Title: EEM calculations in FRED
1EEM calculations in FRED
- Mikkel Brydegaard
- Applied Molecular Spectroscopy Remote
SensingAtomic Physics, LTH
2Project
3Components
- Fiber tip source
- Off axis parabolic mirror
- Quartz prism
- Achromat
- Sample window
- Chassis
4Source
- Ocean Optics ø230 µm circular fiber tip
- 24 acceptance angle
- Apperture to 6 for collimating mirror
- Simulated spectral content 300, 350, 400, 500
and 900 nm
5Collimating mirror
- Newport. Off-Axis Replicated Parabolic Mirror,
1.500 in. Dia, 4.000 in. EFL, Protected Aluminum
Coating
6Prism
- Esco products
- Equilatheral fused silica
- 40mm sides
- 35 mean dispersion 300-900nm
- 40cos(3035/2)27mm apperture
- Displacement of optical axis from center of
circumcicle of prism 8.93 mm
7Achromat
- Edmund optics
- Uncoated triplet
- CaF2 Fused Silica CaF2
- Provided raytracing object from edmund upon
signed request
8Sample slit
- Quartz microscope cover slide
- Mirror coating
- Nano church
9Chassis
- Imported from Alibre design
- Anodized aluminum
- Black lambertian 4
- Contribution
10Setup
Newportv Off-axis paraboloid
Edmund achromatCaF2-QS
EscoQuartzprism
Quartzwindow
Fiber tip
11Properties of interest
Simulation variables
- Throughput
- Spectral resolution
- Signal to background
- Ray numbers
- Allow reflections
- Allow surface scatters
12Resulting rays
End rays C1 C2 C3
R1 175 3855 24710
R2 2405 66122 430448
R3 14560 418750 2693084
13Simulation time
Calculation time C1 C2 C3
R1 0.933 sec 5.94 secs 31.2 secs
R2 6.25 secs 50.2 secs 8.76 mins
R3 35.6 secs 5.23 mins 54.2 mins
14Throughput
Throughput C1 C2 C3
R1 56,62 56,85 56,85
R2 60,95 61,23 61,23
R3 61,55 61,85 61,85
15FWHM
FWHM mm C1 C2 C3
R1 0.115 0.147 0.147
R2 0.146 0.146 0.146
R3 0.136 0.136 0.136
FWHM nm C1 C2 C3
R1 14.30 18.33 18.33
R2 18.14 18.14 18.14
R3 16.91 16.93 16.93
16Added background
Background C1 C2 C3
R1 0,00E00 4,70E-03 2,16E-04
R2 0,00E00 1,40E-03 1,65E-05
R3 0,00E00 9,09E-04 2,73E-06
17Signal and back ground contributions
1E-3
3.5E0
3E-6
18Spatial profile of background contributions
19Experiences
- Prism displacement
- Ommitment of achromat, in accordance with
Spectrograph design fundamentals / J.F. James. - More stray light along slit than perpendicular to
it - Parabolic mirrors in FRED