Title: Machine-specific QC in Aarhus using EPID
1Machine-specific QC in Aarhus using EPID
- Mai-Britt K. Jørgensen, Lone Hoffmann, L.P.
Muren, Lars H. Præstegaard, Rune Hansen, Jørgen
B.B. Pedersen, Maria F. Jensen - Aarhus University Hospital
- Department of Medical Physics
2Rutine QC
IMRT VMAT
Machine specific Sweeping gap Picket fence (PF) Sweeping gap Picket fence (PF) Dose rate versus gantry speed (DRGS) Dose rate versus MLC speed (DRMLC)
Patient specific PDI (Delta4) (PTW729 array) PDI (Delta4)
3Picket fence, PF
- For every MLC and gab
- Position
- Relative rotation of EPID and collimator
- Intensity (width)
4PF ex. Eye test
5PF ex. Deviation in position, tolerance 0.3mm
6PF ex. Deviation in intensity, tolerance 300
7PF ex. Eye test version 2
8PF In total 30 test on each accelerator, one
every month
9Dose rate vs. gantry speed, DRGS
10DRGS ex. typical deviation
11DRGS ex. cable issue
12DRGS In total 30 test on each accelerator, one
every month
13Dose rate vs. MLC speed, DRMLC
14DRMLC ex. segment 4 DR120, 0.4cm/s
15DRMLC ex. segment 2 DR600, 2.4cm/s
16DRMLC ex. segment 4 DR120, 0.4cm/s segment
1 DR480, 1.6cm/s
17DRMLC In total 30 test on each accelerator, one
every month
18Time consumption
- Measurements 10 min
- Analysis (home-made software) 1 min
- Total 11 min
19Effect of chain adjustments on the DRGS and DRMLC
tests
- Mehtod
- 3 DRGS and 3 DRMLC test with a loose chain
- 3 DRGS and 3 DRMLC test with a thight chain
- Result, comparing loose and thight chain
- DRGS Standard deviation was unchanged
- DRMLC Std 0.6 (loose) -gt 0.2 (tight)
20Loose chain unstable characteristic
21Tight chain Stable characteristic
22Discussion
- Deviations in PF are covered in next, next
session - Deviations in DRGS and DRMLC
- Chain adjustments have an effect on DRGS and
DRMLC test, but it does not solve everything - Cable issues
- Beam adjustments?
- For more information on the calculations and
tolerances - Med. Phys. 38, 1425 (2011)