Quantitative PCR Session 4: Plate Preparation and Data Analysis - PowerPoint PPT Presentation

About This Presentation
Title:

Quantitative PCR Session 4: Plate Preparation and Data Analysis

Description:

Title: Slide 1 Author: NFSTC Last modified by: Administrator Created Date: 3/1/2006 3:55:59 PM Document presentation format: On-screen Show Company – PowerPoint PPT presentation

Number of Views:175
Avg rating:3.0/5.0
Slides: 67
Provided by: nfs8
Category:

less

Transcript and Presenter's Notes

Title: Quantitative PCR Session 4: Plate Preparation and Data Analysis


1
Quantitative PCR Session 4 Plate Preparation
and Data Analysis
  • Presented by
  • Robert O'Brien
  • Training Specialist Forensic Biology

2
Quantitative PCR
  • Session 4- Plate preparation and Data analysis
  • Importance of pipetting technique with
    Quantifiler
  • Preparing standards
  • Preparing reactions
  • Avoiding and detecting contamination during qPCR
    set up
  • Starting a run
  • Analyzing data
  • IPC
  • Using quantifiler results

3
Importance of pipetting technique with
Quantifiler
  • The standards in quantifiler are the basis by
    which all quantity determinations are made using
    the SDS software.
  • All quantities of samples are calculated based on
    the curve that is made from the standards.
  • Because of this great care must be taken when
    preparing the standards.
  • Most problems associated with quantifiler runs
    and the subsequent data are due to problems with
    standards which are usually because of mistakes
    in pipetting.

4
Importance of pipetting technique with
Quantifiler
  • Tips for improving accuracy and precision when
    pipetting
  • 1) Prewet the Pipette tip
  • Draw in and expel sample at least 3 times before
    drawing in sample for delivery
  • Purpose is to prevent evaporation that can occur
    in the tip. Prewetting the tip increases the
    humidity within the tip and therefore reduces the
    amount of sample evaporation

5
Importance of pipetting technique with
Quantifiler
  • 2) Work at room temperature
  • Allow liquids and equipment to equilibrate to
    room temperature
  • Volume of sample delivered by air displacement
    pipettes varies with air pressure, relative
    humidity and vapor pressure of the liquid, all of
    these are affected by temperature. Working at a
    single constant temperature minimizes this
    variation

6
Importance of pipetting technique with
Quantifiler
  • 3) Examine the tip before dispensing the sample
  • Look at the outside of the tip to see if there is
    any liquid present.
  • Liquid on the outside of the tip will increase
    the amount of liquid being dispensed. The outside
    of the tip can be wiped before delivery of tip
    contents but this must be done carefully because
    if the material being used to wipe the tip
    contacts the tip opening it can absorb the liquid
    inside of the tip

7
Importance of pipetting technique with
Quantifiler
  • 4) Use Standard Mode pipetting
  • Choose standard mode pipetting rather than
    reverse mode pipetting
  • Standard mode pipetting Push button to first
    stop of pipette then place pipette tip in liquid
    approximately 4mm below surface. Slowly release
    button to draw up liquid.
  • To dispense slowly push button to first stop and
    then push all the way to second stop.
  • Speed of drawing up and dispensing liquid should
    be the same

8
Importance of pipetting technique with
Quantifiler
  • Reverse mode pipetting Push button to second
    stop to draw up liquid and only push to first
    stop to dispense. (Reverse mode is used for
    liquids with a high viscosity)

9
Importance of pipetting technique with
Quantifiler
  • 5) Pause consistently after drawing up liquid.
  • Pause with the tip in the liquid for one or two
    seconds after drawing up the sample. The amount
    of liquid in the tip bounces slightly when the
    plunger stops. Slow even plunger release and a
    consistent brief pause after drawing up the
    liquid minimizes errors resulting from the
    bouncing

10
Importance of pipetting technique with
Quantifiler
  • 6) Pull the Pipette straight out
  • After drawing up the sample pull the pipette
    straight out.
  • Do not allow the tip to touch the sides of the
    container.
  • With very small volumes lt 50µl surface tension
    effects cause the sample volumes to vary if the
    exit angle varies. Touching the tip against the
    sides results in loss of sample.

11
Importance of pipetting technique with
Quantifiler
  • 7) Minimize handling the pipette tip
  • Avoid handling the tip and set the pipette down
    between sample deliveries. Body heat transferred
    to equipment during handling disrupts temperature
    equilibrium. Already stated was that temperature
    can affect volume of sample delivered.
  • In forensic science work, the pipette tip should
    never be handled.

12
Importance of pipetting technique with
Quantifiler
  • 8) Immerse the tip to the proper depth.
  • Immerse the tip 2 to 5mm below the meniscus and
    well clear of the container walls and bottom when
    drawing up the sample.
  • Inserting the tip too far into the liquid causes
    excess droplets of liquid to cling to outside of
    the tip. Pressing or resting the tip against the
    container walls or bottom restricts entry of
    sample

13
Importance of pipetting technique with
Quantifiler
  • 9) Use the correct pipette tip.
  • Securely attach a tip designed for use with the
    pipette. Mismatching a tip and pipette or using
    poor quality tips can result in an inadequate
    seal between the pipette and the tip. Quality
    tips are flexible and have thin walls providing
    an airtight seal and more dependable delivery of
    the sample.

14
Importance of pipetting technique with
Quantifiler
  • 10) Use constant plunger pressure and speed.
  • Depress and release the plunger smoothly with
    consistent pressure and speed for each sample.

15
Preparing standards
  • There are 8 standards that are run in duplicate
    on every plate. The highest concentration is
    50ng/µl and the lowest concentration is
    0.023ng/µl.
  • The first standard is made with the DNA standard
    provided by Applied Biosystems that comes with
    the Quantifiler kit.
  • To this standard is added TE buffer or a TE
    buffer glycogen mixture.
  • (glycogen is added to the TE so the standards
    will last longer once made)

16
Preparing standards
  • From this initial standard and TE buffer mixture
    subsequent standards are made following a
    dilution series.

17
Preparing standards
  • The H standard in Human kits contains 14 to 16
    copies of a diploid single copy locus.
  • The H standard in the Y kit contains 7 to 8
    copies of a haploid single copy locus.

18
Preparing reactions
  • Each well in the 96 well plate must have 23µl of
    PCR primer and reaction mix and 2µl of sample to
    give a total volume of 25µl.
  • The primer/reaction mix is made up with 10.5 µl
    of quantifiler Human Primer (or Quantifiler Y
    primer depending on assay being conducted) and
    12.5 µl of quantifiler reaction mix per well.
  • When calculating how much of each to use make
    sure to include samples, standards and controls
    and a few extra for pipetting error.

19
Preparing reactions
  • First step is preparation of Standards
  • Label 8 tubes for the standard series
  • Vortex quantifiler human standard
  • Pulse spin
  • Prepare serial dilutions using the quantities in
    the table previously shown
  • Remember the standards are done in duplicate for
    each assay.
  • During the preparation of standards great care
    must be taken when doing the dilutions. Change
    pipette tip when going from one standard to the
    next during dilutions.

20
Preparing reactions
  • Before making master mix swirl the reaction mix
    by hand. Do not vortex the reaction mix bottle.
  • Vortex the primer mix and pulse spin.
  • Make up the master mix, 10.5µl of primer mix and
    12.5µl of reaction mix per well.
  • Briefly vortex the master mix and pulse spin
  • Dispense 23µl of master mix into each well being
    used

21
Preparing reactions
  • To the appropriate wells add 2µl of standards
    that were made and to the other wells the samples
  • Seal the plate with an optical adhesive cover,
    sealing all edges and between well spaces with
    sealing tool
  • Remove any air bubbles that are in sample wells
    by centrifuging the plate. This is necessary
    because bubbles in plate can show up as small
    amounts of DNA when there is no DNA present.

22
Avoiding and detecting contamination during qPCR
set up
  • Avoiding contamination during Quantitation
  • Even though the 2µl of sample being used for
    quantitation will not be carried to STR
    amplification it is still important to avoid
    contamination of one sample with another.
  • If two samples are mixed during set up of the
    plate then the quantities will be wrong and the
    incorrect amounts of sample will be taken on to
    the amplification process.
  • There can be cross contamination from one sample
    to another, contamination from an outside source
    from the person setting up the plate or from
    amplified DNA

23
Avoiding and detecting contamination during qPCR
set up
  • Avoiding contamination during Quantitation
  • Ways to prevent sample to sample contamination
  • 1 Clearly label extraction tubes and only have
    one tube open at a time when loading the plate.
  • Arrange extraction tubes in the same order as
    they are going into the plate and move the
    extraction tubes after the sample is loaded to a
    different location to avoid the same sample being
    used twice.
  • 3 Strip caps can be used to cover a row of wells
    on the plate after they are loaded. This prevents
    loading the same row of wells twice. It also
    prevents passing over loaded wells with a tip
    that contains DNA.

24
Avoiding and detecting contamination during qPCR
set up
  • Avoiding contamination during Quantitation
  • Ways to prevent sample to sample contamination
  • 4 When loading samples load them as close to the
    bottom of the well as possible and do not
    depress the pipette any further after it is
    pulled out of well.
  • 5 Once the plate is loaded and sealed with
    optical adhesive cover do not tilt or turn the
    plate in such a way that would cause the samples
    to get onto the cover.

25
Avoiding and detecting contamination during qPCR
set up
  • Preventing contamination from an outside source
  • Always wear PPE while loading plates. Especially
    face masks and gloves.
  • Ensure the area where the plate is being set up
    in has been cleaned with bleach solution before
    plate set up.
  • Have a separate room for plate set up and where
    the instrument is located. The instrument is
    amplifying DNA so it should be in a room
    dedicated for amplification. Set up should be in
    a room where there is not amplified DNA.

26
Avoiding and detecting contamination during qPCR
set up
  • Detecting contamination
  • An optional blank can be loaded into the plate
    during plate set up.
  • This is different from the extraction blank
    because this will specifically detect if there
    was any contamination introduced during plate set
    up or if the plate was contaminated before being
    used.
  • The most common type of contamination problem
    with setting up plates is well to well
    contamination because unlike tubes it is
    difficult to cap each well after it is loaded.
  • It is important to keep an eye on your results.
    If a sample that you expected results gives you a
    negative and the sample next to it gave extremely
    high results then it is possible that a well was
    double loaded and the next well was skipped.

27
Avoiding and detecting contamination during qPCR
set up
  • Detecting contamination
  • Another indication of doubling of samples in a
    well is if a sample is run for both Total Human
    and Y and the Y gives a much higher result than
    the Total Human then it is possible that the Y
    well was double loaded.
  • A lot of ways to detect contamination comes from
    thinking about results obtained from quantitation
    and determining if the results make sense. This
    will become easier with experience.

28
Starting a run
  • Instrument operation
  • Turn on the instrument and allow it time to start
    up
  • Open the AB Prism 7000 or 7500 SDS software,
  • set up the plate document for the sample run.
    Usually there will be a template document that
    can be opened so only sample names will need to
    be filled in.
  • Make sure the correct template is selected.
    Templates may vary based on the assays being
    performed (Total human only or Human and Y
    combination plate).

29
Starting a run
  • Once the sample names are filled in on the Plate
    document, then check the thermal cycler
    conditions.
  • Ist step should read 95ºC for 10 minutes
  • 2nd step 40 cycles at 95ºC
  • 3rd step hold cycle for one minute at 60ºC
  • Verify that the sample volume is set to 25µl and
    that the 9600 Emulation box is checked.
  • Save the Plate document when you have finished

30
Starting a run
  • For 7000 place a compression pad on top of sealed
    plate, gray side down directly over the reaction
    wells.
  • For the 7500 no compression pad is needed.
  • Place the plate in the instrument and close the
    door and start the run.

31
Analyzing the Data
  • Once the run is complete the plate must be
    analyzed
  • First verify all the analysis settings
  • Threshold set at 0.20000
  • Baseline start cycle 6
  • Baseline end cycle 15
  • If any of these are different they must be
    changed before analysis. If after plate is
    analyzed it is noticed these values are not
    correct then they can be changed and the plate
    reanalyzed.

32
Analyzing the Data
  • To view analysis settings before analyzing plate
    under Analysis go to Analysis settings the screen
    should be the same as this one.

33
Analyzing the Data
  • If the plate is already analyzed then check here
    for analysis settings

34
Analyzing the Data
  • To analyze plate go to Analysis and select
    analyze or click green arrow shown below

35
Analyzing the Data
  • After analysis of plate is complete the following
    should be reviewed
  • The standard curve
  • The amplification plot
  • The report

36
Analyzing the Data
  • The standard curve
  • The standard curve gives the following
    information
  • The R² value of regression line which shows how
    well the line fits with values from standards
  • The slope which gives an indication of efficiency
  • The Y- intercept
  • To select the standard curve under Results tab
    got to Standard curve verify slope for human is
    between
  • -2.9 to -3.3 and for Y -3.0 to -3.6
  • R² should be gt 0.98

37
Analyzing the Data
38
Analyzing the Data
  • The values for the R² , and slope should be in
    the ranges shown for the data to be used.
  • If these values are not within range then there
    may have been a problem when setting up the
    standards that made the standard curve. This
    reinforces how careful one must be when making
    standards since if these values are not in range
    the plate may not be able to be used and the
    procedure may have to be repeated.

39
Analyzing the Data
  • Only with the standards for a Y plate can an
    adjustment be made.
  • The STD 8 or H for the Y kit only has 7 to 8
    copies per 2µl of the haploid target locus.
  • Because of stochastic effects when using such a
    low amount of DNA the CT values for this
    concentration of DNA are more variable. This
    variability can throw off the line of best fit
    and this in turn will throw off R² value.
  • So the STD 8 or H for the Y kit can be dropped by
    omitting the wells and reanalyzing.

40
Analyzing the Data
  • To omit well select the well then go to Well
    inspector and click on omit well

41
Analyzing the Data
  • After well is omitted under the Plate tab the
    well should be crossed out. Notice that the arrow
    is green indicating reanalysis is necessary.

42
Analyzing the Data
  • Troubleshooting Amplification Plot
  • R² value lt 0.98
  • Quantity values entered for quantitation
    standards in well inspector are incorrect
  • Serial dilutions were not made properly
  • Standards were not loaded correctly
  • Reaction for one of the standards failed
  • Slope not in range
  • Assay not set up correctly
  • Software not set up correctly
  • Reagents may be expired or wrong ones used
  • Instrument may malfunction

43
Analyzing the Data
  • Y intercept
  • There is no fixed value for the Y intercept
  • The Y intercept is the expected CT value for a
    sample that has a quantity of 1ng/µl
  • The Y intercept may shift from one kit lot number
    to the next. The Y intercept from all runs
    performed with a certain kit lot number should be
    within an acceptable range of the intercept value
    when the kit was quality checked. The range is
    lab dependant.

44
Analyzing the Data
  • After the plate is analyzed then the
    amplification plot is checked to see what samples
    gave results and which were negative for DNA.
  • The amplification plot is a plot of ?Rn vs Cycle
    number.
  • This screen shows the corresponding wells for the
    samples. Each individual sample can be checked to
    see if the sample crossed the Threshold or not.
    Samples not crossing the threshold are negative
    no DNA present.
  • Also the IPC can be checked for all of the
    samples.
  • All IPCs should cross unless inhibition occurred.

45
Analyzing the Data
46
Analyzing the Data
  • The individual wells can be selected one at a
    time
  • Also putting the cursor on a plot will show which
    sample it corresponds too.
  • Under Detector there is a drop down menu to
    select.
  • ALL, IPC, Quantifiler Human, or Total Y

47
Analyzing the Data
48
Analyzing the Data
  • To see either the Quantifiler Human or IPC by
    itself then go to Detector drop down box and
    select what plot you would like to view

49
Analyzing the Data
  • The report
  • The report will show the quantities calculated
    for the samples.
  • The IPC well has no value under quantity because
    no quantity calculations are made for IPC,
  • If more than one sample has the same name then
    they are grouped together by the software and the
    Mean value and Standard deviation for the samples
    are calculated.
  • During practice plates duplicating samples is a
    good way to test precision of pipetting technique.

50
Analyzing the Data
51
IPC
  • IPC Internal PCR control
  • IPC is included as part of each reaction well. It
    undergoes the same PCR reaction as the sample but
    because its target sequence does not exist in
    nature it does not compete with the primers and
    probes that target human DNA.
  • However because IPC is undergoing PCR along with
    the sample anything that negatively affects PCR
    of sample will also affect PCR of IPC.

52
IPC
  • Since the IPC is affected by problems with PCR
    that affect the sample it can be used as a
    quality control check.
  • IPC can detect the following
  • Presence of inhibitors
  • Problems with Assay set up
  • Chemistry or Instrument failure

53
IPC
  • If the IPC crosses the threshold and the sample
    does not then the sample is considered a true
    negative.

54
IPC
  • The CT range for the IPC VIC dye should be
    between 20 to 30,

55
IPC
  • If both the Human specific target and the IPC
    fail to amplify it is not possible to determine
    if the negative is due to no DNA present to
    amplify or if it is due to something affecting
    the PCR reaction like inhibition.
  • If inhibition is suspected then the sample can be
    cleaned.
  • Cleaning can be done by concentrating the sample.
    Some inhibitors will pass through the filters
    during the washes.
  • If the sample was already concentrated then it
    can be diluted. This action can reduce the
    concentration of the inhibitor and in turn reduce
    its effect on the PCR reaction.

56
IPC
  • If neither of these work then the sample can be
    re-extracted using a cleaner extraction method.
    Some extraction techniques like organic
    extraction produces a cleaner end product.
    Inhibitors are usually removed during organic
    extraction.
  • If this is not possible then a new sample may be
    taken from the evidence that may be free of dyes
    or dirt or other materials that could have caused
    the inhibition.

57
IPC
  • If the IPC does not cross the threshold and the
    sample has a low CT value and a high ?Rn that
    means there is a lot of target DNA present. In
    this case the DNA may have out competed the IPC
    for certain components necessary for
    amplification.

58
IPC
  • Earlier it was said that the IPC and target DNA
    have different primers so they did not compete
    with each other for those. However there can be
    competition for polymerase and other components
    like Mg needed for PCR.
  • If the sample has a high CT value and a low ?Rn
    and the IPC did not amplify it could mean there
    was partial inhibition. Depending on how low or
    high the quantity is the sample may still amplify
    and give a result during STR amplification or the
    sample may have to be cleaned up before
    continuing with the DNA analysis process.

59
Using Quantifiler results
  • Should a sample that gives a negative result be
    taken on for further DNA analysis?
  • Some laboratories choose not too carry the sample
    any further in an attempt to conserve money and
    time. During quantifiler the sample is put
    through 40 cycles of PCR. The PCR that occurs in
    the next stage of DNA analysis only puts the
    sample through 28 cycles. Theoretically if a
    sample does not amplify in 40 cycles it will not
    amplify in 28 cycles.
  • As a result a true negative quantifiler result
    will not be expected to generate a profile. The
    practice of dropping a sample after quantitation
    is laboratory dependant.

60
Using Quantifiler results
  • If after quantitation there is DNA in your sample
    then the extract must be taken on to STR
    amplification.
  • The final value from quantitation will be in the
    units
  • ng/µl
  • This simply means that for every 1 µl of extract
    there is a certain amount of DNA.
  • STR amplification kits require a certain amount
    of input DNA to give optimum results. Meaning a
    full profile with good peak heights.
  • Usually targeting an input of 1ng of DNA for STR
    amplification will produce a good profile.

61
Using Quantifiler results
  • If final quantity is 1ng/µl then by simply adding
    1µl one will have the necessary amount of target
    DNA.
  • What if value is 4.25 ng/µl.
  • For every 1µl there is 4.25 ng of DNA. So 1µl of
    extract will be greater than optimum input level
    for the amplification kit. This sample needs to
    be diluted before it can be amplified.
  • Calculation
  • 4.25ng 1µl
  • 1ng 1 / 4.25 0.235µl

62
Using Quantifiler results
  • Calculation
  • 0.235 µl of a sample is too little to be added.
    Pipetting error is common when using such low
    quantities of sample. By diluting the sample,
    more of diluted sample can be added and pipetting
    errors can be reduced.
  • Dilution by a factor of 10
  • 1 µl of extract to 9µl of TE buffer.
  • After dilution by a factor of 10 simply multiply
    volume by 10
  • 0.235µl x 10 2.35 µl
  • So for a quantity of 4.25ng / µl. It should be
    diluted by a factor of 10 and 2.35 µl of diluted
    sample be added to amplification tube.

63
Using Quantifiler results
  • At end of quantitation the quantity of DNA is
    0.365ng/µl.
  • So 1µl of sample will not give 1ng of DNA.
  • To calculate how much µl will be needed to give
    1ng of DNA.
  • 0.365 ng 1µl
  • 1 ng 1/ 0.365 2.73 µl
  • So you will need to add 2.73 µl of extract to get
    1 ng of DNA for amplification.

64
Using Quantifiler results
  • If you are using a 25 µl reaction for the STR
    amplification then the amount of sample being
    added is limited.
  • If the quantity of DNA is too low then it will
    require more than the amount of DNA that can be
    added to the reaction.
  • In these cases it is advisable to concentrate DNA
    and quantitate again before going on to
    amplification.

65
Using Quantifiler results
  • At the end of this process samples should be
    ready for STR amplification.
  • As long as the quantifiler results show there is
    enough DNA that the amplification kits require to
    generate a full profile then the analyst should
    be confident they will be able generate a full
    profile at end of STR analysis.

66
Using Quantifiler results
  • One important limitation to consider is if there
    is a mixture of DNA in the sample the quantity
    generated is how much total DNA is present.
  • The resulting profiles may be two or more
    incomplete profiles or one full profile and one
    incomplete profile.
  • Total Human quantitation cannot tell if the final
    value is from more than one person.
Write a Comment
User Comments (0)
About PowerShow.com