy
- Quick and easy switch from air to Heliox (and
vice versa) without going to standby - User friendly interface
- Cost efficient Heliox therapy thanks to the
SERVO-is low gas consumption.
13SERVO-i WITH HELIOX OPTION
Heliox adapters
- Four different adapters are available
- DISS (both air (A) and Heliox (B))
- NIST/NIST
- AGA/NIST
- French/NIST
14SERVO-i WITH HELIOX OPTION
Heliox adapters
- Color marked for Heliox
- Brown/green for Heliox DISS
- Brown/white for Heliox NIST
- Attached to ventilator with safety wire
- Attach to air module when using Heliox
15SERVO-i WITH HELIOX OPTION
Connecting Heliox to the ventilator Heliox
adapters NIST / AGA / French
- Remove adapter from holder.
- Connect adapter to Heliox high pressure hose.
- Connect adapter to air module inlet.
16SERVO-i WITH HELIOX OPTION
Connecting Heliox to the ventilator Heliox and
air adapters DISS
- Remove adapter from holder.
- Connect adapter to Heliox high pressure hose.
- Remove air adapter from air inlet.
- Connect Heliox adapter to air inlet.
17SERVO-i WITH HELIOX OPTION
Preparations before using Heliox with SERVO-i
18SERVO-i WITH HELIOX OPTION
Automatic gas identification
- Performed by the ultrasonic expiratory flow
sensors - Available during Pre Use Check and during
ventilation - Active as long as the set O2 concentration is 75
or below - Time for identification is determined by
tidal/minute volume and O2 concentration used
19SERVO-i WITH HELIOX OPTION
Pre Use Check
20SERVO-i WITH HELIOX OPTION
Automatic identification when changing from air
to Heliox during ventilation
21SERVO-i WITH HELIOX OPTION
Gas type change manually initiated compensation
for Heliox gas
22SERVO-i WITH HELIOX OPTION
Gas type change manually initiated compensation
for Heliox gas
23SERVO-i WITH HELIOX OPTION
Gas type change manually initiated compensation
for Heliox gas
24SERVO-i WITH HELIOX OPTION
Things to bear in mind when using Heliox with
SERVO-i
- When system is compensated for Heliox
- Flow trigger sensitivity is automatically reduced
to a level of 1 if set to a more sensitive level. - O2 alarm limits are automatically adjusted from
5 to 7. - When ventilating below 100 ml in infant range, it
is recommended to use additional patient
monitoring.
25SERVO-i WITH HELIOX OPTION
Things to bear in mind when using Heliox with
SERVO-i
- When changing gas type from air to Heliox and
vice versa - When changing from air to Heliox or vice versa,
measured values relating to expiratory flow may
be affected for a short while. - Changing gas during ventilation may trigger O2
concentration high and/or Gas supply pressure low
alarms. - After O2 concentration is changed stabilization
of expiratory measurements may be delayed. - During nebulization there may be a higher
deposition of drugs used compared to when
ventilating with air.
26SERVO-i WITH HELIOX OPTION
Non-invasive application
27SERVO-i WITH HELIOX OPTION
Infant application
28SERVO-i WITH HELIOX OPTION
Order information
Part Description Part Number
Heliox software, SERVO-i 66 75 580
Heliox adapter kit, new SERVO-i DISS NIST/NIST AGA/NIST French/NIST 66 75 585 66 75 582 66 75 590 66 75 587
Heliox adapter kit, onsite upgrade DISS NIST/NIST AGA/NIST French/ NIST 66 77 552 66 77 555 66 55 560 66 77 557
SERVO-i up to serial number 40.000 also requires PC board 1784 rev 16 or later PC board 1785 rev 3 or later 64 67 984 64 68 016
High pressure Heliox hose DISS, 5 m 66 75 592
High pressure Heliox hose NIST, 5 m 66 75 607
- The Heliox option is not compatible with the O2
Sensor and Y Sensor measuring.
29SERVO-i 4.0 OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
30SERVO-i 4.0 OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
31SERVO-i 4.0 OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
32SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
33SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
34SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
35SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Showing previous Pre Use Check/Suction Support
FiO2 retention
36SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Maintenance notifications/Changed calculation of
time for next PM
37SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Previous mode information/Initiating a Pre Use
Check
38SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
New alarm symbol/Clearer difference between
medium and high priority alarms
39SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
New patient category symbols
40SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Time scale on the waveforms