Title: JWST OVerview
1The JWST Detector Characterization Project in
the Independent Detector Testing
Laboratory Don Figer, STScI/JHU January 9,
2003 AAS 201 Seattle, WA
2Detectors Are Important for JWST
3JWST NIR Detector Requirements
4- IDTL Role in
- JWST Detector Program
5NIR Detector Characteristics
- Dark current
- Read noise
- Linearity
- Latent charge (persistence)
- Relative and Absolute Quantum efficiency (QE)
- Intra-pixel sensitivity
- Thermal stability
- Radiation immunity
6- Past and present personnel
Eddie Bergeron Data Analyst
Tom Reeves Lab Technician
Robert Barkhouser Optical Engineer
Bernie Rauscher Project Scientist
Utkarsh Sharma Graduate Student
Mike Telewicz Intern
Gretchen Greene Mechanical Engineer
Steve McCandliss JHU Lead
Ernie Morse Data Analyst
Monica Rivera Intern
Scott Fels Intern
Don Figer Director
Sito Balleza Systems Engineer
Mike Regan System Scientist
Russ Pelton Technician
7IDTL
- IDTL provides unbiased testing of detectors and
has been operating for 2 years - Completed Target Field Simulation Testbed project
(18 months, 8/02 end) - Will finish JWST Detector Characterization
project (2 years, 3/03 end) - In pre-proposal phase
- JDC Figer
- 3/03-6/10
- perform optimal-use testing
- develop High Volume Testing (HVT) program
- collaborate on HVT
- perform concurrence testing
- perform anamoly mitigation testing
- perform ground calibration
- TFST Sivaramakrishnan
- 24 months
- produce 'day in the life of' JWST WFSC
- determine what factors impact OPD measurement
- provide regression test data for pipelines, WFSC
- SNAP Figer
- 18 months
8IDTL Test System
Controller Electronics
Dewar
Entrance Window
Vacuum Hose
He Lines
9Hawaii 1R with 5.0 mu Cutoff
10Rockwell Raytheon 2K Devices
11IDTL Measurements Dark Current
- Lowest measured dark current is 0.006
e-/s/pixel.
12IDTL Measurements Read Noise
- Read noise is 10 e- for large number of reads.
(system read noise is 2.5 e-)
13IDTL Measurements Gain
14Hawaii 1R with 5 um Cutoff
15IDTL Test System
16IDTL Measurements AQE
- QE measurements have an accuracy of 10.
17IDTL Measurements RQE
- Relative QE maps show significant variations
across field
18JWST Detectors
HST/NICMOS!
80 megapixels of flight-qualified
detectors!!! (10 times all detectors on