Title: Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC)
1Physikalisch-Meteorologisches Observatorium
Davos, World Radiation Center (PMOD/WRC)
The PMOD/WRC is the World Radiation Center of the
World Meteorological Organisation,
And is signatory of the Mutual Recognition
Arrangement by the Committee of Weights and
Measures (CIPM MRA).
2PMOD/WRC
PMOD/WRC
World Radiation Center
Absolute radiometry (ARS)
Infrared radiometry (IRS)
Optical depth WORCC
UV radiometry (WCC-UV)
3EMRP-ENV03 Solar UV
http//projects.pmodwrc.ch/env03/
- Enhance the reliability of spectral solar UV
radiation measured at the Earth surface - Develop new techniques and devices for
traceability better than 2 (now 5) and for
cost-effective array-spectroradiometer in UV
monitoring networks - Intercomparison Campaigns and Workshops
End-User Devices
Calibrated UV Network
Transfer Standard
Reference Spectroradiometer
Primary Irradiance Standard
- Project Coordination Julian Gröbner, Davos
- Duration 2011 - 2014
- Total Budget 3.9 M
- 8 Partners EU-NMI 2 Industry 2 Universities gt
3 Collaborators (Stakeholders)
4Traceability chain for spectral irradiance
Wm-2nm-1
Cryogenic radiometer 0.01
Detector
Source
cw-Laser Sources
Si-trap detector aperture
Spectrally tuneable source
Filter Radiometer
Blackbody aperture 0.3
Spectroradiometer
Spectral irradiance standard 1-3
Portable transfer standard (used for stability
check) 4
Spectroradiometer 5
Gröbner Sperfeld, Metrologia, 2006
5Validation of the QASUME irradiance reference at
PTB
Blackbody BB3200pg at PTB
QASUME Validation at PTB in 2004 and 2013
Gröbner J., and P. Sperfeld, Direct traceability
of the portable QASUME irradiance scale to the
primary irradiance standard of the PTB,
Metrologia, 42, 134139, 2005.
6UV Quality Assurance Program
- Quality site audit with the QASUME reference
spectroradiometer
- Status 2002 - 2013
- 63 site visits
- 34 sites
- gt150 spectroradiometer intercomparisons
Ny Ålesund, Spitzbergen, 2009
QASUME
Gröbner et al., Appl. Opt. 2005 Gröbner et al.,
Metrologia, 2006
http//www.pmodwrc.ch/euvc/euvc.php?topicqasume_a
udit
7RBCC-E _at_ El ArenosilloThe sunny place
- 1999, 2005, 2007, 2009, 2011, 2013
2005
8Daily ratios to QASUME The ideal case
9Brewer cosine error
10Brewer cosine error
11Single Brewer Stray light
12Brewer Temperature dependency
6
13Summary from 6th RBCC-E (2011)
14Wavelength scale
15What needs to be addressed?
- Temperature dependency
- Absolute calibration
- Wavelength dispersion
- Single Brewer Stray Light
- Angular Response
- Apply Correction,
- or better
- Entrance optic modification
16Spectral solar UV Intercomparison at Davos
Final ENV03 Workshop, 7-16 July 2014
International Pyrheliometer Comparison (IPC) 2010
Presentations, Guidelines, Publications can be
found at the project web-site
http//projects.pmodwrc.ch/env03/
17UV Quality Assurance Program
- Quality site audit with the QASUME reference
spectroradiometer
- Status 2002 - 2013
- 63 site visits
- 34 sites
- gt150 spectroradiometer intercomparisons
http//www.pmodwrc.ch/euvc/euvc.php?topicqasume_a
udit
18QASUME irradiance reference stability 2004 to
2012
19QASUME irradiance reference
- Based on 7 transfer standards directly calibrated
against the primary irradiance normal of the
Physikalisch-Technische Bundesanstalt (PTB), in
Germany.
PTB
Transfer standard 1000 W FEL lamp
Transfer standard 1000 W FEL lamp
Transfer standard 1000 W FEL lamp
EUVC/QASUME
250 W portable lamp
Portable QASUME irradiance reference
250 W portable lamp
250 W portable lamp
20Spectral UV irradiance site audit procedure
PMOD/WRC
Measuring Station
Spectroradiometer Brewer163
QASUME
Quality Assurance
Regular intercomparisons
Portable QASUME reference spectroradiometer
Local Spectroradiometer
QASUME irradiance reference traceable to PTB
Audit report and calibration certificate sent to
local operator and published on
www.pmodwrc.ch/euvc/euvc.html
21Comparison of spectroradiometers
QASUME
TEST
matSHIC
- Same wavelength scale
- Uniform slit width (1 nm default)
Ratio TEST/QASUME
22Wavelength scale