Title: Current Meter Maintenance HS Model OSS-B1
1Current Meter MaintenanceHS Model OSS-B1
- Hydrological Services Pty Ltd
- Presented to HiQ Engineering
- Brunei
- JULY 2009
2Section 1Prior To Discharge Measurements
3Prior To Discharge Measurements
- Remove fan assembly from carry case
- Turn fan carrier anti-clockwise
Anti-Clockwise
4Prior To Discharge Measurements
- Separate fan carrier from shaft assembly (keep
away from contact with any dirt particles) - Hold the fan upright and ¾ fill with approved
HS current meter oil
Use Lint Free Tissue
5Prior To Discharge Measurements
- Replace axle and fan carrier, excess oil will be
forced up through the capillary gap around the
axle, ensure fan carrier is tightened -
- Wipe the excess oil of the meter using lint free
paper towel
Clockwise
6Prior To Discharge Measurements
- Do not forget to close the oil bottle straight
after usage to avoid contamination of the oil - Slide the fan assembly into the body and use the
stubby screwdriver to tighten the retaining
screw to lock the fan assembly in place
7Prior To Discharge Measurements
- Prior to spin test hold the current meter in the
horizontal position and spin by hand for approx.
one minute to get rid of the hydraulic pressure - Spin test the meter against time to test the
performance of the shaft and bearings
8Prior To Discharge Measurements
- Spin test results are dependent on the fan type
- Connect your current meter to the counter using
the lead provided (this will test that the lead,
reed switch and counter are working prior to use)
OSS-B1 Spin Test OSS-B1 Spin Test
Fan No. Minimum Time (secs)
1 90
A 110
2 90
4 90
9Section 2During and After Discharge Measurements
10During and after Discharge Measurements
- After every discharge measurement, (or one
hour of use), the current meter should be cleaned
and refilled with oil, if continuing a discharge
measurement.
11During and after Discharge Measurements
- Undo the cleaning jar cap and ¾ fill with white
spirit - Remove fan assembly from meter body by undoing
the retaining screw and separate the body from
the fan
Cleaning Jar
12During and after Discharge Measurements
- Remove fan carrier and shaft assembly (do not
place in contact with dirt particles) - Empty the old oil in a recycling container and
dispose (do NOT re-use)
13During and after Discharge Measurements
- Put the shaft and bearing assembly into the
cleaning jar. - Agitate the jar
- Fill up the fan with white spirit, place your
thumb over the hole and shake vigorously for a
few seconds.
14During and after Discharge Measurements
- Blow dry the shaft assembly, fan carrier and fan.
- Re-assemble the fan, dry if discharge
measurement is complete
15During and after Discharge Measurements
- Wipe box clean and return the fan assembly to the
case - If continuing a discharge measurement, refill the
fan with oil
16During and after Discharge Measurements
- Spin test the current meter against time as shown
in section 1 - After the spin test is performed connect your
current meter to the counter using the lead, to
test the lead, reed switch and counter.
17Section 3Office Cleaning
18Office Cleaning
- Office cleaning should be performed on a
monthly basis or whenever the meter is dirty
19Office Cleaning
- Separate fan carrier from shaft assembly
Anti-Clockwise
20Office Cleaning
- Separate fan carrier (1) from shaft assembly (2)
1
2
21Office Cleaning
Remove Bearing
Remove Nut
22Office Cleaning
Remove Outer Spacer
Remove Inner Spacer
23Office Cleaning
Fan assembly fully dismantled ready to clean
Remove Bearing
24Office Cleaning
Place items in sink, filled with hot soapy water,
soak for at least 5 minutes
Clean items with a bottlebrush cleaner
25Office Cleaning
Cleaning Fan
Cleaning Fan Carrier
26Office Cleaning
Cleaning outer spacer
Rinse full assembly with tap water
27Office Cleaning
- Place full assembly on lint free tissue
- Ready to dry and re-assemble
- Rinse all items in white spirit prior to
re-assembly
28Office Cleaning
Dry Fan
Dry Fan Carrier
29Office Cleaning
Dry Inner Spacer
Dry Outer Spacer
30Office Cleaning
Blow Dry Bearing Nut
31Office Cleaning
Assemble Inner Spacer
Start Shaft Assembly, Bearing First
32Office Cleaning
Assemble Bearing
Assemble Outer Spacer
33Office Cleaning
Dip shaft assembly in white spirit to lubricate
bearings
Assemble axle nut
34Office Cleaning
Blow Excess White Spirit
Assemble Axle With Fan Carrier
35Office Cleaning
- Re-assemble fan and place in carry case
Clockwise
36Section 4Calibration
37Calibration
- Re-calibrate the current meter every
300 hours of use or once a year which ever
comes first (see Hydrological Services
calibration certificate for details) - To minimise freight costs, send only
the body and fans requiring re-calibration - Note Hydrological Services provides a
re-calibration and repair service for all types
of current meters.