Achieving USP Compliance for PW and WFI Waters - PowerPoint PPT Presentation

1 / 54
About This Presentation
Title:

Achieving USP Compliance for PW and WFI Waters

Description:

Title: Achieving USP Compliance for PW and WFI Waters Author: Todd E Blonshine Last modified by: Z. Meral Maden Created Date: 10/2/2006 5:57:27 PM – PowerPoint PPT presentation

Number of Views:290
Avg rating:3.0/5.0
Slides: 55
Provided by: ToddEBl
Category:

less

Transcript and Presenter's Notes

Title: Achieving USP Compliance for PW and WFI Waters


1
TOC and Conductivity for USP and EP Water Systems
2
Content
  1. Conductivity
  2. TOC analysis
  3. Validating a TOC analyzer
  4. Introduction to the PAT700 combined TOC and
    Conductivity Analyzer

3
Dissociation and Conductivity
  • Conductivity measures ionic activity
  • Ionic compounds dissociate into ions, then
    re-associate back into the compound
  • E.g. Sodium Chloride (NaCl) made from Na and Cl-
  • Add energy and the Na and Cl- split, or
    dissociate
  • The ions then contribute to the conductivity of
    the water
  • Some substances dissociate more readily and
    contribute more strongly to conductivity
  • Different temperature compensation curves for
    different waters

4
Conductivity
Conductivity Meter
Na
Cl-
Measurement Cell
5
Conductivity varies with temperature
Conductivity TOC
Temperature
6
USPlt645gt Conductivity Temperature
  • Conductivity meters will report different results
    depending on
  • The compounds present in the water
  • The temperature compensation curve selected
  • Hence pharmacopoeias call for un-compensated,
    e.g. USPlt645gt

7
Content
  1. Conductivity
  2. TOC analysis
  3. Validating a TOC analyzer
  4. Introduction to the PAT700 combined TOC and
    Conductivity Analyzer

8
On-line TOC analysis principle of operation
Conductivity Meter
Measurement Cell
UV lamp
9
On-line TOC analysis principle of operation
Conductivity Meter
UV
UV
10
On-line TOC analysis - Principle of Operation
Conductivity
Time
UV on
11
Letter of the Law
  • USP (till recently) share the objective of
    completely oxidizing the organic molecules in an
    aliquot of sample water to carbon dioxide
  • EP states have in common the objective of
    completely oxidizing the organic molecules in
    the sample water to produce carbon dioxide
  • Only Hachs, End Point Detection method, strictly
    meets this requirement for online TOC analyzers

12
Dynamic End Point Detection
USP lt643gt Requirement Instrument Must
Completely Oxidize the Sample
13
Content
  1. Conductivity
  2. TOC analysis
  3. Validating a TOC analyzer
  4. Introduction to the PAT700 combined TOC and
    Conductivity Analyzer

14
Validating the TOC analyser
  • Validate that the instrument matches the URS
  • Validate that installation is as per manufacturer
  • Verify Calibration
  • TOC
  • Conductivity
  • Temperature sensor
  • System suitability
  • Verify connections to remote systems
  • Performance qualification
  • Calibration is stable
  • Analyser works correctly

15
Tests to validate the analyser
  • Validate that the instrument matches the URS
  • Validate that installation is as per manufacturer
  • Verify Calibration
  • TOC
  • Conductivity
  • Temperature sensor
  • System suitability
  • Verify connections to remote systems
  • Performance qualification
  • Calibration is stable
  • Analyser works correctly

16
Validation documents procedures
  • Demonstrate URS requirements are met through,
    Commissioning, IQ and OQ
  • PQ demonstrates that the analyzer is stable and
    can support required calibration and system
    suitability intervals

6 monthly testing
17
Tests to validate the analyser
  • Verify that the instrument matches the URS
  • Verify that installation is as per manufacturer
  • Verify Calibration
  • TOC
  • Conductivity
  • Temperature sensor
  • System suitability
  • Verify connections to remote systems
  • Performance qualification
  • Calibration is stable
  • Analyser works correctly

18
TOC Calibration factory on site
Factory Calibration
0 Reported TOC 1000
0 Injected TOC 1000
19
TOC Calibration on site
TOC
20
Certified traceable standards
21
Calibration As Found, As Left
22
System Suitability Test Must Be Performed
  • System Suitability is not calibration!
  • System suitability is the process of validating
    whether your system (i.e. TOC analyzer) is
    acceptable for providing useful analytical data
    without any bias.
  • This is typically done by
  • Analyzing a material that is easy-to-oxidize
    (sucrose)
  • Analyzing a material that is difficult-to-oxidize
    (1,4 - benzoquinone)
  • Calculating the ratio of the responses

23
System suitability USPlt643gt EP2.2.44
24
System suitability calculation
  • Sucrose analysis result 475ppb TOC

Benzoquinone analysis result 485ppb TOC
25
Verifying conductivity USPlt645gt
  1. Verify meter accuracy (/- 0.1mS)
  2. Verify cell constant (/-2)
  3. Verify temperature sensor (compensation)

26
Temperature sensor verification
27
Verifying conductivity meter accuracy
28
Verifying conductivity meter accuracy
29
Verifying conductivity meter accuracy
30
Verifying conductivity cell constant
  • Function area and distance

31
Verifying conductivity cell constant
  • Introduce certified conductivity standard
  • /-2 of certified standard
  • Cell constant known within /-2

32
Content
  1. Conductivity
  2. TOC analysis
  3. Validating a TOC analyzer
  4. Introduction to the PAT700 combined TOC and
    Conductivity Analyzer

33
PAT700 TOC Analyzer with OASISTM
  • Onboard, Automated Standards Introduction System

34
Introduction to the ANATEL PAT700
  • On-line TOC analysis with complete sample
    oxidation
  • OASISTM Onboard, Automated Standards
    Introduction System
  • Calibration and system-suitability test SOP
    built-in via color touch-screen
  • Individual 4-20mA outputs for TOC, temperature
    and conductivity
  • Main and standby auto-switching UV lamps with UV
    Detect technology
  • IP 56 stainless-steel enclosure improves
    protection from water and particulates

35
PAT700 Supports all test requirements
  • USP lt643gt Total Organic Carbon
  • USP lt645gt Conductivity
  • USP lt788gt Water for Injection
  • USP lt789gt Opthalmic Water
  • EP 2.2.38 Conductivity
  • EP 2.2.44 Total Organic Carbon
  • JP lt60gt
  • 21 CFR Part 11

36
OASISTM Technology
  • Saves time and delivers improved confidence in
    quality
  • All bottles for a test installed at once and
    tests automated
  • Information about each standard stored in RFID
    tag on the bottle
  • contents, concentration, certified value, lot
    number, expiration date
  • (Manual data entry possible for customer prepared
    standards)
  • Analyzer voids bottle following test
  • OASIS can be used for grab sample analysis

37
Non-Hach Standards Warning Message
  • If the user selects No, then the test is
    cancelled and the user is placed back at the Run
    Standards selection screen. 
  • If the user selects Yes, then the user is
    prompted to enter all of the bottle information
    manually and the test is allowed to continue as
    if the bottle did not have an RFID tag at all.

38
PAT700 TOC Analyzer with OASISTM
  • Grab-sample analysis

39
Grab Sample Analysis
  • Utilizes OASIS sample bay
  • Up to 4 grab sample bottles at a time
  • Grab Sample Bottle ID nine character alphanumeric
  • (ex. PW POU 01 )

40
PAT700 TOC Analyzer with OASISTM
  • OOS sample capture and root-cause analysis support

41
OOS sample capture capability
  1. Utilizes OASIS sample bay
  2. OOS triggers automatic collection of two samples
  3. Re-tests 1 bottle to confirm OOS
  4. Saves one sample for analysis in laboratory to
    support root-cause analysis

42
PAT700 TOC Analyzer with OASISTM
  • 21 CFR data export

43
21 CFR Part11 Data Export
  • TOC Dataview PC Software
  • Allows the viewing of encrypted data
  • Allows decrypted data to be saved
  • Encrypted file exported as filename.hef
  • Read-only .pdf for record retention
  • .txt files to load into LIMS database

44
PAT700 TOC Analyzer with OASISTM
  • Clean-In-Place verification

45
On-line Monitoring of CIPFinal Rinse Water
Verification
  • Conductivity can be used for detecting acid wash
    residues
  • TOC can be used as broad-spectrum test to detect
    presence of drug material residues
  • Unit programmed to meet users specific criteria
  • Either conductivity threshold
  • Or time based
  • After TOC analysis, reports results and return to
    idle mode

46
On-line Monitoring of CIPFinal Rinse Water
Verification
  • Automated Mode (plc controlled)
  • Unit idle until called for
  • TOC analysis triggered via
  • Digital input, or
  • Serial command, or
  • Modbus over Ethernet
  • Reports results and return to idle mode

47
PAT700 TOC Analyzer with OASISTM
  • Laboratory TOC analysis

48
PAT700 with auto-sampler
  • Autosampler for laboratory use can load up to 36
    vials
  • e.g. point-of-use and rinse water
    samples.(0-2000ppb)
  • Provides a laboratory TOC analyzer without
    reagents or carrier gas

49
PAT700 TOC Analyzer with OASISTM
  • Water loop monitoring and root-cause analysis
    support

50
Dual-stream option PAT700
  • Two inputs
  • One on the PW/WFI loop start, one on the return
  • Brackets loop quality
  • Helps root-cause investigation to find
    contamination source, e.g. feed-water or
    point-of-use

51
Bracketing POU for On-line Release
Contamination Source ?
Points-Of-Use (POUs)
Contamination Source ?
Pure Steam Generation
Distillation - WFI
52
PAT700 TOC Analyzer with OASISTM
  • 12-month service and calibration intervals

53
UV DetectTM Diagnostics
  • Direct measurement of UV source performance
  • Diagnosis of UV status
  • Fast, real-time lamp feedback
  • On-line Verified after every TOC analysis
  • Off-line Can be conducted in diagnostic
    mode after lamp replacement
  • Detects rouge contamination
  • Provides confidence for on-line water release
  • Standby UV lamp automatically switched on if main
    lamp fails avoids gaps in analysis data
  • More reliable than simple hours-of-operation
    counters
  • Reduces cost-of-ownership analyzer can support
    12-monthly calibration and service intervals

54
Questions?
  • PAT700 TOC and Conductivity Analyzer
Write a Comment
User Comments (0)
About PowerShow.com