Title: Perchlorate
1Perchlorate
An Integrated Approach to Ensure Reliable
Perchlorate Analytical Results
April 14, 2004
2Agenda
- Perchlorate Chemistry and Sources
- Current Methods for Testing
- Ion Chromatography
- LC/MS/MS
- An Integrated Approach
- A Success Story
- Summary and Conclusions
3Perchlorate Ion
- Perchlorate is a compound containing one chlorine
atom and four oxygen atoms. - Perchlorate is an ion, meaning it carries an
electrical charge, as indicated by the negative
sign.
4Natural Sources of Perchlorate
- Chilean fertilizer deposits
- New Mexican potash
- Canadian potash
- Californian hanksite
- Bolivian playa crusts
5Anthropogenic Sources of Perchlorate
- Rocket Fuel
- Fireworks
- High explosives
- Flares
- Herbicides
- Automobile airbags
- Tracer munitions
- Detergents ?
6How is perchlorate measured?
- EPA Method 314.0
- Ion chromatography/conductivity detector
- Enhanced (Modified) 314.0
- Suppressed Conductivity Detector
- SW-846 Method 9058 (draft)
- Ion Chromatography/conductivity detector
7How is perchlorate measured?
- Other analytical methods that can detect
perchlorate - IC Improved Method 314.1 approved use of 2nd
column confirmation and pre-concentration of
samples. - IC/MS (ion chromatography/mass spectrometry)
Proposed Method 330.0 - LC/MS (liquid chromatography/mass spectrometry)
- LC/MS/MS (liquid chromatography/mass
spectrometry/mass spectrometry) Method 8321
Modified, Proposed Method 331.0
8Ion Chromatography
9Ion Chromatography
- An aqueous sample, comprised of components A and
B, is injected into the ion chromatograph. - The sample enters the analytical column which is
packed with a material that can adsorb the
components of the sample. - Some components are more strongly adsorbed.
Diagram from Fundamentals of Analytical
Chemistry, Skoog, 1988.
10Ion Chromatography
- Fresh solvent is continuously injected into the
instrument. - The solvent carries the sample components, A and
B, through the column. - Component B is more strongly adsorbed onto the
column and is moving thought the column at a
slower rate. - The boundary between A and B is indistinct at
this point.
Detector
Diagram from Fundamentals of Analytical
Chemistry, Skoog, 1988.
11Ion Chromatography
- Component B is almost completely separated from
component A. - The boundary between the two components is more
distinct.
Detector
Diagram from Fundamentals of Analytical
Chemistry, Skoog, 1988.
12Ion Chromatography
- Component B is completely separated from
component A. - Component A exits the column through the detector
at time t3. - The detector records component A as a peak
(pictured below the column) in the sample
chromatogram (graph).
Detector
Diagram from Fundamentals of Analytical
Chemistry, Skoog, 1988.
13Ion Chromatography
Solvent
- Component B exits the column through the detector
at time t4. - The detector records component B as a peak
(pictured below the column) in the sample
chromatogram.
Detector
Diagram from Fundamentals of Analytical
Chemistry, Skoog, 1988.
14Diagram from Fundamentals of Analytical
Chemistry, Skoog, 1988.
15 Identification of Perchlorate by Ion
Chromatography
- 11.2.4 The width of the retention time window
used to make identifications should be based upon
measurements of actual retention time variations
of standards measured over several days. Three
times the standard deviation of retention time
may be used as a suggested window size but the
retention time window should not extend beyond
5 of the retention time for perchlorate. The
experience of the analyst should weigh heavily in
the interpretation of these chromatograms.
EPA Method 314.0
16IC Perchlorate Standard
17High Conductivity
18Carry Over
19Matrix Spike Confirmation
20Matrix Spike Non- Confirmation
21Matrix Spike Non- Confirmation
22Liquid Chromatography/MassSpectrometry/Mass
Spectrometry
23LC/MS/MS
- Leading edge technology
- DoD Recommended
- Pending Regulatory Approval
24(No Transcript)
25API MS/MS
26Bowling Balls and Feathers a simplification of
mass spec separations
27Identification By LC/MS/MS
- Liquid chromatography separates perchlorate from
other sample components in the same manner that
ion chromatography does - Mass spectrometers (MS) are used as detectors in
place of the conductivity detector used in IC. - Unlike conductivity detectors, an MS is a
specific detector
28Detection By LC/MS/MS
- The MS counts the number of ions that have a
specific molecular weight. Perchlorate has a
characteristic molecular weight (MW). - The first MS counts the number of ions that have
the MW of perchlorate 1 Chlorine (35)
4 Oxygens (16) 99 - The second MS removes one of the oxygen atoms and
then counts the number of daughter ions that
have that MW 1 Chlorine (35) 3 Oxygens (16)
83 - Chlorine has a less abundant isotope (MW 37), so
ions with MW 101 are also counted by the first MS
and ions with MW 85 are counted by the second MS. - The retention time of the peak for MW 99 must be
the same as the retention time of the peak for MW
83 for perchlorate to be confirmed. If the
retention times are different, then the two
molecular weight were detected from two different
components and perchlorate is not confirmed.
29Non-detect Analysis by LC/MS/MS
- Analysis by LC/MS/MS
- Reporting limit is 2 ?g/L.
- Bottom chromatogram shows the counts of ions with
molecular weight 83 (1 chlorine 3 oxygens). - Top chromatogram shows the counts of ions with
molecular weight (the isotope of chlorine). - Perchlorate is not detected.
30Comparison of LC/MS/MS
Non-Detect
Detect
31Detect Analysis by LC/MS/MS
- Reporting limit 2 ?g/L.
- Bottom left shows MW 83 (1 chlorine 3 oxygens).
- Top left shows MW 85 (the isotope of chlorine).
- Right chromatogram show the 18O internal standard
(recovery 88). - Perchlorate is detected.
32Perchlorate Decision Tree
33 Success Using the Perchlorate Decision Tree
- Approximately 1000 samples were collected and
analyzed for perchlorate from August 2003
through January 2004. - 89 samples (about 9 of total) were pretreated
and reanalyzed. - 39 samples (about 3 of total) were reanalyzed by
LC/MS/MS - 24 detects and 15 nondetects. - 10 perchlorate detects (about 0.4 of total) were
confirmed by LC/MS/MS 58 false positive rate
and a 0 false negative rate.
34Summary
- Why use 314.0?
- Cost
- Capacity
- Availability
- Promulgated Methods
- When to use LC/MS/MS?
- Definitive Data
- Critical Samples
- Small Projects
35Thank You
Questions and Comments
- Contact
- Elizabeth Wessling at 303.935.6505
- elizabeth.wessling_at_amec.com