Title: InfraCal TOG/TPH Analyzer
1InfraCal TOG/TPH Analyzer,Model CVH Extraction
Procedure
www.wilksIR.com
2- The sample extraction can be done using either a
prescription bottle with a - septa cap or a separatory funnel
3- Extraction using a prescription bottle with a
septa cap
4- Collect a measured amount of sample
5- For samples that will be stored for more than a
day or for samples with a high pH - Adjust the pH to less than 2 with Hydrochloric
acid or Sulfuric Acid (typically 3-5 drops
depending on buffers in sample)
610 parts sample to 1 part solvent
For example For 150 ml of sample add 15 ml of
solvent
7 8- Open the cap on the bottle during the 2 minute
shaking to relieve the pressure
9- Solvent is heavier than water and will sink
10 11- Use the syringe to remove the solvent
12- Fill the cuvette with solvent
13- Place the cuvette in the InfraCal frosted side to
the front
14 15- Oil concentration displayed
14
16- Extraction using a separatory funnel
17- Pour measured sample into separatory funnel
18- Measure one tenth of the sample size of solvent
For example For 150 ml of sample add 15 ml of
solvent
19- Pour the solvent into the sample collection
container to rinse interior surfaces and cap of
left over oil
20- Pour the solvent from the sample container into
separatory funnel containing sample
21 22- Stop to vent during the 2 minutes shaking to
relieve the pressure
23- Put the separatory funnel into the stand
24- Allow the solvent to sink to the bottom
25- Fill the cuvette with solvent
26- Place the cuvette in the InfraCal frosted side to
the front
27 28- Oil concentration displayed
14
29Silica Gel Extraction The difference between TPH
(Total Petroleum Hydrocarbon) and TOG (Total Oil
and Grease) is the polar organics are removed
from the extract using silica gel. The remaining
hydrocarbons are the non-polar components
considered to be TPH
30- Silica Gel Extraction
- Put filter paper into funnel
- Add silica
31- Silica Gel Extraction
- Put filter paper into funnel
- Add silica
- Put funnel into bottle
- Put solvent extract through
- silica gel
32- Silica Gel Extraction
- Put filter paper into funnel
- Add silica
- Put funnel into bottle
- Put solvent extract through
- silica gel
33- Silica Gel Extraction
- Put filter paper into funnel
- Add silica
- Put funnel into bottle
- Put solvent extract through
- silica gel
- Put solvent extract into cuvette
34- Silica Gel Extraction
- Put filter paper into funnel
- Add silica
- Put funnel into bottle
- Put solvent extract through
- silica gel
- Put solvent extract into cuvette
- Put cuvette into InfraCal frosted side to the
front
35 36- Oil concentration displayed
14
37InfraCal TOG/TPH Analyzer,Model CVH Extraction
Procedure
For questions please contact us
at tech_at_WilksIR.com, or 203-855-9136