Title: Achieving USP Compliance for PW and WFI Waters
1TOC and Conductivity for USP and EP Water Systems
2Content
- Conductivity
- TOC analysis
- Validating a TOC analyzer
- Introduction to the PAT700 combined TOC and
Conductivity Analyzer
3Dissociation and Conductivity
- Conductivity measures ionic activity
- Ionic compounds dissociate into ions, then
re-associate back into the compound - E.g. Sodium Chloride (NaCl) made from Na and Cl-
- Add energy and the Na and Cl- split, or
dissociate - The ions then contribute to the conductivity of
the water - Some substances dissociate more readily and
contribute more strongly to conductivity - Different temperature compensation curves for
different waters
4Conductivity
Conductivity Meter
Na
Cl-
Measurement Cell
5Conductivity varies with temperature
Conductivity TOC
Temperature
6USPlt645gt Conductivity Temperature
- Conductivity meters will report different results
depending on - The compounds present in the water
- The temperature compensation curve selected
- Hence pharmacopoeias call for un-compensated,
e.g. USPlt645gt
7Content
- Conductivity
- TOC analysis
- Validating a TOC analyzer
- Introduction to the PAT700 combined TOC and
Conductivity Analyzer
8On-line TOC analysis principle of operation
Conductivity Meter
Measurement Cell
UV lamp
9On-line TOC analysis principle of operation
Conductivity Meter
UV
UV
10On-line TOC analysis - Principle of Operation
Conductivity
Time
UV on
11Letter of the Law
- USP (till recently) share the objective of
completely oxidizing the organic molecules in an
aliquot of sample water to carbon dioxide - EP states have in common the objective of
completely oxidizing the organic molecules in
the sample water to produce carbon dioxide - Only Hachs, End Point Detection method, strictly
meets this requirement for online TOC analyzers
12Dynamic End Point Detection
USP lt643gt Requirement Instrument Must
Completely Oxidize the Sample
13Content
- Conductivity
- TOC analysis
- Validating a TOC analyzer
- Introduction to the PAT700 combined TOC and
Conductivity Analyzer
14Validating the TOC analyser
- Validate that the instrument matches the URS
- Validate that installation is as per manufacturer
- Verify Calibration
- TOC
- Conductivity
- Temperature sensor
- System suitability
- Verify connections to remote systems
- Performance qualification
- Calibration is stable
- Analyser works correctly
15Tests to validate the analyser
- Validate that the instrument matches the URS
- Validate that installation is as per manufacturer
- Verify Calibration
- TOC
- Conductivity
- Temperature sensor
- System suitability
- Verify connections to remote systems
- Performance qualification
- Calibration is stable
- Analyser works correctly
16Validation documents procedures
- Demonstrate URS requirements are met through,
Commissioning, IQ and OQ - PQ demonstrates that the analyzer is stable and
can support required calibration and system
suitability intervals
6 monthly testing
17Tests to validate the analyser
- Verify that the instrument matches the URS
- Verify that installation is as per manufacturer
- Verify Calibration
- TOC
- Conductivity
- Temperature sensor
- System suitability
- Verify connections to remote systems
- Performance qualification
- Calibration is stable
- Analyser works correctly
18TOC Calibration factory on site
Factory Calibration
0 Reported TOC 1000
0 Injected TOC 1000
19TOC Calibration on site
TOC
20Certified traceable standards
21Calibration As Found, As Left
22System Suitability Test Must Be Performed
- System Suitability is not calibration!
- System suitability is the process of validating
whether your system (i.e. TOC analyzer) is
acceptable for providing useful analytical data
without any bias. - This is typically done by
- Analyzing a material that is easy-to-oxidize
(sucrose) - Analyzing a material that is difficult-to-oxidize
(1,4 - benzoquinone) - Calculating the ratio of the responses
23System suitability USPlt643gt EP2.2.44
24System suitability calculation
- Sucrose analysis result 475ppb TOC
Benzoquinone analysis result 485ppb TOC
25Verifying conductivity USPlt645gt
- Verify meter accuracy (/- 0.1mS)
- Verify cell constant (/-2)
- Verify temperature sensor (compensation)
26Temperature sensor verification
27Verifying conductivity meter accuracy
28Verifying conductivity meter accuracy
29Verifying conductivity meter accuracy
30Verifying conductivity cell constant
- Function area and distance
31Verifying conductivity cell constant
- Introduce certified conductivity standard
- /-2 of certified standard
- Cell constant known within /-2
32Content
- Conductivity
- TOC analysis
- Validating a TOC analyzer
- Introduction to the PAT700 combined TOC and
Conductivity Analyzer
33PAT700 TOC Analyzer with OASISTM
- Onboard, Automated Standards Introduction System
34Introduction to the ANATEL PAT700
- On-line TOC analysis with complete sample
oxidation - OASISTM Onboard, Automated Standards
Introduction System - Calibration and system-suitability test SOP
built-in via color touch-screen - Individual 4-20mA outputs for TOC, temperature
and conductivity - Main and standby auto-switching UV lamps with UV
Detect technology - IP 56 stainless-steel enclosure improves
protection from water and particulates
35PAT700 Supports all test requirements
- USP lt643gt Total Organic Carbon
- USP lt645gt Conductivity
- USP lt788gt Water for Injection
- USP lt789gt Opthalmic Water
- EP 2.2.38 Conductivity
- EP 2.2.44 Total Organic Carbon
- JP lt60gt
- 21 CFR Part 11
36OASISTM Technology
- Saves time and delivers improved confidence in
quality - All bottles for a test installed at once and
tests automated - Information about each standard stored in RFID
tag on the bottle - contents, concentration, certified value, lot
number, expiration date - (Manual data entry possible for customer prepared
standards) - Analyzer voids bottle following test
- OASIS can be used for grab sample analysis
37Non-Hach Standards Warning Message
- If the user selects No, then the test is
cancelled and the user is placed back at the Run
Standards selection screen. - If the user selects Yes, then the user is
prompted to enter all of the bottle information
manually and the test is allowed to continue as
if the bottle did not have an RFID tag at all.
38PAT700 TOC Analyzer with OASISTM
39Grab Sample Analysis
- Utilizes OASIS sample bay
- Up to 4 grab sample bottles at a time
- Grab Sample Bottle ID nine character alphanumeric
- (ex. PW POU 01 )
40PAT700 TOC Analyzer with OASISTM
- OOS sample capture and root-cause analysis support
41OOS sample capture capability
- Utilizes OASIS sample bay
- OOS triggers automatic collection of two samples
- Re-tests 1 bottle to confirm OOS
- Saves one sample for analysis in laboratory to
support root-cause analysis
42PAT700 TOC Analyzer with OASISTM
4321 CFR Part11 Data Export
- TOC Dataview PC Software
- Allows the viewing of encrypted data
- Allows decrypted data to be saved
- Encrypted file exported as filename.hef
- Read-only .pdf for record retention
- .txt files to load into LIMS database
44PAT700 TOC Analyzer with OASISTM
- Clean-In-Place verification
45On-line Monitoring of CIPFinal Rinse Water
Verification
- Conductivity can be used for detecting acid wash
residues - TOC can be used as broad-spectrum test to detect
presence of drug material residues - Unit programmed to meet users specific criteria
- Either conductivity threshold
- Or time based
- After TOC analysis, reports results and return to
idle mode
46On-line Monitoring of CIPFinal Rinse Water
Verification
- Automated Mode (plc controlled)
- Unit idle until called for
- TOC analysis triggered via
- Digital input, or
- Serial command, or
- Modbus over Ethernet
- Reports results and return to idle mode
47PAT700 TOC Analyzer with OASISTM
48PAT700 with auto-sampler
- Autosampler for laboratory use can load up to 36
vials - e.g. point-of-use and rinse water
samples.(0-2000ppb) - Provides a laboratory TOC analyzer without
reagents or carrier gas
49PAT700 TOC Analyzer with OASISTM
- Water loop monitoring and root-cause analysis
support
50Dual-stream option PAT700
- Two inputs
- One on the PW/WFI loop start, one on the return
- Brackets loop quality
- Helps root-cause investigation to find
contamination source, e.g. feed-water or
point-of-use
51Bracketing POU for On-line Release
Contamination Source ?
Points-Of-Use (POUs)
Contamination Source ?
Pure Steam Generation
Distillation - WFI
52PAT700 TOC Analyzer with OASISTM
- 12-month service and calibration intervals
53UV DetectTM Diagnostics
- Direct measurement of UV source performance
- Diagnosis of UV status
- Fast, real-time lamp feedback
- On-line Verified after every TOC analysis
- Off-line Can be conducted in diagnostic
mode after lamp replacement - Detects rouge contamination
- Provides confidence for on-line water release
- Standby UV lamp automatically switched on if main
lamp fails avoids gaps in analysis data - More reliable than simple hours-of-operation
counters - Reduces cost-of-ownership analyzer can support
12-monthly calibration and service intervals
54Questions?
- PAT700 TOC and Conductivity Analyzer