SIAMOIS:%20optical%20test%20bench%20BOSS%20(Banc%20Optique%20SiamoiS) - PowerPoint PPT Presentation

About This Presentation
Title:

SIAMOIS:%20optical%20test%20bench%20BOSS%20(Banc%20Optique%20SiamoiS)

Description:

Title: Diapositive 1 Author: Pernelle Bernardi Last modified by: Pernelle Bernardi Created Date: 4/20/2006 10:10:52 AM Document presentation format – PowerPoint PPT presentation

Number of Views:96
Avg rating:3.0/5.0
Slides: 9
Provided by: Perne1
Category:

less

Transcript and Presenter's Notes

Title: SIAMOIS:%20optical%20test%20bench%20BOSS%20(Banc%20Optique%20SiamoiS)


1
SIAMOIS optical test benchBOSS (Banc Optique
SiamoiS)
  • Pernelle Bernardi
  • Observatoire de Paris-Meudon, LESIA

2
The test bench objectives
  • Simulation of all the components, except the
    interferometer
  • Far field of the optical fiber
  • Qualification of the optical fiber connectors
  • Definition of the mechanical interface of the 2
    fibers
  • Imaging spectra on the camera
  • Testing acquisition softwares and data processing

3
Design of the test bench
4
Far field of the fiber
  • The far field of the optical fiber defines the
    size of the 5 zones of the steps mirror.

Acquisition of the beam profile with a CCD.
5
Qualification of the optical fiber
  • Two qualifications at -80C
  • Optical fiber only ? optical tests during cooling
  • Optical fiber connectors ? optical tests
    before/after cooling
  • Purchase of several fibers
  • Two manufacturers
  • Two different jackets PVC and metal

6
Mechanical interface of the fibers
  • Specification distance of 128 µm between the two
    fiber cores (signal calibration) at the
    collimator focus
  • Design of the mechanical interface of the 2
    fibers
  • Test of the robustness of this component
  • Evaluation of the constraints on the fibers

7
Implementation of the camera
  • Model iDus DU440
  • Active pixels 1024 x 255
  • Pixel size 26 x 26µm
  • Back illuminated, AR coated for optimal
    performance in the 350nm region
  • Full well capacity 465 ke- (typical)
  • Read noise 4 _at_ 33kHz
  • Typical minimum temperature -70C (air cooled)
  • Andor software (LabView)
  • Acquisition
  • Real-time data display
  • Shutter control
  • Data processing
  • 2 spectra
  • Calibration source

8
Conclusion
  • State of the test bench today (in Meudon labs,
    ex CoRoT CCD test bench)
  • Waiting for equipmentafter many difficulties to
    order it!
  • Beginning of the tests scheduled for june 2006
  • 2 students
Write a Comment